4
Do not use any other input power cord other than the one provided with the unit. If the
input power cord is lost or missing contact the seller to procure a replacement cord.
The outlet must consist of a Line, Neutral and Ground (DO NOT plug the power conditioner into
an ungrounded outlet or use any kind of ground removing plug (commonly called “cheater
plugs”) to power the conditioner. If the power conditioner is equipped with a power on circuit
breaker switch turn the switch to the
ON
position. The blue power
ON
indicator (light) on the
front of the unit will illuminate. Once the power conditioner is plugged in and turned on, insure
the “Site Wiring Fault” LED (120 VAC North American models only) on the rear of the unit is not
flashing. If this LED is flashing contact a qualified electrician immediately to insure proper wiring
of the mains outlet. The power conditioner must not be plugged into this outlet until the wiring
problem is corrected. If the “wiring fault” LED is not ON or flashing, you can turn on the load
equipment/electronic devices. If you have any questions in this regard, contact the seller.
Note: If the power conditioner is mounted to a wall the installation can be in the horizontal position
facing left or right. If the power conditioner is mounted vertically the unit must be installed with front
plastic panel facing the ground.
All
NXT POWER
Conditioners are equipped with overload protection and this protection will turn
off the power conditioner if overloaded. If the power conditioner turns off due to an overload,
remove some of the load equipment/electronic devices from the conditioner and turn the power
conditioner on again.
Overload Protection device cannot be replaced.
These power conditioners are not rated or intended for use in the presence of flammable gas or
anesthetics as there will be a risk of a of an explosion.
All power conditioner
s
use a basic type insulation and a grounded power network must
be used.
All power conditioners listed below are considered Class 1.
All power conditioners listed below are considered Type B.
There are no user serviceable parts inside the power conditioner, contact the seller for
service or repair.
All power conditioners use hazardous voltage.
All power conditioners should only be used in ordinary locations.
All Power conditioners can be operated in continuous duty cycle.
All Power conditioners enclosure are IP40 rated.
CAUTION
CAUTION
DANGER