User Reset

Reset the CTS calibration to the user’s default settings by using the user reset function. 

Temperature user calibration is not affected by the user reset function.
1.  Press MENU/

v

 

to enter setup window. Scroll down by pressing MENU/

v

 

to select Reset. 

Press HOLD/

   

to select Reset. The display shows User Reset and Factory Reset.

2. Press HOLD/

   

to select User Reset. 

3.  The display automatically shows No and Yes. Scroll down by pressing MENU/

v

  

to toggle between No and Yes. 

4.   Press  HOLD/

   

to confirm either No or Yes. Display shows User Reset option with a  . 

Factory Reset

Reset the CTSTestr to the factory default settings by using the factory reset function. 
1.  Press MENU/

v

 

to enter setup window. Scroll down by pressing the MENU/

v

  

to select Reset. Press HOLD/

   

to select Reset. The display shows User Reset and  

Factory Reset.

2.  Scroll down by pressing MENU/

v

 

to toggle between the resets. Press HOLD/

    

to select Factory Reset.

3.  The display automatically shows No and Yes. Scroll down by pressing MENU/

v

  

to toggle between No and Yes. 

4.  Press HOLD/

   

to confirm either No or Yes. The display shows the Factory Reset 

option with a  . 

Measure
Settings 
Temp Set

Factory Reset

hold

No

 

Yes

menu

hold

User Reset

Reset

Measure
Settings 
Temp Set

User Reset

hold

hold

No

 

Yes

menu

hold

Reset

Factory Reset

Specifications

CTSTestr 50 

Conductivity

Conductivity range

0.0 to 200.0 μS, 200 to 2000 μS, 2.00 to 20.00 mS

Resolution

0.1 μS, 1 μS, 0.01 mS

Relative accuracy

±1% full scale

Normalization temperature

25.0ºC (77ºF)

Temperature co-efficient

0.0% to 10.0%

Calibration points

Up to 3 points

TDS

TDS range

0.0 to 100.0 ppm, 100 to 1000 ppm, 0.10 to 10.00 ppt (TDS Factor 0.5)

Resolution

0.1 ppm, 1 ppm, 0.01 ppt

Relative accuracy

±1% full scale

Calibration points

Up to 3 points

TDS factor

0.40 to 1.00 (selectable)

Salinity

Salinity range

0.00 to 10.00 ppt

Resolution

0.10 ppt

Relative accuracy

±1% full scale

Calibration points

1

Temperature

Temperature range

0 to 60°C (32.0 to 140.0°F)

Temperature resolution

0.1°C / 0.1°F

Temperature accuracy

From 0 to 50ºC (±0.5ºC / ±0.9ºF + 1 LSD);  

from 50 to 60ºC (±1.0ºC / ±1.8ºF + 1 LSD)

Temperature compensation

Yes (Automatic Temperature Compensation)

General

Display

Graphics, dot matrix 80 x 100 pixel

Auto off

8.5 minutes (from last key press)

Reset

User / Factory

Power requirement

Four AAA 1.5 V batteries

Battey life

>150 hours

Water proofing

IP67

Environmental operating conditions

Ambient operating 

temperature

5 to 45°C / 41 to 113°F 

Relative humidity

5% to 85% noncondensing

Storage temperature

–20 to 60°C / –4 to 140°F

Storage humidity

5% to 85% noncondensing

Ordering Code

Product Description

35634-55

CTSTestr 50P pin-style pocket tester with case, lanyard, and batteries

35634-65

CTSTestr 50C cup-style pocket tester with case, lanyard, and batteries

35634-57

Replacement sensor module for CTSTestr 50P pin-style tester

35634-07

Replacement sensor module for CTSTestr 50C cup-style tester

35634-09

Replacement sensor cap

09376-00

Replacement alkaline batteries; AAA, 1.5 V. Pack of 12

17101-45

NIST-traceable calibration with data for pocket testers

Accessories

HOLD Function

This feature helps to freeze the display for a delayed observation.
1.   Press  HOLD/

   

button to freeze    

2. Press HOLD/

   

again to release    

    

measurement.  

 

                    measurement. 

1065O100_MAN_35634-55,-65 

                  August, 2017  

www.4oakton.com

Replacing the Batteries

The CTSTestr 50 Pocket Tester uses four 

AAA 1.5 V batteries. 
1.  To remove the battery cover, see Figure 1. 

Clear the front catch and then the back 

catch, before sliding the cover off. 

2.  To remove the battery plate, push the 

center tab towards the front of the tester 

as shown in Figure 2. Once unlocked, 

remove the plate to access the batteries.

3.  Invert the tester upside down to remove 

the batteries. Each side uses two  

AAA batteries. Orient each battery with 

positive terminal facing downward. 

4.  To lock the battery plate, align the small 

tabs (Figure 3) into the guide ribs on  

the housing and then press down.  

See Figure 4.

Fig. 3: Align tabs

Fig. 4: Push down to lock

Fig. 5: Battery plate unlocked

Fig. 6: Battery plate locked

Fig. 2: Push to unlock

Fig. 1: Removing battery cover

Sensor Replacement

Replace the sensor module at the fraction of the cost of a new tester. When the 

tester fails to calibrate or gives fluctuating readings in calibration standards, you 

need to change the electrode.
1.  With dry hands, grip the ring with sensor facing you. Twist the ring counter-

clockwise. Save the ring for later use.

2. Pull the old sensor module away from the tester.
3.  Align the four tabs on the new module so that they match the four slots on  

the tester.

4.  Gently push the module onto the slots to sit it in position. Push the smaller 

O-ring fully onto the new sensor module. Push the other O-ring over the  

module and thread it into place by firmly twisting clockwise.

Note: It is necessary that you recalibrate your tester prior to measurement after  

a sensor replacement.

Self-Diagnostic Messages

           

Batteries are weak and need replacement soon.

stable error

Appears when calibration is attempted but the reading is not yet stable. Wait for 

the reading to stabilize or manually confirm the calibration by pressing enter.

over range

The reading is above the measuring range of tester.

Sensor Maintenance

1.  Always keep the sensors electrodes clean. Rinse the electrodes with  

deionized water and wipe them dry with clean cloth before storing with its 

protective cap. For cup type electrodes, remove the white plastic cup and 

insert to thoroughly clean viscous solutions. Never scratch electrodes with 

a hard substance.

2.  For better performance, soak the electrode in alcohol for 10 to 15 minutes and 

rinse with de-ionized water before starting any measurement process. This is 

to remove dirt and oil stains on the electrode which may affect the accuracy 

of the measurements.

 

COND

      μs/cm 

        820 
      23.5 °C

      μs/cm 

       820 

      

23.5 °C

COND

Press Hold  
to release 

 

COND

      μs/cm 

       820 

      

23.5 °C 

 

hold

Press Hold  
to release 

Warranty

This instrument is supplied with a warranty against manufacturing defects for  

a period of one year from the date of purchase.

Return of Items

Authorization must be obtained from your distributor before returning items 

for any reason. When applying for authorization, please include information 

regarding the reason the item(s) are to be returned.
We reserve the right to make improvements in design, construction and 

appearance of products without notice. Prices are subject to change without 

notice.

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

 

COND

      μs/cm 

        820 
      23.5 °C

1065O100_MAN_35634-55,-65.indd   2

9/5/2017   3:07:29 PM

Reviews: