8
3
TIDAL VOLUME:
0.5 litres
BREATHS PER MINUTE:
10
I:E RATIO:
1 : 2
AUTOMATIC FLOW RATE: 15.0 L/Min
MANUAL FLOW RATE: 15.0 L/Min
DELAY TO AUTOMATIC CYCLING RE-START
AFTER MANUAL BUTTON DEPRESSION: 20 SECONDS
POSITIVE PRESSURE BLEED FLOW: 2 L/Min
AUTO SHUTOFF
5 to 8 s
DEMAND BREATHING FLOWRATE: 0 – 120L/Min
INSPIRATORY RESISTANCE @ 60 L/Min: 0 to - 6 cmH
2
0
INPUT PRESSURE: 45 - 87 PSI
3.1 - 6 Bar
MAXIMUM AIRWAY PRESSURE: 60 cmH
2
0
58.8 mBar
OPERATING TEMPERATURE: -18
o
C to + 50
o
C
0
o
F to +122
o
F
STORAGE TEMPERATURE: - 40
o
C to + 60
o
C
- 40
o
F to +140
o
F
INPUT CONNECTION: 9/16” DISS
PATIENT CONNECTION: 15 / 22 mm
PATIENT VALVE DEAD SPACE: 8 ml
WEIGHT: 16 OZ / 0.45 Kg
SIZE: 140 x 63 x 73 MM
5.5 x 2.5 x 2.9IN.
CYLINDER DURATION:
@ 10 BPM / 0.5 Litres V
T
82 minutes
(Based on an Aluminum “D” size cylinder containing 415 Litres of air or
oxygen.)
Should the patient vomit into the mask during resuscitation the
following steps should be followed to clear the foreign material:
1.
Remove the mask from the patient’s face and clear any foreign
material from the patients airway. Depress the manual button
or allow the resuscitator to cycle automatically for a few breaths
to clear the mask and valve of foreign material.
2.
If depressing the manual button repeatedly or automatically
cycling the resuscitator does not clear the foreign material from the
patient valve,turn the selector to the OFF position, remove the
facemask and unscrew patient valve swivel housing (e) from the
resuscitator body being careful to ensure that the diaphragm is
retained (Fig.3).
3.
Shake out any foreign material from the resuscitator, diaphragm,
face mask and patient valve swivel housing.
4.
Reassemble the facemask, patient valve and diaphragm and
attach to the resuscitator.