Nuvo Video Display Systems – User Manual 

Nuvo Video Display System Specifications 

 

 

Dimensions for Product V1360 Nuvo Video Display Systems 

 

Feature Dimension 

Counterbalance arm outer clevis to finish floor 

78" (198 cm) 

Suspension arm inner hub center line to outer hub 
center line 

26.75" (68 cm) 

Suspension arm inner hub center line to outer most 
point on monitor frame 

67" (170 cm) 

Monitor size 

19" (48 cm) 

Minimum ceiling height 

Maintaining a 78" (198 cm) minimum 
counterbalance arm outer clevis to finished floor 

 

No light 

8' 0" (244 cm) 

 

Single light 

8' 7" (262 cm) 

 

Dual light 

9' 0" (274 cm) 

 
 

Monitor Specifications 

 

Dimension 

Feature 

19" Vector3 Monitor 

19" Radiance/EndoVue Monitor 

Screen size 

19" (48 cm) 

19" (48 cm) 

Brightness 250 

cd/m

2

340 cd/m

2

Resolution 

1024 x 768 Extended 
Graphics Array (XGA) 

1280 x 1024 (SXGA) 

Dot Pitch 

0.294 mm 

0.294 mm 

Viewing angle 

170° horizontal 
170° vertical 

170° horizontal 
170° vertical 

Contrast ratio 

600:1 

650:1 

Input sources 

Composite 
Y/C (S-Video) 
RGBS 
RGBHV 

Composite 
Y/C (S-Video) 
RGBS 
RGBHV 

Storage height (head clearance) 

78" (198 cm) 

78" (198 cm) 

Low position (floor clearance) 

49" (124 cm) 

48" (124 cm) 

 

 

 

 

NOTE:

 

The above monitor specifications are applicable to the National Display Systems medical grade 
monitors offered by Nuvo Surgical.  Alternative models may be provided by others or, for special 
configurations, by Nuvo Surgical. 

 
 
 
 

M2Z00167 Page 

16 

Summary of Contents for V1360A

Page 1: ...uvo Video Display Systems User Manual Nuvo Surgical Video Display System Model V1360A USER MANUAL For Parts or Technical Assistance USA and CANADA 800 663 1152 INTERNATIONAL 814 899 422 M2Z00167 Page...

Page 2: ...o Signal for Display at a Monitor Using the Monitor Handle 10 To Remove the Sterile Handle and Reinstall 11 To Adjust the Monitor Image 11 Maintenance 12 Cleaning 12 Sterilization 12 Sterilizable Hand...

Page 3: ...se manuals Where applicable this manual is intended to supplement the Nuvo Surgical Light User Manual S2Z00125 WARNING It is recommended that a complete back up display system be available at all time...

Page 4: ...ingle or dual monitor mounts Monitor Features 19 flat screen medical grade Liquid Crystal Display Up to four input channels with optional switching system Wall mounted input output selector with optio...

Page 5: ...e counterbalance arm The monitor pivots 300 inside the arm It can be moved completely out of the way when not in use without affecting the light The combination of these individual elements allows med...

Page 6: ...pressing a button on the wall controls or on the sterilizable handle at the monitor To Turn the Monitor On Off Turn the monitor system on by using the on off switch located on the back of the monitor...

Page 7: ...deo cable supplied by others into a Video Input Panel having the same video format Composite Video S Video RGB Video NOTE The composite video cable must be twisted clockwise to ensure that the cable i...

Page 8: ...is compatible with a the signal to be recorded or displayed and b the device to which it will be connected NOTE Output Panels are labeled Output A and Output B The signal delivered to Output A duplica...

Page 9: ...rol where the bottom panel is dedicated to video switching Press the associated button to the left of the panel for Monitor A to the right for Monitor B repeatedly until the desired image appears on t...

Page 10: ...the same function as the selection button at the optional wall control panel NOTE This operation should only be performed by a surgeon or other sterile staff member Operation by a non sterile member o...

Page 11: ...e Sterilizable Handle must be sterile during surgical procedures Therefore suitable precautions should be taken during installation to ensure that sterility is not compromised Either use a sterilized...

Page 12: ...fully and use them in accordance with the manufacturer s instructions Take care to avoid the use of cleaning materials that contain high concentrations of alcohol or chlorine as these may lead to prem...

Page 13: ...before use NOTE Service Life of Sterilizable Handle Frequent sterilization causes natural degradation of the replaceable handle If signs of material fatigue such as cracking or discoloration occur th...

Page 14: ...ature are functioning as originally designed Pay particular attention to The condition of the video input and output connectors System drift adjustment See the picture below for Allen wrench sizes Con...

Page 15: ...sing the wall control Incorrect monitor selection Select correct inputs for A and B monitors Incorrect input selection Select correct inputs for A and B monitors Monitor not in sync with source Turn m...

Page 16: ...Monitor Specifications Dimension Feature 19 Vector3 Monitor 19 Radiance EndoVue Monitor Screen size 19 48 cm 19 48 cm Brightness 250 cd m2 340 cd m2 Resolution 1024 x 768 Extended Graphics Array XGA...

Page 17: ...mum Current Rated Frequency 100 240 V 2 0A 50 60 Hz Ceiling Structure Vertical Load Characteristics Description Specification Monitor suspension with two monitors 175 lb 79 kg Monitor and single light...