NUVE BM 15 Instruction Manual Download Page 5

 

1.

 

INTRODUCTION 

1.1.

 

USE AND FUNCTION 

BM  15  and  BM  30  Water  Baths  are  designed  to  be  used  in  bacterial  culturing,  cell  culture, 
coagulation  tests,  coliform  determination,  enzyme  reactions  and  other  general  and  special 
applications. 

The BM Series provide temperatures between 5°C above the ambient temperature and 99.9 °C 
and keeps the temperature stable within the given tolerances. 

BM  Series  have  circulation  function.  The  circulation  pump  and  etched  foil  heaters  which  are 
placed at the outer sides of the tank provide absolutely homogenous temperature distribution 
in the tank. They offer excellent temperature control of liquid producing a uniform and stable 
heating environment for the applications. 

The baths has a microprocessor control system. The heating is based on PID controlled system. 
There are two digital displays for heating function. One of them is temperature display the other 
is timer which can be programmed up to 99.9 hours and including HOLD position. 

All  components  which  are  in  contact  with the  liquid  are  made  of high  grade  stainless  steel  to 
resist corrosion. The tank is insulated with glass wool for operating homogenous temperature 
distribution. 

The BM Series Water Baths are manufactured according to the following standards:  
EN 61010-1, EN 61000-6-3, EN 50419. 
 
This device is in compliance with WEEE Regulation. 
 
If mentioned warnings in this manual are not considered, NÜVE will not be responsible from 
their results. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for BM 15

Page 1: ...NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A Ş BM 15 BM 30 STIRRED WATER BATHS INSTRUCTION MANUAL Z14 K 25 098 Rev No 03 Rev Tarihi 01 2016 ...

Page 2: ...erate damaged instrument do not operate the instrument with a damaged cable do not move the instrument during operation In case of a problem contact your Nüve agent for an authorized service or maintenance The validity of the guarantee is subject to compliance with the instructions and precautions described in this manual Nüve reserves the right to improve or change the design of its products with...

Page 3: ...onditions use of the product in non standard environmental conditions including but not limited to failure to meet requirements of ambient temperature lubrication humidity or magnetic field influences from the defects in maintenance negligence bad functioning of auxiliary equipment in the case of force majeure or accident and incorrect power supply 5 Any injury loss or damage caused due to a failu...

Page 4: ...ION 8 5 3 UNPACKING 9 5 4 MAINS SUPPLY 9 5 5 POSITIONING 9 5 6 GENERAL PRESENTATION 10 5 7 CONTROL PANEL 10 5 8 PRIOR TO OPERATION 11 5 8 1 FILLING LIQUID 11 6 OPERATING PRINCIPLES 12 6 1 OPERATING THE UNIT 12 6 2 PROGRAMMING 12 6 3 COMPLETION OF THE OPERATION 13 7 PERIODIC MAINTENANCE AND CLEANING 14 7 1 PERIODIC MAINTENANCE 14 7 2 CLEANING 14 8 DISPOSAL MANAGEMENT CONCEPT 14 9 TROUBLESHOOTING 14...

Page 5: ...of liquid producing a uniform and stable heating environment for the applications The baths has a microprocessor control system The heating is based on PID controlled system There are two digital displays for heating function One of them is temperature display the other is timer which can be programmed up to 99 9 hours and including HOLD position All components which are in contact with the liquid...

Page 6: ...nute 16 Internal Material Stainless Steel External Material Epoxy polyester powder coated stainless steel Internal Dimensions WxDxH mm 300x240x200 300x503x195 External Dimensions WxDxH mm 332x280x375 332x540x375 Packing Dimensions WxDxH mm 430x535x420 430x680x495 Net Packed Weight kg 8 11 11 14 At ambient 22 C and closed lid 2 2 OPTIONAL ACCESSORIES A 08 021 Tube rack 27xØ18 mm wide 70 mm A 08 049...

Page 7: ... be used by authorized and trained staff after the instruction manual has been read carefully Only authorized technical staff can handle the product in case of a failure Only original spare parts and original accessories supplied by Nüve should be used Correctly grounded power supply should be used Liquids should not be heated in sealed containers The samples which may liquefy and expand should no...

Page 8: ...e Earthed Wall Socket BM 15 Fuses 2 x 5A BM 30 Fuses 2 x 9A 5 INSTALLATION 5 1 ENVIRONMENTAL CONDITIONS The water bath is designed to operate safely under the following conditions Indoor use only Ambient temperature 5 C to 40 C Maximum relative humidity for temperature up to 22 C 80 Maximum altitude 2000 m Temperature for maximum performance 15 C 25 C 5 2 HANDLING AND TRANSPORTATION All handling a...

Page 9: ... at the back of the steam sterilizer Always plug in the instrument to correctly grounded sockets A supply fitted with a circuit breaker should be used for protection against indirect contact in case of an isolation fault 5 5 POSITIONING Check that no damage occurred during transportation Check that the positioning is suitable for the users Check that the water bath is stable on its four feet Check...

Page 10: ...10 5 6 GENERAL PRESENTATION Figure 1 Figure 2 5 7 CONTROL PANEL Figure 3 CONTROL PANEL STIRRED TANK 3 9 1 2 10 4 5 6 7 8 ...

Page 11: ...me 1 minute 99 9 hours and Hold position 8 Time Value Increase Decrease Keys These keys are used to increase or decrease the values on the displays 9 Heat Led This led flashes during heating process 10 Alarm Led It turns on when the program ends and if any failure occurs during the operation 5 8 PRIOR TO OPERATION 5 8 1 FILLING LIQUID Fill the tank with distilled water or liquid and put the sample...

Page 12: ...ons of command and display panel see Section 5 7 Set the values and start the operation Please check of the liquid level in the tank during long operations and refill to the may line if necessary 6 2 PROGRAMMING Push the temperature set key Set the temperature by pushing the value increase decrease keys on the temperature adjustment side Push the temperature set key again to save the temperature v...

Page 13: ...ey again to save the settings Push the start stop key to start the program The set time start to count after the device reaches the set temperature 6 3 COMPLETION OF THE OPERATION Ensure that the program is over If there is no adjusted program you can terminate the heating process by pushing start stop button Take the samples out at the end of the operation Take precautions while handling the samp...

Page 14: ...ividual component All parts which may comprise potentially infectious materials have to be disinfected by suitable validated procedures i e autoclaving chemical treatment prior to disposal 9 TROUBLESHOOTING If the device fails to operate please check the followings The power switch is on The fuse is not blown The plug is plugged in properly The plug is not defective The installation of the plug is...

Page 15: ...15 10 ELECTRICAL CIRCUIT DIAGRAM ...