5 MEASUREMENT
5.1
Taking pH Measurements
5.1.1
Automatic Temperature Compensation
Automatic Temperature Compensation only occurs when a compatible temperature sensing
element is plugged into the meter.
If there is no temperature sensor plugged into the meter, the
default manual temperature
setting is 25 °C
. You can manually set the temperature to match your working conditions
using a separate thermometer.
For automatic temperature compensation (ATC) simply
plug the temperature probe into the meter (see page 6
for directions). The ATC indicator will light up on the
LCD. See figure on right.
Figure 13: Automatic Temperature
Compensation
NOTE
The temperature sensing element must be in the
sample that you are calibrating or measuring.