INSTRUCTION MANUAL 

 

Oakton

®

 Temp-340 

Thermistor 

Datalogging Thermometer

 

Instruction Manual 

 

Temp-340 Thermistor  

Datalogging Thermometer 

Part of Thermo Fisher Scientific  68X536601

 

R0 Mar 2010

    

www.calcert.com

[email protected]

1.800.544.2843

0

5

10

15

20

25

30

Summary of Contents for Temp-340 Thermistor

Page 1: ...mp 340 Thermistor Datalogging Thermometer Instruction Manual Temp 340 Thermistor Datalogging Thermometer Part of Thermo Fisher Scientific 68X536601 R0 Mar 2010 www calcert com sales calcert com 1 800...

Page 2: ...MENT MODE 14 11 HOLD FUNCTON 15 12 MIN MAX and AVE FUNCTION 15 13 DATA LOGGING 15 14 SETUP MODE 16 15 GENERAL SETUP SCREEN 17 16 CALIBRATION SCREEN 19 17 ALARMS SCREEN 21 18 DATA LOGGING SCREEN 24 19...

Page 3: ...up to 2000 points USB output Operator selection of Celsius or Fahrenheit scale Resolution of up to 0 01 C F Large backlit LED and dot matrix graphic LCD display Hold feature for temporarily retain a r...

Page 4: ...be changed in an area known to be non hazardous CAUTION 1 Do not use or store this instrument in microwave ovens or any abnormally hot or cold areas 2 Weak batteries should not be left in the instrum...

Page 5: ...0 1 above 99 99 Accuracy From 40 00 to 99 99 F 40 00 to 99 99 C 0 06 F 0 03 C From 100 0 to 257 0 F 100 0 to 125 0 C 0 1 F 0 1 C From 257 0 to 302 0 F 125 0 to 150 0 C 0 9 F 0 5 C Display Backlit Dot...

Page 6: ...tion 50V rms Storage 40 C to 65 C 40 F to 149 F Humidity 10 to 90 non condensing Battery Life Size Three AA 1 5V Alkaline Life 400 hours continuous typical without backlighting Dimensions Without Armo...

Page 7: ...ess protection Meets IEC 529 IP 54 for dust and water resistant enclosures probe attached CE Compliance EN61326 1 A1 1998 EU EMC Directive www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20...

Page 8: ...ted but the battery life indicator 6 Indicator Voltage Cell 3 bars More than 4 1 V Cell 2 bars More than 3 6 V Cell 1 bars More than 3 0 V Cell empty bars flashing More than 2 85 V Adaptor power suppl...

Page 9: ...eing replaced 1 Before changing battery turn instrument off and disconnect Thermistor 2 Loosen screw and lift battery cover off back of case 3 Remove the three AA batteries 4 Insert three new batterie...

Page 10: ...er before pushing the bottom edges of meter down to set it into position Lift up the stand at the back of meter for bench top applications if necessary 2 To remove thermometer from armor push out from...

Page 11: ...BLING OPTIONAL HANDSFREE ACCESORIES You can use the optional magnets and strap in the Handsfree Kit accessories for hands free operations 9 www calcert com sales calcert com 1 800 544 2843 0 5 10 15 2...

Page 12: ...mating connector on the top of the instrument If no probe is connected the display will read open Thermistors are sensitive at the tip or sensing element When taking measurements allow time for the re...

Page 13: ...s F2 Toggle between F and C display F3 Toggle between menu and measure mode hold Freeze display on off light Turns meter on and off press and hold for 3 seconds to turn off Press momentarily to turn o...

Page 14: ...atrix display features a large primary display smaller secondary displays for channel info or min max ave and helpful annuciators for added measurement data 12 www calcert com sales calcert com 1 800...

Page 15: ...PC data logging indicator 10 Main reading display 11 Current reading unit indicator 12 Min Max Avg elapsed time 13 Current Min Max Avg reading indicator 14 Meter logging memory location 15 Measurement...

