O-Two CAREvent CA-G05 User Manual Download Page 3

 

 

TIDAL  VOLUME:   

 0.5   litres 

 

BREATHS  PER  MINUTE:                            

                10 

 

I:E RATIO: 

  1 : 2 

AUTOMATIC FLOW RATE:                                             15.0 L/Min 

MANUAL FLOW RATE:                                                  15.0 L/Min 

DELAY TO AUTOMATIC CYCLING RE-START 
AFTER  MANUAL  BUTTON  DEPRESSION:                    20  SECONDS 

POSITIVE  PRESSURE  BLEED  FLOW:                            2  L/Min  

AUTO SHUTOFF  

 

 

 

 

   5 to 8 s 

 

DEMAND  BREATHING  FLOWRATE:                              0 – 120L/Min  

INSPIRATORY RESISTANCE  @  60  L/Min:                     0 to - 6 cmH

2

 

INPUT  PRESSURE:                                                            45 - 87 PSI 

     3.1 - 6 Bar 

 

MAXIMUM  AIRWAY  PRESSURE:                                   60 cmH

2

0  

                                                                                           58.8 mBar 

 

OPERATING  TEMPERATURE:                                          -18

o

C to + 50

o

0

o

F to +122

o

 

STORAGE  TEMPERATURE:                                           - 40

o

C to + 60

o

                                                                                         - 40

o

F to +140

o

F   

INPUT CONNECTION:                                                       9/16” DISS 

PATIENT CONNECTION:                                                  15 / 22 mm 

PATIENT VALVE DEAD SPACE:                                          8 ml 

WEIGHT:                                                                           16 OZ / 0.45 Kg 

 

SIZE:                                                                                       140 x 63 x 73 MM 

5.5 x 2.5 x 2.9IN. 

CYLINDER DURATION: 

  @ 10 BPM / 0.5 Litres   V

T                                                                                

82  minutes 

 

(Based on an  Aluminum “D” size cylinder  containing 415 Litres of air or 

oxygen.) 

 

 

 

Should the patient vomit into the mask during resuscitation the 
following steps  should be followed to clear the foreign material: 

 

1.

 

Remove  the  mask  from  the  patient’s  face  and  clear  any  foreign 
material from  the  patients  airway.   Depress  the  manual  button 
or  allow  the resuscitator to cycle automatically for a few breaths 
to clear the mask and valve of foreign material. 
 

2.

 

If  depressing  the  manual  button  repeatedly  or  automatically 
cycling the resuscitator does not clear the foreign material from the 
patient  valve,turn  the  selector  to  the  OFF  position,  remove  the 
facemask  and  unscrew  patient valve swivel housing (e) from the 
resuscitator  body  being  careful  to  ensure  that  the  diaphragm  is 
retained (Fig.3). 
 

3.

 

Shake out any foreign material from the resuscitator, diaphragm, 
face mask and patient valve swivel housing. 
 

4.

 

Reassemble  the  facemask,  patient  valve  and  diaphragm  and 
attach to the resuscitator. 

 
 

   

Summary of Contents for CAREvent CA-G05

Page 1: ...dible Alarm 2 Ventilation Frequency and Tidal Volume 3 Manual Ventilation 4 Demand Valve Function and Automatic Circuit Shut Off 5 Bleed Flow The supply hose provided is attached to the oxygen inlet on the rear of the resuscitator and is tightened finger tight fig 1 The use of excessive force in tightening the supply hose may damage the seal and or thread The facemask is attached to the patient co...

Page 2: ...serve the patient s chest movements If there is any leak from around the mask or any obstruction in the patients airway blow off valve will operate reposition patients head and adjust mask and hand position to ensure a good airway and mask to face seal If the patient is intubated or if the patient is to be intubated following mask ventilation remove the face mask from the 22mm connector and attach...

Page 3: ...ze cylinder containing 415 Litres of air or oxygen Should the patient vomit into the mask during resuscitation the following steps should be followed to clear the foreign material 1 Remove the mask from the patient s face and clear any foreign material from the patients airway Depress the manual button or allow the resuscitator to cycle automatically for a few breaths to clear the mask and valve o...

Page 4: ...accessories for correct function should be undertaken by a responsible member of staff on a routine basis This check is to ensure that all of the accessories and resuscitator components are present the air or oxygen cylinder is full and that the resuscitator is in working order Regulator working pressure suction if equipped and ventilator limiting pressures should be checked at least every six mon...

Page 5: ...tilator support in pulmonary resuscitation in emergency medical and rescue environment where toxic gas may be present The ventilator is suitable for use in Pulmonary resuscitation during respiratory and or cardiac arrest Short term ventilatory support in the confined space rescue in IDLH Immedi ately Dangerous to Life and Health environment pre hospital Intra hospital inter hospital and air ambula...

Page 6: ...ssemble unit connect to an air or oxygen supply to check operation prior to packaging for emergency use 17MP9039 Reusable patient valve swivel housing 17MP1528 Reusable Patient Valve Diaphragm 01FG6500 Mask Retaining Insert 01FG6501 Mask Insert Extraction Tool 01FG6502 Head Harness System 17MP7010 Single Use PEEP Valve 01FV4314 Conversion Hose O2 Air 1 1 1 Introduction 1 1 2 Warranty Information 1...

Page 7: ... ...

Reviews: