Nutricia Flocare Infinity+ Short Manual Download Page 2

Flocare Infi nity+

 

Problem solving

instructions

Error messages and instructions

Please refer to detailed Instructions for Use for additional information

171101 / BP&MC 11090802

>

  Press                   to switch off the pump  

 Insert the Flocare

®

 Infi nity

 set into the pump following 

instructions for use  

>

  Restart the pump.

>

  The pump has been untouched in hold mode for over 3 minutes  

>

  Extend the hold mode by a further 

3 minutes by pressing                             

>

  Program the pump and press                            to start the pump.

>

  The programmed dose / volume has been administered  

>

  Switch off the pump by pressing                   

for 2 seconds  

 Clear the memory by pressing                   .

>

  No fl ow rate is installed  

 Make sure the correct fl ow rate is programmed.

>

  The door is not correctly closed  

 Make sure the door is properly closed prior to starting a feeding 

program.

>

  Check if the set is inserted correctly and check the air sensor  

 Remove the air in the set by using                  

                       key.

>

  Connect the adaptor to the pump and mains and charge the pump for approx. 6 hours. During charging 

the pump can be used.

>

  Battery failure  

>

  Please refer the pump for service.

>

  Press                          

>

  Check the feeding set (IN) or feeding tube (OUT) for blockages  

>

  Reinsert 

the set and start the pump, if necessary clean the sensors.

>

  Electronic error  

>

  Switch the pump off, make sure the pump door is closed and turn the pump back 

on  

>

  If the error persists, please refer the pump for service.

>

  The wall outlet or the adaptor does not work  

>

  Connect to another wall outlet  

>

  If there is still no 

symbol, request replacement of the adaptor.

NO SET

PUSH STRT

END OF DOSE

PROG

DOOR

AIR

BATT

BATT  E + F

OCC IN

>

  The programming feature is blocked in the set-up mode of the pump  

>

  Refer to instruction booklet 

to de-block.

LOCK

ER01 - ER99 

NO PLUG SYMBOL

OCC OUT

Problem

Instructions

(while the pump is 

connected to the mains)

clr

fi ll set

start/stop

start/stop

start/stop

>

  Stop the pump by pressing the                          

>

  Let it run briefl y by repressing the

key  

>

  Stop the pump again with                         key, ensuring there has been no OCC OUT alarm  

>

  

Remove the set from the pump and reinsert it into the pump  

>

  Restart the pump by pressing  

                       .

OCC OUT

(repeated)

start/stop

start/stop

start/stop

start/stop

Reviews: