background image

57

56

DO Probe Care

The probe is a galvanic measuring element which
produces an output proportional to the oxygen 
present in the medium in which it is placed. 
The galvanic probe design lets you take measure-
ments immediately—without the typical 15
minute wait of other dissolved oxygen probes.

The probe consists of two parts:
• an upper part consisting of an anode, 

a cathode, and cable

• a lower part consisting of a membrane cap, 

membrane, and electrolyte solution. 

Under typical operating conditions, the probe
should last for several years.

Proper care and 

maintenance will help you receive the maximum
probe life and ensure more accurate readings.

Since any deposits on the membrane surface act
as a barrier to oxygen diffusing through the mem-
brane, the membrane must be cleaned at regular
intervals to assure maximum reliability.

After using the probe, rinse the probe in clean
water and wipe it with a soft cloth or paper to
avoid any hardening of deposits. If growth 
develops on the probe, use a disinfecting 
chemical to clean.

NOTE:

Although the membrane is strong and 
not easily damaged, wipe it gently while
cleaning it. If the membrane is damaged
or torn, the probe will no longer function.

There are no special probe storage requirements.

9. Probe Care and Maintenance

pH Electrode Care

Since your pH electrode is susceptible to dirt and contamination, clean it every one
to three months depending on the extent and condition of use.

NOTE:

for specialty electrode care, consult the instruction manual included with
your electrode.

pH electrode storage

For best results, always keep the pH bulb wet. Use the rubber cap filled with 
OAKTON electrode storage solution to store your electrode (see page 64-65 for
ordering information). Also, you can store in a pH 4 buffer with 1/100 part of 
saturated KCl. Other pH buffers are OK for storage, but NEVER use distilled 
water for storage.

After measuring

1.

Rinse the pH electrode and reference junction in de-ionized water.

2.

Store the electrode as recommended above in “pH electrode storage,” or as 
recommended by the manufacturer.

3.

Prior to next use, rinse the liquid junction with de-ionized water and tap dry—
never wipe electrode.

NOTE:

If this does not restore electrode to normal response, see “Reactivating the
pH electrode” section on page 57.

pH electrode cleaning

Salt deposits:

dissolve the deposits by immersing the electrode in tap water for ten

to fifteen minutes. Ten thoroughly rinse with distilled water.

Oil/grease film:

wash electrode pH bulb gently in some detergent and water. Rinse

electrode tip with distilled water or use a general purpose electrode cleaner (see
page 65 for ordering information).

Clogged reference junction:

heat a diluted KCl solution to 60 to 80°C. Place the

sensing part of the electrode into the heated solution for about 10 minutes. Allow the
electrode to cool in some unheated KCl solution.

Protein deposits:

prepare a 1% pepsin solution in 0.1 M of HCl. Set the electrode in

the solution for five to ten minutes. Rinse the electrode with distilled water.

9.1

Reactivating the pH electrode

If stored and cleaned properly, your pH electrode should be ready for immediate
use. However, a dehydrated bulb may cause sluggish response. To rehydrate the
bulb, immerse the electrode in a pH 4 buffer solution for 10 to 30 minutes. If this
fails, you probably need to replace the pH electrode. Never touch or rub glass bulb.
Contact builds up an electrostatic charge.

Anode

Membrane cap

(holds membrane

and electrolyte

solution)

Probe 
guard

Cathode

9.2

Summary of Contents for 35632-Series

Page 1: ...s Portable Waterproof pH Dissolved Oxygen Meter pH Dissolved Oxygen C F Meter pH DO 300 Series WATERPROOF MODE HOLD ENTER ON OFF CAL MEAS SETUP ATC READY MEAS C IP 67 Notes Write down the name and inf...

Page 2: ...efault settings 38 8 6 saturation mode Program P1 0 Dissolved oxygen parameters Pressure adjustment mode 39 8 7 saturation mode Program P2 0 Offset adjustment 40 8 8 saturation mode Program P3 0 Previ...

Page 3: ...lay shows the temperature of the reading The display also shows error messages keypad functions and program functions ATC READY MEAS CAL HOLD ERR F pH SETUP ON OFF C ppm mg l ppt 2 1 Primary display 5...

Page 4: ...pm calibration mode NOTE Temperature calibration is available from both pH and DO mg l ppm calibration modes see page 19 for directions In advanced set up mode Press CAL MEAS to return to main menu fr...

Page 5: ...connector Pull probe away from the meter CAUTION Do not pull on the probe cord or the probe wires might disconnect DO probe with 6 pin connector locking ring pH Electrode with BNC connector top view o...

Page 6: ...libration at pH 1 68 4 01 7 00 10 01 and 12 45 to ensure accuracy across the entire pH range of the meter The meter automatically recognizes and calibrates to these standard buffer values which makes...

Page 7: ...de and DO probe thoroughly with de ionized water or a rinse solution Do not wipe the probe this causes a build up of electrostatic charge on the glass surface 3 Dip both probes into the calibration bu...

Page 8: ...and returns to Measurement mode See figure ATC READY MEAS C READY CAL ATC READY MEAS C 4 5 A A B C B C Notes The reading in the primary display in step 3 must read at 50 or above for the calibration...

Page 9: ...reading in the primary display in step 3 must read at 10 or under for the calibra tion to work correctly Whenever an error occurs during calibration the ERR indicator appears in the lower left hand c...

Page 10: ...l ppm Measurement mode 2 Press the CAL MEAS key to enter pH or mg l ppm calibration mode The CAL indicator will appear above the primary display See figure 3 While in pH or mg l ppm calibration mode p...

Page 11: ...the READY indicator on or off in the SETUP program P5 1 for saturation measurements page 43 or program P4 1 page 51 for mg l ppm measurements Taking measurements with the auto endpoint feature selecte...

Page 12: ...of the LCD The ATC indicator appears in the lower right hand corner to indicate Automatic Temperature Compensation If the ATC indicator does not appear make sure DO probe is plugged into the meter Se...

Page 13: ...9 1 Switch the meter on Press the MODE key to select pH mode 2 Press the CAL MEAS key to enter pH calibration mode The CAL indicator will appear above the primary display 3 While in pH calibration mod...

Page 14: ...mmest light to level 8 brightest light 6 Press the ENTER key to confirm selection and to return to the subgroup menu Press the CAL MEAS key to return to measurement mode SETUP SETUP A A B B A This fea...

Page 15: ...E Probe data COF Configuration rSt Reset meter to factory default CAL Calibration data dPr Dissolved oxygen parameters pH Saturation mg l ppm LCd Backlit display rSt Reset meter to factory default Fro...

Page 16: ...ctions on page 45 SETUP rSt Reset to factory default Reset to factory default settings Instructions on page 46 SETUP From the mg l ppm measurement mode dPr Dissolved oxygen parameters Select Hg or Pa...

Page 17: ...recall previous calibration data which helps you know when to recalibrate your meter This is a view only mode 8 2 pH SETUP A A pH SETUP B pH SETUP C B C P2 1 Electrode offset From measurement mode 1...

Page 18: ...f ON switches the READY indicator on ON and HOLD together switches the auto endpoint feature on 6 Press the ENTER key to confirm selec tion and to proceed to step 4 of P3 2 Press the CAL MEAS key to r...

Page 19: ...ration mode after you have calibrated to your selected number of points P3 3 Selecting C or F This meter lets you select between C and F units for temperature readings From measurement mode 1 Press th...

Page 20: ...meters This sub group lets you adjust the barometric pressure Pressure adjustment mode Barometric pressure is vital to correct dissolved oxygen measurements You need to enter the correct barometric pr...

Page 21: ...ant to return to measure ment mode press CAL MEAS SETUP READY MEAS SETUP A A B B saturation mode P3 0 Previous calibration information This sub group shows you the previous calibration data along with...

Page 22: ...ey to enter SETUP mode 3 Press the and keys to scroll through subgroups until you view parameter P5 0 in the lower display See figure 4 Press the ENTER key to select parameter P5 1 rdY Ready See figur...

Page 23: ...e brightness of the backlit LCD Selecting a dimmer backlighting level helps conserve batteries 8 11 From measurement mode 1 Press the MODE key to select the saturation measurement mode 2 Press SETUP k...

Page 24: ...barometric pressure and salinity P1 1 Pressure adjustment mode Barometric pressure is vital to correct dissolved oxygen measurements You need to enter the correct barometric pressure of the area you...

Page 25: ...P1 0 If you want to return to measure ment mode press CAL MEAS SETUP ppt SETUP D D E E mg l ppm mode P2 0 Previous calibration information This sub group shows you the previous calibration data along...

Page 26: ...function The first program lets you select READY indicator on to indicate when the reading is stable READY indicator off for faster meter response Auto endpoint function on Select auto endpoint on to...

Page 27: ...Selecting C or F temperature readout This meter lets you select between C and F units for temperature readings From measurement mode 1 Press the MODE key to select mg l ppm mode 2 Press SETUP key to e...

Page 28: ...rogram lets you reset all parameters to factory default settings This clears all calibration data memory and any other setup functions you might have changed The LCD contrast setting will be reset to...

Page 29: ...solution to store your electrode see page 64 65 for ordering information Also you can store in a pH 4 buffer with 1 100 part of saturated KCl Other pH buffers are OK for storage but NEVER use distill...

Page 30: ...ert notches on installation tool into slots on membrane lock Then unscrew membrane lock from membrane cap B C B C 5 Using the installation tool unscrew and remove the membrane lock from the membrane c...

Page 31: ...ies not in place a Check that batteries are in no display place and making good contact b Batteries not in correct b Reinsert batteries with polarity and correct polarity c Weak batteries c Replace ba...

Page 32: ...00 0 0 0 to 50 0 C 20 00 mg l ppm 32 0 to 122 0 F Resolution 0 01 pH 0 01 mg l ppm 0 1 0 1 C 0 1 F Accuracy 0 01 pH 1 5 full scale 1 5 full scale 0 3 C 0 5 F Calibration up to 5 points one or two poin...

Page 33: ...dy have 35640 74 Replacement DO membranes 5 pack 35640 75 Replacement DO membranes 25 pack 35640 79 DO membrane installation tool Required for membrane installation Not required if you purchase replac...

Page 34: ...sate for these factors This meter automatically compensates for tempera ture readings It also allows you to enter a salinity correction factor and the baromet ric pressure to correct for this variabil...

Page 35: ...authorized returns NOTE We reserve the right to make improvements in design construction and appearance of products without notice 16 Appendix 3 Factory Default Settings Resetting the meter to factory...

Reviews: