O-Two 01FC Quick Start Manual Download Page 2

OPERATING INSTRUCTIONS:

A.   PRE-USE CHECK: 

1. 

Visually inspect the fl ow controller for any signs of grease or oil.  Do not use the device if grease 

or oil is suspected or found. 

2. 

Visually inspect the intended gas symbol which is marked on the fl ow controller label. It shall 

match the contents of the pressure source to be used.

3. 

Visually inspect the inlet and outlet connectors of the fl ow controller for damage and to ensure 

compatibility with the pressure source.

4. 

Connect the fl ow controller to a pressure source and turn the fl ow knob to “OFF” position.

5. 

If required, slowly and gradually turn on the pressure source.

6. 

Confi rm that the contents of the pressure source are suffi  cient for the intended use.

7. 

Apply a compatible leak test solution to fi ttings and threads of the inlet and outlet of the fl ow 

controller to check for leakage.

8. 

If leakage occurs between the fl ow controller and the pressure source, close the pressure source 

and tighten fi ttings as required to eliminate all external leaks. DO NOT over tighten threaded 

connections. Never tighten fi ttings when under pressure.

B.   OPERATION: 

1. 

Ensure that the Flow Controller knob is in the “OFF” position.

2. 

Connect the fl ow controller to a pressure source and tighten the connection if applicable.

3. 

Connect the fl ow controller to a patient through therapy tubing and mask.

4. 

If required, gradually open the pressure source.

5. 

Rotate the fl ow-selecting knob anti-clockwise to the required fl ow setting.    

C.   PREVENTATIVE MAINTENANCE: 

The fl ow controller can be cleaned with a clean, soft and damp cloth after each use. 

The user shall also test the fl ow controller (therapy fl ow and leak) periodically to ensure proper 

performance and safety. The frequency of such testing should be established according to usage, but 

it is recommended that it be performed at least every 6 months. 

O-Two Medical recommends that the fl ow controller be returned to the manufacturer or its authorized 

agents for evaluation for damage, contamination, wear and function every 5 years. The maximum 

service life for the fl ow controller is 15 years providing the above maintenance recommendations are 

followed.

THERAPY FLOW CONTROLLER FOR USE WITH MEDICAL GASES:

INTENDED 

MEDICAL GASES

O – Oxygen (O

2

)

A – Medical Air

INLET CONFIGURATION

1 - DISS Inlet- Male

2 - NIST Inlet 

3 - Schrader Female quick disconnect

4 - Hansen Female quick disconnect

5 - British Female quick disconnect

6 - Schrader male probe

7 – British Probe

8 – DISS Inlet- Female

9 – UNIFOR Probe

10- AFNOR Probe

11- DIN Probe

12- AGA Probe

13- DISS Inlet- Hand tight

14- DISS Inlet- Hand tight (US version)

PRESSURE OUTLETS

D – 9/16 DISS outlet

B – Therapy barbed outlet

01FC

THERAPY

06 – 0-  6 L/min (± 20%)

08 – 0-  8 L/min (± 20%)

15 – 0-15 L/min (± 20%)

25 – 0-25 L/min (± 20%)

*(± 30% for fl ows 1.5 l/min or less)

6. 

When medical gas is no longer required, turn off  the gas cylinder or disconnect the fl ow controller 

from hospital pipeline system and vent any trapped pressure inside the fl ow controller.

7. 

Turn the fl ow-selecting clockwise back to the “OFF” position and then remove the fl ow controller 

from the pressure source.

8. 

Store the fl ow controller in a clean and dry place.

 WARNING       

Setting the fl ow-selecting knob between adjacent settings may cause the output fl ow to cease!

Reviews: