Operating Manual

EcoTestr   CTS1 Pocket Tester

  The batteries have been pre-installed. Please pull out the 

insulating paper before first-time use. When replacing batteries, 

the positive side (+) of every single battery must face up. Please 

check the detail in "Battery Installation" section.

• 

 Before shipment, the factory adds in some water droplets to the 

tester's cap. This ensures that the probe keeps its sensitivity. This 

practice of adding several drops of water into the cap to keep 

sensor activation is highly recommended when storing the tester.

Calibration  

1. The tester only needs to be calibrated for conductivity, then 

    press       to switch to Conductivity → TDS→ Salinity.

2. Press       and switch to conductivity mode, rinse the probe in 

distilled water and shake off excess water.

3. Long press       to enter calibration mode; short press       to exit.

4. Dip the probe into 1413 μS/cm calibration solution. Stir gently, 

leave it to stand. Wait for the measurement stability icon (

) to 

appear and stay on the display, then short press       to complete 

the 1st calibration. Tester returns to measurement mode, and 

calibration icon 

“M” 

appears on the bottom left side of display.

5. Rinse probe in distilled water and shake off excess water. Repeat 

#3 and #4 to complete 2nd calibration in 12.88 ms/cm calibration 

solution. Tester returns to measurement mode, and calibration 

icon 

“M” 

and 

“H” 

will appear on the bottom left side of display.

1. Short press       to turn on the tester, short press       to switch to 

the desired measurement mode (Conductivity, TDS or salinity). 

Rinse probe in distilled water and shake off excess water.

2. Stir the probe in the sample solution gently, leave it to stand. Wait 

for the stability icon (

) to stay on screen, then take the reading.

3. Rinse off the probe thoroughly in distilled water after each test.

•  

If probe response is slow, we recommend soaking the probe in 

12.88 mS/cm solution for 15 to 30 minutes to restore its 

sensitivity. Re-calibrate the tester after rinsing. Please add 

several drops of water into the cap to keep sensor activation 

when storing the tester.

•  

The tester adopts 1413 μS/cm and 12.88 mS/cm standard 

calibration solutions. User can use 1 or 2 point calibrations as 

    needed. For most circumstances, calibrating in 1413 μS/cm to 

   Restore to factory default (P5) description: Select "Yes" in P5 to 

restore the calibration to the theoretical values and parameter 

setting to original values. When tester's calibration or 

measurement performs abnormally, this function can be adopted 

so the tester returns to factory default setting and then users can 

conduct calibration or take measurements again.

 
 
 
 

Conductivity of the solution 

T

DS

 factor (P2) 

0

 to 1

00

 

µ

S/

cm

 

0

.6

0

 

1

00

 to 1

000

 

µ

S/

cm

 

0

.71 

1 to 1

0

 m

S/

cm

 

0

.81 

1

0

 to 1

00

 m

S/

cm

 

0

.94 

  Conductivity range (P1) description: Aut - all ranges: 0~2000 

    μS/cm, 0.01~20.00mS/cm. Once the readings are greater than 

2000 μS/cm, the unit will automatically become mS/cm. 

    1 mS/cm = 1000 μS/cm. μS - only μS/cm range: 0~2000 μS/cm; 

mS/cm - only mS/cm range: 0.01~20.00 mS/cm.

  TDS factor (P2) description: User can adjust TDS factor by 

experimental data or experience. The following chart lists some 

commonly used TDS factors for reference: 

S

ymbol 

Menu setting 

S

election 

Factory default 

P1 

S

elect conductivity range*                Aut* – 

μS

 – m

S

   

Aut 

P2 

S

elect T

DS

 factor 

 

0

.4

0~

1.

00

 

0

.71 

P3 

 

S

elect salinity unit 

  ppt – g

/

ppt 

P4 

S

elect temperature unit 

 – 

 

/

 

P5 

Restore to factory default 

No 

– 

Y

es 

No 

 

When test is off, long press       to enter setup mode → short 

press      to switch P1-P2...P5, short press       and parameter will

flash →press       to choose desired parameter → short press       to
confirm parameter selection → long press       to return to

measurement mode.

Keypad Functions
Short press 

=< 2 seconds

Long press 

=> 2 seconds 

CAL

MODE

1. Short press to turn on; long press to turn off.

2. When turned off, long press to enter setup mode.
3. In mode setting , short press to change parameter.

1. In measurement mode, long press to enter calibration mode, 

    short press to switch Conductivity, TDS and Salinity.

2. In calibration mode, short press to confirm calibration.
3. In mode setting, short press to confirm parameter selection.

MEAS

complete 1st point calibration will meet testing requirements.

• 

 The tester has already been calibrated before leaving factory. 

User can use the tester immediately or can test it in the standard 

calibration solutions to evaluate its accuracy. If error is large, 

calibrate the tester before using.

  We recommend replacing the calibration solution after 5 to 10 

calibrations to maintain accuracy.

•  

The surface coating of probe sensor bar is firm platinum black 

processed with special techniques. It improves measuring 

performance and coating firmness. If the probe is contaminated. it 

can be gently brushed with a soft brush in warm soap water or 

alcohol.

Self-Diagnostic Messages 

Symbol Self-diagnosis

information 

How to fix 

• Check if calibration solution is 

  correct. 

• Check if probe is damaged. 

           is pressed 

before measurement 

is stable (

 appears 

and stays)

Wrong calibration 

solution, or measured 

value is not within the 

range of the tester

Wait for the measurement stability 

icon (

) to appear and stay, 

then press 

TM

Special Notes

Measurement (Conductivity, TDS & Salinity)

Notes:

Setting the Parameter

Notes:

Reviews: