6.0. Powering Off
At the home or gas reading screen, hold down the On/Off button for a couple of seconds.
The Analyzer will display “OFF” and then go blank.
7.0. Factory Reset
In case it is necessary to reset the Pro CO
2
Analyzer to the factory settings, power on the
Analyzer pressing at the same time for more than one second the “On/Off” and “Adjust”
buttons. On the display will appear “res” and the instrument will go to the reading page.
In case of reset, the instrument will delete all the alarms settings and the full scale value.
Before using again the instrument, it may be necessary to program again the alarm
values and the full scale value. All the analysis concentration shown on the display could
be wrong. Wrong Carbon Dioxide analysis may lead to death. You should calibrate your
analyzer immediately after a reset before using.
8.0. Maintenance
8.1. Analyzer Care
Analyzers immersed in liquid or stored in wet environments may not operate properly.
This may result in incorrect readings. Incorrect gas analysis may result in personal
injury or death.
Protect the analyzer from excessive shock and impact. Excessive shock and impact may
result in incorrect readings. Incorrect gas analysis may result in personal injury or
death.
Protect the analyzer from exposure to hyperbaric environments. Exposure to hyperbaric
environments may result in incorrect readings. Incorrect gas analysis may result in
personal injury or death.