13 ADDENDUM 1: METER FACTORY DEFAULT
SETTINGS
Resetting the meter to factory default settings clears all calibration data and most other setup
functions you might have changed. The following settings will remain as you have set them:
•
Temperature unit of measure (°C or °F) –
only for pH 310 meter
•
The temperature offset calibration value.
•
Clock
function
–
only for pH 310 meter
Type Parameter
P
No.
Default
Remarks
CLr
Memory clear
P1.0
No
Retains your current settings
CAL
View pH calibration data
P2.1
––
Calibration data for 1st buffer
P2.2
––
Calibration data for 2nd buffer
P2.3
––
Calibration data for 3rd buffer
P2.4
––
Calibration data for 4th buffer
P2.5
––
Calibration data for 5th buffer
(only in pH 310 meter)
P2.6
––
Calibration data for 6th buffer (DIN only)
ELE
View electrode offset
P3.1
0.00 mV
No offset adjustment
View electrode slope
P3.2
100.0 %
No slope adjustment
View relative mV offset
P3.3
0 mV
No mV adjustment
COF
Ready indicator
P4.1
Ready On
Ready indicator on; auto endpoint off
# pH calibration points
P4.2
3
3 pH calibration points available (1 – 5 pt
range)
Calibration buffer selection
sets(
pH 310 meter only
)
P4.3 USA
°C or °F (
pH 310 meter
)
P4.4
No default
°C or °F remains as selected
CLO
Setting year, date and time
(
in pH 310 meter only
)
P5.0
No default
Year, date and time remain as selected
rSt
pH factory default
For pH 300 meter
For pH 310 meter
P5.0
P6.0
No
Retains your current settings