5.5.1 According to the pH isothermal measurement principle, the closer the test sample’s temperature is to 
the calibration solution’s, the higher the accuracy of the measurement. This principle is recommended to 
follow when conducting tests for the best result. Example: If users were to test samples at 150˚F, we 
recommend warming up the calibration solutions to the same temperature before performing calibrations in 
order to get the most accurate readings.

 

5.5.2 The instrument has a function to return to factory default setting, which can be set up in P5 (refer to 
section 5.6). Returning to factory default setting is to restore the meter to theoretical value (zero potential pH 
is 7.00, slope is 100%), and set all the parameters to default settings. When the meter’s calibration or 
measurement is performing abnormally, users can use this function to let the meter return to factory default 
mode, and then conduct calibration and test again. Please note that this function is irreversible once used.

 

5.5.3 Self-Diagnosis

 

In the process of calibration and measurement, the meter has a self-diagnosis function. Please see table 5.

 

Table – 5 Self-diagnosis Information

 

Display 

Icons

 

Contents

 

How to fix

 

 

Wrong pH buffer solution or the buffer 
solution is out of range.

 

1. 

Check whether pH buffer solution is correct.

 

2. 

Check  whether  the  meter  connects  the 
electrode properly.

 

3. 

Check whether the electrode is damaged.

 

4. 

Make  sure  there  is  no  bubble  in  the 
electrode’s bulb.

 

 

pH  electrode  zero  electric  potential 
out of range (<-60mV or >60mV)

 

 

1. 

Make sure there is no bubble in the 
electrode’s bulb.

 

2. 

Check whether the electrode is damaged.

 

3. 

Check if pH buffer solutions are correct. 

 

4. 

The pH Electrode is aged. Replace a new 
one.

 

 

pH electrode slope is out of range

 

(<85% or >110%)

 

 

Press          key  when  reading  is  not 
stable during calibration.

 

Press          key

 

when          icon appears and stays 

on the screen.

 

 

During calibration, the measuring value 
is not stable for ≥3min.

 

1. 

Make sure there is no bubble in the 
electrode’s bulb.

 

2. 

Check whether the electrode is damaged.

 

3. 

The pH Electrode is aged. Replace a new 
one.

 

 

 

* If you find any air bubble in the glass bulb of the pH sensor, invert the probe for a few times to remove 
it.  The  existence  of  an  air  bubble  in  the  glass  bulb  will  significantly  decrease  the  stableness  of 
measurement. 

 

 

 

 

 

Summary of Contents for PH550

Page 1: ...PH550 Benchtop pH Meter Instruction Manual Oakton Instruments www 4oakton com Manual Version 3 2...

Page 2: ...4 5 Data storage recall and deletion 4 4 6 Max Min Function 5 5 pH Measurement 5 5 1 Preparation 5 5 2 pH Electrode Information 5 5 3 Information regarding Calibration 6 5 4 pH Meter Calibration 8 5 5...

Page 3: ...g technology intelligently improves the meter s response time and measuring accuracy with stable reading display mode available The kit comes with a 3 in 1 combination pH Electrode measuring pH and te...

Page 4: ...bration icons and numberings 3 Measurement unit 4 Temperature and reminder icons 5 The pH unit displayed during calibration 6 The pH value displayed during calibration numberings for storage and recal...

Page 5: ...on and go back to measurement mode Short Press In calibration mode press to conduct calibration In parameter setting press to confirm selection Short Press Store measuring data Long Press Recall store...

Page 6: ...deletion Storage When the measurement is stable press key the meter displays M icon and storage numberings on LCD and stores the measuring data In Figure 4 the meter stores the first group of the meas...

Page 7: ...ectrode holder The electrode holder is composed of a metal stand base a supporting pole and an electrode clip First install the metal stand base to the meter screw it on Please note that the metal bas...

Page 8: ...water Gently stir the electrode for a few seconds after it s dipped into the test solution and then let it stand still Wait for the stable measurement a smiley face appears and stays and then record...

Page 9: ...n of electrodes and the requirement of the accuracy For High Accuracy meaustements 0 02pH the meter should be calibrated before test every time For ordinary accuracy measurements 0 1pH once calibrated...

Page 10: ...d water dry it and dip it into pH 4 00 buffer solution Stir the solution gently and let stand still in the buffer solution until a stable reading is reached The meter s display will show the recogniti...

Page 11: ...e purified water A small addition of KCL solution will help with the measurement time 2 Purified water such as distilled water and deionized water are recommended for only rinsing the probe for the be...

Page 12: ...bration and measurement the meter has a self diagnosis function Please see table 5 Table 5 Self diagnosis Information Display Icons Contents How to fix Wrong pH buffer solution or the buffer solution...

Page 13: ...menu from P1 to P2 P5 For details please see Table 7 Table 7 pH Parameter Setting P1 Select pH buffer series USA NIS 1 Long press to enter P1 2 Press USA flashes press to select USA or NIS press to co...

Page 14: ...calibration accuracy the pH of the standard buffer solution must be reliable The buffer solution should be refreshed often especially after heavy use We recommend the Oakton buffer assortment pack 059...

Page 15: ...easuring value use ORP standard solution to test mV value and see whether ORP electrode or meter works properly Table 9 is the data of standard Zobell ORP solution for 222 mV Table 9 C 10 15 20 25 30...

Page 16: ...lfunctioned or damaged product attributable to responsibility of OAKTON INSTRUMENTS for a period of THREE YEARS SIX MONTHS for the probe from the delivery This limited warranty does not cover any dama...

Reviews: