5.2 The tester adopts 1413 µS and 12.88 mS standard calibration solutions. Users 

can use 1-2 point calibrations as needed. For most circumstances, calibrating in 

1413 µS to complete 1

st

 point calibration will meet testing requirements. 

 

5.3 Unit Information: 1 mS/cm = 1000 µS/cm 

Once the readings are greater than 2000µS/cm, the unit will automatically 

become mS/cm. 

 

5.3 The tester has self-diagnosis functions: 

Symbol  Self-Diagnosis information 

How to fix 

 

 

 
 

Wrong calibration solution, 
which exceeds the 
recognizable range of the 
meter. 

1. Check if calibration solution is 
correct 
2. Check if probe is damaged. 

 

 

 

       is pushed before 
measurement is stable 
(      comes up and stays) 

Wait  for  the  smile  icon  to  stay,  
and then short press      

5.4 The tester has already been calibrated after manufacture. Usually users can 

use the tester right away, or test it in the calibration solutions to test its 

accuracy. When error is large, calibrate it before using. 

6. Parameter Setting 

6.1 Parameter setting reference chart: 

Symbol 

Parameter Setting content 

Code 

Factory Default 

P1 

Select conductivity ranges 

Aut-µS-mS 

Aut 

P2 

Restore to factory default 

No 

 Yes 

No 

 

6.2 How to setup parameters: 

When turned off, long press          to enter setup 

 short press           to switch  

P1-P2

 Short press          , parameter flickering

short press            to choose, 

Summary of Contents for EC1

Page 1: ...EC1 Pocket Conductivity Tester Instruction Manual OAKTON INSTRUMENTS www 4Oakton com v2 2...

Page 2: ...TION OF BATTERIES WILL CAUSE DAMAGE TO THE TESTER AND POTENTIAL HAZARDS 2 Keypad Functions Short press 2 seconds Long press 2 seconds 1 Short press to turn on long press to turn off 2 When turned off...

Page 3: ...ete 2nd point calibration in 12 88 mS cm calibration solution tester returns to measurement mode calibration icons display on bottom left of LCD 4 Conductivity Measurement 4 1 Short press to turn on t...

Page 4: ...calibration solution is correct 2 Check if probe is damaged is pushed before measurement is stable comes up and stays Wait for the smile icon to stay and then short press 5 4 The tester has already b...

Page 5: ...original values When meter s calibration or measurement performs abnormally this function can be adopted so the meter goes back to factory default setting and then users can conduct calibration or ta...

Page 6: ...e 190 165 140mm 438g 9 Warranty We warrant this instrument to be free from defects in material and workmanship and agrees to repair or replace free of charge at option of OAKTON INSTRUMENTS any malfun...

Page 7: ...ntenance v modifications vi combination or use with any products materials processes systems or other matter not provided or authorized in writing by us vii unauthorized repair viii normal wear and te...

Page 8: ...OAKTON INSTRUMENTS PO Box 5136 Vernon Hills IL 60061 Tel 888 462 5866 www 4Oakton com...

Reviews: