5. Measurement mode

CAUTION: Do not submerge the probe in a liquid or powder or else damage 

to the probe may occur. Probe is for ambient air use only.

Each time you turn on the meter it will enter the measuring mode after it 
displays a test screen and the temperature sensor offset value. The measuring 
mode lets you measure temperature, % relative humidity, and dewpoint 
(dewpoint readings are available with model WD-35612-10 only).

Measuring humidity: 

1. 

Press ON

to power up meter. The LCD

test display appears for 1 second. 
Next, the temperature sensor offset
value will appear for 2 seconds.

2. 

The meter displays the percent relative
humidity and the temperature detected
by the probe. 

See figure

3. 

Move the probe around or cover the
end with your hand to see the tempera-
ture and relative humidity change. 
If this does not happen, and/or you
receive an error message, see page 21
for an explanation of error messages
and what actions you need to take.

1. 

Calibrate the meter for humidity if
needed (see pages 10-12 for directions).

2. 

While in humidity measurement mode,
press NEXT

. The meter will show “td”

to the lower left of the humidity display
to indicate dew point measurement
mode. The lower display will show the
dew point temperature. 

See figure

3. 

Press NEXT

to return to humidity 

measurement.

Measuring dewpoint 

(model WD-35612-10 only)

I

J

Fast humidity calibration:

Follow the directions below to calibrate humidity more quickly. Please note that
using this calibration method may result in an accuracy slightly less than specifica-
tion. If you want to ensure the highest possible accuracy, follow the “Automatic
Humidity Calibration” directions on pages 10-11. 

1. 

Place the meter, probe, and calibration
salt bottles in a place where the 
temperature is constant. 

2. 

Press OFF.

3. 

Press ON and PRG simultaneously
and hold until the screen displays
“CAL”. 

Release the keys. The screen

displays the temperature offset value.

See figure

CAUTION: 

Be careful when inserting 

the sensor probe into the calibration salt
bottles. Rough movement may damage 
the bottle.

4.

Gently open the cover of the 33% RH
calibration salt bottle and carefully
insert the probe into the bottle.
The probe will hit a stopper when 
it is fully inserted.

5.

Press RST twice. 

The meter will 

beep each time the key is pressed. 
The meter is now in low RH point 
calibration mode.

See figure

6. 

Wait until the low RH reading stabi-
lizes. Press PRG to save the calibration
value. The meter automatically moves
to high RH point calibration.

See figure

7. 

Gently open the cover of the 75% RH
calibration salt bottle and carefully
insert the probe into the bottle. 
The probe will hit a stopper when 
it is fully inserted.

8. 

Wait until the high RH reading stabi-
lizes. Press PRG to save the calibration
value and return to measurement mode.

MIN

%RH

G

G

MAX

%RH

H

H

F

F

Summary of Contents for 35612-series

Page 1: ...mer misuse Out of warranty products will be repaired on a charge basis 13 Return of items Authorization must be obtained from our Customer Service Department before returning items for any reason When...

Page 2: ...mode 17 7 2 Min Max memory mode 18 7 3 Hold mode 19 7 4 C F mode 19 8 Advanced Operations 20 9 Troubleshooting Error Messages 21 10 Specifications 22 11 Accessories 23 12 Warranty 24 13 Return of Item...

Page 3: ...e next digit for alarm setting Displays the next alarm setting value Selects the next digit for calibration PRG Switches from measuring program to alarm check mode Switches from alarm check to alarm s...

Page 4: ...The probe cable has a notched 6 pin connector to attach the probe to the meter NOTE Do not substitute other probes For a replacement probe see the Accessories section page 23 NOTE Keep connector dry...

Page 5: ...new temperature offset value Tnew by adding the results in steps a and b and subtracting the results in c Tnew To Tm Ts Tnew __________ Other Connections Three other connections are available Serial...

Page 6: ...es not meet your accuracy requirements If the values are only slightly different and you do not require the highest possible accuracy then calibration is not necessary Two point relative humidity cali...

Page 7: ...nt Measuring dewpoint model WD 35612 10 only I J Fast humidity calibration Follow the directions below to calibrate humidity more quickly Please note that using this calibration method may result in a...

Page 8: ...re value in C press Mn Mx to place or remove a sign in front of the old value ii Press NEXT if necessary one or more times until the digit you need to change is flashing See figure iii Press Mn Mx unt...

Page 9: ...s 2 through 4 Otherwise press RST to return to the measurement mode To cancel an alarm setting 1 In the alarm set mode select the alarm set point you wish to cancel i In the measurement mode press PRG...

Page 10: ...ays SA for saved The screen then displays RH See figure 6 Press OFF When the meter is turned on again it will default to the new temperature measuring mode X X Min Max memory mode Press Mn Mx to displ...

Page 11: ...Indicates Cause Solution Readout flashing OP Beeping Probe incorrectly Improper setup Check probe connected or faulty wire connection 0 0 Probe is Rough handling or Replace the 99 9 damaged defective...

Page 12: ...Adapter 110 VAC WD 35615 08 AC Adapter 220 VAC 10 Specifications Mode Relative Dew Point Temperature Humidity WD 35612 10 only Range 0 to 99 9 RH 51 0 to 49 0 C 0 to 50 0 C 60 0 to 120 0 F 32 0 to 12...

Reviews: