22
Device Settings
Turn the device on by pressing the Power button.
Press the Increase or Decrease
buttons to select the desired flow rate.
Turn the device off by pressing the Power button .
THE PROPER FLOW FATE IS PRESCRIBED BY YOUR PHYSICIAN.
DO NOT CHANGE THESE SETTINGS WITHOUT CONSULTING WITH YOUR PHYSICIAN.
PULSE DOSE SETTINGS MUST ACCOMMODATE THE INDIVIDUAL PATIENT’S LIFESTYLE INCLUDING REST,
TRAVEL AND EXERCISE.
OXYGEN DELIVERY SETTINGS OF THE DEVICE SHOULD BE PERIODICALLY REASSSESSED FOR EFFECTIVENESS
OF THE OXYGEN THERAPY.
Consult your healthcare provider if you are feeling unwell which may indicate too much or too little
oxygen. These side effects are not immediate or life threatening.
When the device is powered on it will enter a seven (7) minute warm up period to reach desired
performance. The warm up period is indicated on the display screen by a green ring that becomes a
green circle once warm up is complete.
The device will detect your breath and supply a measured pulse of oxygen or bolus. The breath detect
icon will flash on the control panel with each breath.
Flow Settings & Pulse Volumes (ml)
Breaths
per
Minute
Setting 1
Setting 2
Setting 3
Setting 4
Setting 5
15
8.0
16.0
24.0
32.0
40.0
20
8.0
16.0
24.0
32.0
40.0
25
6.4
12.8
19.2
25.6
32.0
30
5.3
10.7
16.0
21.3
26.7
35
4.6
9.1
13.7
18.3
22.9
40
4.0
8.0
12.0
16.0
20.0
The settings of the Oxlife Freedom might not correspond with continuous flow oxygen.
The settings of other models or brands of oxygen therapy equipment do not correspond with the
settings of the Oxlife Freedom.