Use the System Setup to customize operation of your 2700 series meter. Press  

SETUP

  from the 

measurement screen and then press 

ENTER

 when 

SYSTEM 

is highlighted to access these settings. 

The following settings can be customized for each model:

n

  

Stability

  To aid with measurement and calibration, your meter has a unique feature that provides an 

obvious visual indication when the reading becomes stable. If the Stability feature is 

enabled

,  

the digits on the primary reading will be faded when unstable, becoming solid when stable.  

If this feature is 

disabled

, the primary reading will always be solid.  

n

  

Stability Criteria

  The 2700 allows adjustment of the Stable indicator which relates to how fast and frequent it 

appears. There are three levels of adjustment; 

SLOW

MEDIUM

, & 

FAST

.

  To display a ‘Stable’ reading more quickly and more often, use “

FAST

” setting. 

MEDIUM

 or 

SLOW 

is recommended for most applications. When Auto Read is enabled, a stability criteria setting of 

SLOW

” is recommended to for best results.

 

SLOW: 

The Stable indicator will take longer to appear and will appear less frequently. Use this 

setting if you want the best results and don’t mind waiting longer to get them. 

 

MEDIUM:

 This is the factory default setting. It provides a balanced response which works best for 

most applications.  

 

FAST:

 The Stable indicator will appear quickest of any setting. If you find that the Stable indicator 

frequently appears and disappears, you may want to select a slower setting. This setting is not 

recommended when 

AUTO READ

 is enabled. It is also not suitable for most non-refillable 

electrodes which are generally slower to respond. 

n

  Navigation Of Tabs

SYSTEM SETUP & CONFIGURATION

PREV

NEXT

ESC

Scroll to previous  

menu option

ESCAPE – Use to  

cancel or exit out of  

current screen

Scroll to next  

menu option

SETUP

MEM IN

MEM OUT

REPORT

Enters Setup mode;  

use to change  

System settings or  

measurement parameters

Access stored values  

in memory

Inserts measured value  

into memory

Access Calibration  

Report Data

TEMP

SORT

CLR–C

NEXT–P

Access Temperature  

Calibration

Clears current  

calibration

Sorts stored data  

(memory out) by  

selected available  

option

Skip to Next Page / 

Next Parameter  

of menu

MEAS

CAL

MODE

PRINT

Resumes live  

measurement  

when Auto  

Read is flashing

Toggles available  

measurement modes

Begin Calibration  

Mode using current  

measured value

Sends data to  

printer or PC

STABILITY
STABILITY CRITERIA
AUTO READ
BACKLIGHT
DATALOG
PRINT SETUP
PRINT FORMAT
PRINT SET
DATE & TIME
PASSWORD
CLEAR DATALOG
FACTORY RESET
CONTRAST

3

2

Summary of Contents for Bench 2700 Series

Page 1: ...e five digit number to identify samples Measuring Range of displayed value 1 thru 5 Applicable to Conductivity TDS Salinity or Resistivity measurements only Automatic Temperature Compensation is activ...

Page 2: ...red values in memory Inserts measured value into memory Access Calibration Report Data TEMP SORT CLR C NEXT P Access Temperature Calibration Clears current calibration Sorts stored data memory out by...

Page 3: ...ing more quickly and more often use FAST setting MEDIUM or SLOW is recommended for most applications When Auto Read is enabled a stability criteria setting of SLOW is recommended to for best results S...

Page 4: ...single sample over time Note that MEM IN is not available during TIMED setting As data is collected the display will indicate how much memory has been used To view stored data use MEM OUT Note the mor...

Page 5: ...onsecutive times the meter returns to measurement mode In the event the password can not be recovered a password can be provided via a written request to Eutech Instruments Oakton Instruments The inst...

Page 6: ...he primary display is the un calibrated measured value The 2700 automatically selects the appropriate value from your buffer group in the secondary display This value will blink when the Stable indica...

Page 7: ...on Automatic Recognition 1 Press CAL if needed for calibration mode Rinse electrode then dip into your conductivity standard either 84 S 1413 S 12 88 mS or 111 8 mS 2 Use up down arrows to adjust cell...

Page 8: ...n mode rinse the probe do not dry 2 Dip the probe into sample of known DO concentration i e determined by titration or another instrument Press CAL 3 Use up down arrows to adjust the upper DO reading...

Reviews: