Ethernet  settings  should  be  adjusted  by  technical  service  staff  for  the  first  usage. 
Otherwise, this function does not work. 

 

“E-Mails” is the page which e-mail addresses will be written on. 

 

 “Login” is the adjustments for the e-mail addresses. 

 

“E-Mails”  is  selected  on  the  e-mail  page  by  the 
value  increase  (F1)  and  decrease  (F2)  keys.  The 
page  on  the  left  appears  when  enter  key  (F4)  is 
pushed. The e-mail addresses can be typed in this 
page and e-mail is sent to these addresses in case 
of any failure.  

 

Each  character  of  e-mail  addresses  is  typed  one  by  one.  The  first  character  is  entered  by 
pushing value increase (F1) and decrease keys (F2) and tab key (F3) is pushed to enter the 
second  character.  The  second  character  is  written  by  pushing  value  increase  (F1)  and 
decrease keys (F2) and tab key (F3) is pushed to enter the next character. After all characters 
are  written  completely in  this  way,  push  the enter  key  (F4)  to  enter the mail  server. After 
mail server is written in the same way, push the enter key (F4) to write next e-mail address. 
After third email address is written, push enter key (F4) to enter other e-mail addresses. 

 

After all e-mail addresses are written as mentioned above, push enter key (F4) and return to 
main page. 

 

Tab key (F3) is used for changeover from one character of address to another. 

 

 

“Login”  is  selected  on  the  e-mail  page  by  the 
value  increase  (F1)  and  decrease  (F2)  keys.  The 
page  on  the  left  appears  when  enter  key  (F4)  is 
pushed.  Each  character  of  e-mail  addresses  is 
written one by one. The first character is entered 
by pushing value increase (F1) and decrease keys 
(F2) and tab key (F3) is pushed to enter the next 
character.  The  second  character  is  entered  by 
pushing value increase (F1) and decrease keys  

(F2) and tab key (F3) is pushed to enter the next character. After all characters of user’s name 
are entered by this way, push enter key (F4) to enter the mail server of the e-mail address. After 
mail server is entered in the same way, push enter key (F4) to pass the “password” submenu. 
Enter the password and push enter key (F4).  
The  port  provided  by  internet  server  is  entered  on  the  part  of  “Port”.  After  port  is  entered, 
enter  key  (F4)  is  pushed  to  return the main  menu.  Tab  key  (F3)  is  used  for  changes from  one 
character of address to another. 

Summary of Contents for KN 120

Page 1: ...N VE SANAY MALZEMELER MALAT VE T CARET A KN 72 KN 120 KN 294 KN 504 BLOOD BANK REFRIGERATORS USER S MANUAL 1984 Z14 K 25 253 Rev No 07 Rev Date 01 2016...

Page 2: ...damaged instrument 4 do not operate the instrument with a damaged cable 5 do not move instrument during operation In case of a problem contact your N ve agent for an authorized service or maintenance...

Page 3: ...nditions use of the product in non standard environmental conditions including but not limited to failure to meet requirements of ambient temperature lubrication humidity or magnetic field influences...

Page 4: ...ROGRAMMING 13 6 1 1 OPERATING MODE 13 6 1 2 SETTINGS 14 6 1 3 MEMORY 15 6 1 3 1 INTERNAL MEMORY 15 6 1 3 2 EXTERNAL MEMORY USB STICK 17 6 1 4 SMS 17 6 1 5 E MAIL 18 6 1 6 SET TEMPERATURE 20 6 1 7 GRAP...

Page 5: ...nt unauthorized access KN series have audible and visual alarm system alarm systems for high and low temperature conditions power failure temperature sensor failure and other faults associated with th...

Page 6: ...att Door Insulated steel frame with triple glass window and key lock Drawers Stainless steel drawers Number Of Drawers 3 pcs 5 pcs 6 pcs 12 pcs Control System N SmArt Programmable microprocessor contr...

Page 7: ...ginal spare parts and original accessories supplied by N ve should be used Correctly grounded power supply should be used There should not be any material in the chamber that may damage the device The...

Page 8: ...72 KN 120 KN 294 Fuses 2 x 10 A KN 504 Fuses 2 x 13 A 5 INSTALLATION 5 1 ENVIRONMENTAL CONDITIONS The instrument is designed to operate safely under the following conditions Indoor use only Ambient t...

Page 9: ...trument Always plug in the instrument to correctly grounded sockets A supply fitted with a circuit breaker should be used for protection against indirect contact in case of isolation fault 5 5 POSITIO...

Page 10: ...10 5 6 GENERAL PRESENTATION Figure 1 CONTROL PANEL EXTERNAL PROBE ACCESS PORT COMMUNICATION UNIT POWER SWITCH KEY LOCK DRAWER GLASS DOOR DOOR HANDLE CHART RECORDER DOOR LOCK...

Page 11: ...4 The functions of F1 F2 F3 and F4 buttons depend on the meaning of the corresponding symbol appearing on the display The following table shows the meaning of these symbols FUSES POWER CORD NC CONTACT...

Page 12: ...e alarm is activated in case of error This symbol denotes backspace You can return the previous page or exit from the page by pushing the button corresponding to this symbol This symbol denotes the va...

Page 13: ...el See Section 5 7 6 OPERATING PRINCIPLES 6 1 PROGRAMMING For temperature homogeneity never take out the plexiglas dividers inside the drawers and do not overload the drawers 6 1 1 OPERATING MODE When...

Page 14: ...key F4 is pushed Time is shown in the format of hour minute and the cursor is on hour part when Date Time Settings page comes to the screen Hour is adjusted by pushing the value increase F1 and decre...

Page 15: ...fter old password value is entered by pushing the value increase and decrease keys push the enter key New password value can be entered in the New Password and Retype New sections If you want to exit...

Page 16: ...mn since temperature cannot be read by the sensor History GRAPH submenu is selected on the memory page by the value increase F1 and decrease F2 keys Select Record Date page appears when enter key F4 i...

Page 17: ...e display returns the main page If you want to exit this page without starting to record use backspace key F3 You can remove USB stick when you want to finish the record process Copy Failure History A...

Page 18: ...y F3 is pushed to enter the second digit The second number is entered pushing value increase F1 and decrease keys F2 and tab key F3 is pushed to enter the next digit After all digits of the phone numb...

Page 19: ...address After third email address is written push enter key F4 to enter other e mail addresses After all e mail addresses are written as mentioned above push enter key F4 and return to main page Tab...

Page 20: ...the condenser surface with a vacuum cleaner every 3 months 7 2 CLEANING Clean the refrigerator at room temperature after disconnecting the power cable Remove the shelves from their shelf carriers Cle...

Page 21: ...nterruption push the key corresponding to question mark F3 in order to see the time when power interruption occurs Err1 It appears when temperature sensor is defective Err2 It appears when there is a...

Page 22: ...e Figure 7 SIM card is provided by the user The cost of SIM card and SMS differs according to the GSM providers and all charges will be covered by the user SIM card which will be used for GSM module s...

Page 23: ...ppear when SMS submenu on the SMS page is selected as off 10 2 N veWarn REMOTE ALARM SYSTEM KN series blood bank refrigerators have remote alarm feature This feature can be used only if an optional re...

Page 24: ...it and take it out Put on the battery supplied with device to the hole at the back of the unit by paying attention to and poles Please use always leak protected high quality batteries Put the battery...

Page 25: ...25 11 ELECTRICAL CIRCUIT DIAGRAM 11 1 KN 72 AND KN 120 ELECTRICAL CIRCUIT DIAGRAM...

Page 26: ...26 11 2 KN 294 ELECTRICAL CIRCUIT DIAGRAM...

Page 27: ...27 11 3 KN 504 ELECTRICAL CIRCUIT DIAGRAM...

Reviews: