Manual Calibration

In manual calibration, the tester allows 1-point calibration for each measuring range. You can 

use customized calibration solutions with known conductivity/TDS values to calibrate the tester.

The following table shows acceptable conductivity/TDS ranges of calibration solutions for each 

measuring range. Make sure your calibration solutions are within the given ranges.

Selected Measuring Range 

Acceptable Calibration Standard Range

 

 

  Conductivity TDS/Salt

LO   

 

 

200 - 2000 μS/cm 

200 - 2000 ppm

HI   

 

 

2.00 - 20.00 mS/cm 

1.00 - 10.00 ppt

AUTO 

 

 

Select a calibration standard nearer to application sample

To prepare calibration standards:

Use fresh calibration solutions. Measure conductivity/TDS values of the solution with a meter 

known to be accurate. Prepare each solution in two beakers - one for rinsing and the other for 

calibration. Rinse the electrode in de-ionized water before calibration.

To begin manual calibration:

1. Switch on the tester. Make sure the tester is in measuring mode. Rinse the electrode with the 

calibration standard that you intend to calibrate and then dip the electrode in the other beaker 

with same calibration standard. Swirl gently to create a homogenous sample and allow time 

for the reading to stabilize.
2. Press INC or DEC key to enter calibration mode. The ‘CAL’ indicator appears in LCD. The display 

briefly shows ‘CAL’ and the number of points the tester will be calibrated.
3. The upper display shows the measured conductivity/TDS reading of the solution based on 

previous calibration (if any) and the lower display shows the default (uncalibrated) conductivi-

ty/TDS reading.

Note

: The tester shows error message ‘Er.1’ :

(a) If the reading is over range (Or) of selected measuring range of the tester , or
(b) If the default (uncalibrated) reading is not within the acceptable calibration standard ran-

ge.
Use INC and DEC keys to adjust the upper display to the correct conductivity/TDS value of the 

calibration solution.

Note

: The calibration adjustment window is ± 50% from the default reading.

Note

: If you do not press INC or DEC key within 5 seconds, the tester shows the confirmation 

‘CO’ and returns to the measurement mode. However, the tester is not calibrated to new values 

yet. The old calibration is still active. If this happens, press INC or DEC key once again to enter 

calibration mode.
4. Wait for 5 seconds for the tester to automatically confirm the calibration by displaying ‘CO’ 

and return to the measurement mode.

Note:

 To exit calibration mode without confirming the calibration, press HOLD/ENT key before 

the automatic confirmation takes place.

Note

: The tester shows error message ‘Er.0’ and returns to measuring mode if the temperature 

of the calibration solution is not within 0°C to 50°C.
5. For multi-point calibration rinse the electrode in de-ionized water and repeat step the above 

steps with another standard solution.

TDS Factor Setting

TDS factor is only applicable for TDSTester11 models. The factory default TDS factor is 0.71. You 

can adjust the TDS factor to suit different samples of your applications.
To change TDS factor:
1. Switch on the tester. Make sure the tester in measurement mode. Press HOLD key to bring 

the tester to the HOLD mode.
2. Press INC or DEC key to enter the TDS factor setting mode.
3. The upper & lower displays of LCD show the last configured TDS factor. The upper display is 

adjustable. Use the INC or DEC key to adjust the TDS factor.
The adjustable range is 0.4 to 1.0

Note

: If you do not press INC or DEC key within 5 seconds, the tester shows the confirmation 

‘CO’ and returns to measurement mode.
4. Wait for 5 seconds for the tester to automatically confirm the new setting by displaying ‘CO’ 

and return to the measurement mode.

Temperature Calibration

Temperature calibration need not be performed every time, unless the temperature reading 

differs from that of an accurate thermometer. If temperature calibration is performed, Conduc-

tivity/TDS/Salt calibration is mandatory.
1. Switch on the tester. Make sure the tester is in measuring mode. If required, press °C/°F key 

to select the desired unit of measurement for temperature (Celsius or Fahrenheit). Dip the tester 

into a solution of known temperature and allow time for the temperature reading to stabilize.
2. Press INC or DEC key to bring the tester to the calibration mode. CAL’ indicator appears in 

LCD. Immediately press °C/°F key to switch to the temperature calibration mode.

Note

: When you enter calibration mode, if the conductivity/TDS/salt reading is outside the 

specified range (Or), the tester shows ‘Er.1’ error message. You can still proceed to the tempe-

rature calibration mode by pressing °C/°F key immediately. If the °C/°F key is not pressed within 

2 seconds, the tester exits the calibration mode and returns to the measurement mode.
3. The upper display shows the measured temperature reading based on the last set offset 

(if any) and the lower display shows the default (uncalibrated) temperature reading based on 

factory y settings. Use INC and DEC keys to adjust the upper temperature reading to the known 

temperature value of the solution.

Notes:

 The temperature adjustment window is ± 5°C (± 9°F) from the default reading.

4. Wait for 5 seconds for the tester to automatically confirm the temperature calibration value 

by displaying ‘CO’ and return to the measurement mode.

Note

: To exit temperature calibration mode without confirming the calibration, press °C/°F key 

or HOLD/ENT key before the automatic confirmation takes place.

Note

: The tester shows error message ‘Er.0’ and returns to measuring mode if the temperature 

of the solution is not within 0°C to 50°C.

Reset

Reset option allows you to restore the calibration and other parameters back to factory default 

settings.
1. Switch off the tester. Press and hold the HOLD key and then switch on the tester using ON/OFF 

key. Release HOLD key.
2. The lower display shows ‘rSt’ (reset) and the upper display blinks ‘No’. Use INC or DEC key 

to select ‘Yes’ (to proceed with resting) or ‘No’ (to quit without resetting).

Note

: Press °C/°F key if you wish to skip to measurement mode without making any selec-

tion.

3. Press HOLD key to confirm your selection. LCD shows ‘CO’. If ‘Yes’ is selected, the tester 

resets to its factory default values as listed below. LCD shows power-up sequence and tester 

goes to measurement mode.

Parameter 

    Factory 

Default

User calibration (conductivity/TDS/salt) 

 

(Reset)

Temperature unit of measurement   

 

Celsius (°C)

Temperature 

offset 

   0

Auto calibration (for ECTester11) 

 

 

Enable

1-point 

calibration 

   Enable

Conductivity calibration factor (for ECTester11)   

1.0

TDS factor (for TDSTester11) 

 

 

0.71

Changing Batteries

Replace the batteries when the low battery indicator starts blinking.
1. Open the battery compartment lid (with attached lanyard loop).
2. Remove old batteries by pulling plastic ribbon.
Replace with fresh ones.

Figure 11: Battery compartment

Electrode Maintenance

1. Always keep the sensor electrodes clean. Rinse the electrodes with de-ionized water and wipe 

them dry with clean cloth before storing with its protective cap. 

Model

ECTester11
19 46 81

TDSTester11
19 47 01

SALTTester11
19 47 11

Range: 

Lo
Hi

0 to 2000 μS/cm
0 to 20.00 mS/cm

0 to 2000 μS/cm
0 to 10.00 mS/cm

-
0 to 10.00 ppt

Resolution: 

Lo
Hi

10 μS/cm
0.10 mS/cm

10 ppm
0.10 ppt

-
0.10 ppt

Accuracy

± 1% of Full Scale

Calibration Type

Auto or Manual

Manual

Manual

Calibration Points

1 or 2 points

1 or 2 points

1 point

Calibration Window

± 50% from each point

Calibration Standard

Range (Manual)  

Lo
Hi

 

200 - 2000 μS/cm      

 .

      

2.00 - 20.00 mS/cm

 

200 - 2000 ppm               

 .

 

1.00 - 10.00 ppt

 

1.00 - 10.00 ppt

Sensor Type

Two-pin

TDS Factor

0.4 to 1.0 (Default 0.71)

Temperature
Range in °C
Range in °F
Resolution
Accuracy
Calibration point
Calibration Window

ATC
Temp Coefficient

0.0 to 50.0°C
32.0 to 122°F
0.1°C (0.1°F)
±0.5°C (±0.9°F)
1 point
± 5°C (± 9°F)from factory default
0 to 50°C
2% per °C

Normalization Temp

25.0°C

Auto Off

8.5 minutes after last key press

Operating Temp

0 to 50°C

Power Battery

4 X 1.5V”A76“ micro alkaline battery

Battery Life

>150 hrs

LCD Display

Custom Dual Display 27mm (H) X 21 mm (W)

Dimensions, Weight

Tester: 16.5 cm X 3.8 cm; 90g 

Boxed: 22cm X 6cmX 5cm; 170 g

For cup type electrodes, remove the white 

plastic cup and insert to thoroughly clean 

viscous solutions. Never scratch electrodes 

with a hard substance.
2. For better performance, soak the electrode 

in alcohol for 10 to 15 minutes and rinse 

with de-ionized water before starting any 

measurement process.
This is to remove dirt and oil stains on the 

electrode which may affect the accuracy of 

the measurements.

Electrode Replacement

When the tester fails to calibrate or gives 

fluctuating readings in calibration standards, 

you need to change the electrode module. You 

can replace the electrode module at a fraction 

of the cost of a new tester.
1. With dry hands, grip the ribbed tester collar 

with electrode facing you. Twist the collar 

counter clockwise (see Figure13-A). Save the 

ribbed tester collar and O-ring for later use.
2. Pull the old electrode module away from 

the tester.
3. Align the four tabs of the new electrode 

module so that they match the four slots on 

the tester (see Figure 13-B).
4. Gently push the module into the slots to sit 

it in position. Push the smaller O-ring fully onto 

the new electrode module. Push the collar over 

the module and thread it into place by firmly 

twisting clockwise.

Note

: It is necessary that you recalibrate the 

tester prior to measurement after an electrode 

replacement.

Accessories

194716 

Replacement Sensor MicroDirect  

  

EC/TDS/SALT11

Warranty

The waterproof testers are warranted to be 

free from manufacturing defects for 2 years 

and electrode module for 12 months, unless 

otherwise specified. If repair, adjustment or 

replacement is necessary and has not been 

the result of abuse or misuse within the time 

period specified, please return the tester - 

freight prepaid - and correction will be made 

without charge. Out of warranty products will 

be repaired on a charge basis.

Return of Items

Authorization must be obtained from your 

distributor before returning items for any re-

ason. When applying for authorization, please 

include information regarding the reason the 

item(s) are to be returned.

Note:

 Tintometer reserve the right to make 

improvements in design, construction and 

appearance of products without notice. Prices 

are subject to change without notice.
Technical changes without notice
Printed in Germany 07/11