7
Equinox® Advantage - User manual
1.6. Performance Specifi cations
14. Do not disassemble any part of the unit except where
described in this manual as any unauthorized disassembly will
invalidate the warranty.
15. Before use on a patient, the oxygen concentration of the
delivered gas should be checked at the setting intended for use.
16. Us federal law restricts this device to sale by or on the order
of a physician.
Minimum Flow rate
120 L/min
Exhalation Resistance
0 to 6 cmH
2
O @ 60 L/min
Inhalation Resistance
0 to -6 cmH
2
O @ 60 L/min
Oxygen Concentrations
(±5%V/V at Minimum Demand Flow 20 L/min)
25, 40, 50, 60, 70, 75, 90, 100%
Input Pressure
50 to 70 PSI (3.5 - 4.8 Bar)
Low Input Pressure Alarm
40 ± 1 PSI (2.7 Bar)
Device Cut-off Pressure
35 ± 2 PSI (2.4 Bar)
Demand Valve Triggering Pressure
-1.5 to -2.5 cmH
2
O
Oxygen Enrichment Flow
30 ± 5 L/min
Operating Temperature
41
0
F to 104
0
F (5
0
C to 40
0
C)
Storage Temperature
-40
0
F to 140
0
F (-40
0
C to 60
0
C)
Device Input Connections
Nitrous Oxide: CGA1040 DISS Oxygen: CGA1240 DISS
Supply Hose Connections
Refer to Section 6
Patient Connector
15/22 mm
Patient Circuit (Single use)
Dual limb circuit with one-way valves and 19 mm
Scavenging Port Connection on the exhalation limb
Patient Valve Dead Space
32 ml
Weight
4.4 lb. (2.0 Kg)
Dimensions W x D x H
9.27”x7.6”x4.4” (236x194x112mm)