10~100 mS/cm 

0.94 

 

b) Restore to factory default: 

Select Yes to restore the calibration to the theoretical values and parameter 

setting to original values. When meter’s calibration or 

measurement performs 

abnormally, this function can be adopted so the meter goes back to factory 

default setting and then users can conduct calibration or take measurements 

again.   

 

 

 

 

 

 

7. Technical Specifications 

TDS 

Range 

0

100.0 ppm, 0

1000 ppm, 0

10.00 ppt 

Resolution 

0.1/1 ppm, 0.01 ppt 

Accuracy 

±

1% F.S 

Calibration points 

1-2 points 

Automatic Temp. 

Compensation 

0-50

˚

Temp. 

Range 

0-50

˚

Resolution 

0.1

˚

Accuracy 

±

0.5

˚

 

 

 

 

8. Other Functions & Parameters 