![Oakton CON 6 Instruction Manual Download Page 35](http://html2.mh-extra.com/html/oakton/con-6/con-6_instruction-manual_3142655035.webp)
Instruction Manual CON 6/TDS 6
31
2) Press
▲
or
▼
key until [
tdS
] appears on the LCD. Press
ENTER
key.
3) Press
▲
or
▼
key to select a value between 0.4 to 1.0.
4) Press
ENTER
key to select. The meter will take you back to the menu,
[
tdS
].
5) Press
▲
or
▼
key to move to the next menu or press
CAL
to exit to
measurement mode.
6.5 Temperature
Coefficient
The temperature coefficient is the amount of change in conductivity per degree
of temperature; it is expressed in percent per °C. Entering the exact
temperature coefficient of your solution lets you accurately compensate
temperature for almost any solution. You can adjust 0.0 to 3.0 % per °C.
Meter default is 2.1% per °C.
6) Enter the advanced setup as described in
Section 6.1.
7) Press
▲
or
▼
key until [
t.Co %
] appears on the
LCD. Press
ENTER
key.
8) Press
▲
or
▼
key to select a value between 0.0
to 3.0.
9) Press
ENTER
key to select. The meter will take
you back to the menu, [
t.Co %
].
Press
▲
or
▼
key to move to the next menu or press
CAL
to exit to measurement mode.
%
%
%
HOLD