Page 16: ...n the primary display Pressing the F2 key will toggle reading between F and C display Pressing the F1 key initiates and toggles through Minimum Maximum and Average reading modes Pressing F3 enter acce...

Page 17: ...and maximum values are sensed and automatically stored To exit and clear this function press the F3 to access the Menu functions See the Clear Reset menu section for more details 13 DATA LOGGING Press...

Page 18: ...ions To enter a setup screen press Select F1 key To return to measurement mode press Meas F3 key Following menu options are listed 1 General Setup 2 User field calibration 3 Alarm settings 4 Data logg...

Page 19: ...1 to indicate you want to change the setting of the current parameter or recall or log to move to the next parameter Press recall or log to change the options Press F2 to choose the next setting Whene...

Page 20: ...o reset change password In the event if uses forget his her password 5586 can be used to eset to a new value r On the second page you can set time and date 18 www calcert com sales calcert com 1 800 5...

Page 21: ...for Thermistor off set error It is NOT necessary to perform a field calibration to obtain specified meter accuracy Use the field calibration feature to improve thermometer probe accuracy or to compen...

Page 22: ...or 9 00 F Slope adjusts at two points The two calibration points must be at least 20 00 C 36 00 F apart The 2nd calibration point should be at a higher value than the 1 st calibration point Use the or...

Page 23: ...the alarm setting options Measurement alarm Disable or enable the alarm by pressing recall or log and F1 to accept Increase or decrease individual limit by pressing recall or log Meter in alarm mode...

Page 24: ...sable the countdown alarm and set the countdown time from 5 sec to 1 hour After setting enabling the countdown alarm the measurement screen should look like this 22 www calcert com sales calcert com 1...

Page 25: ...f a countdown will stop the process You will need to restart the countdown by selecting the start key F2 Alternatively you can also choose Menu to go into the timer menu and disabling the alarm 23 www...

Page 26: ...tdown has expired To reset the timer select restart F2 it You display should look like this To repeat the functions select F2 Note The C F function is disabled when the count down function is enabled...

Page 27: ...rides the Auto shutoff until the countdown is completed or manually stopped If the meter is manually or auto shutoff The Countdown Alarm is automatically reset to Disable each time the meter is shut...

Page 28: ...hods as auto or manual If it is auto logging using recall or log to set time interval Its range is from 0min to 60min Data Transfer from Meter to Computer 26 Once the USB connection is establish with...

Page 29: ...MUST BE TURNED ON PRIOR TO CONNECTING USB CABLE OR COMPUTER MAY NOT RECOGNIZE THE INSTRUMENT 19 CALIBRATION REPORT SCREEN The Calibration report will show the time and date along with results of the l...

Page 30: ...o choose which data you want to clear or reset For calibration logged data and reset all you will have to enter the password to proceed Default Password is 9900 28 www calcert com sales calcert com 1...

Page 31: ...ITH DAMP CLOTH Do not clean with abrasives or solvents Use mild detergents never immerse nor use excessive fluid 23 BATTERIES If there is no display when the thermometer is turned on check condition o...

Page 32: ...Connector Display Shows Err If display shows this message other than during the field calibration mode please return the instrument for servicing Cannot connect to PC Ensure that the meter is powered...

Page 33: ...rmour with Stand 35427 80 Hands free Kit Two magnets and a strap 35427 85 General Purpose Probe Immersion into liquids 08491 06 Flexible General Purpose Probe Vinyl sheath 08491 02 USB cable 35427 86...

Page 34: ...hose resulting from product misuse or abuse may be invoiced to you 27 PRODUCT RETURN To limit charges and delays contact the seller or Manufacturer for authorization and shipping instructions before r...

Page 35: ...and other quality standards Conformity ISO IEC 17025 2005 accredited NIST Handbook 150 2000 Edition ANSI NCSL Z540 2 1997 NIST Technical Note 1297 ISO 9000 2000 Fast Service Our substantial inventory...

Page 36: ...on you need to meet your most stringent quality requirements for the control of inspection measuring and test equipment Certification includes certificate of calibration with test data including descr...