Flocare Infi nity+
Problem solving
instructions
Error messages and instructions
Please refer to detailed Instructions for Use for additional information
171101 / BP&MC 11090802
>
Press to switch off the pump
>
Insert the Flocare
®
Infi nity
™
set into the pump following
instructions for use
>
Restart the pump.
>
The pump has been untouched in hold mode for over 3 minutes
>
Extend the hold mode by a further
3 minutes by pressing
>
Program the pump and press to start the pump.
>
The programmed dose / volume has been administered
>
Switch off the pump by pressing
for 2 seconds
>
Clear the memory by pressing .
>
No fl ow rate is installed
>
Make sure the correct fl ow rate is programmed.
>
The door is not correctly closed
>
Make sure the door is properly closed prior to starting a feeding
program.
>
Check if the set is inserted correctly and check the air sensor
>
Remove the air in the set by using
key.
>
Connect the adaptor to the pump and mains and charge the pump for approx. 6 hours. During charging
the pump can be used.
>
Battery failure
>
Please refer the pump for service.
>
Press
>
Check the feeding set (IN) or feeding tube (OUT) for blockages
>
Reinsert
the set and start the pump, if necessary clean the sensors.
>
Electronic error
>
Switch the pump off, make sure the pump door is closed and turn the pump back
on
>
If the error persists, please refer the pump for service.
>
The wall outlet or the adaptor does not work
>
Connect to another wall outlet
>
If there is still no
symbol, request replacement of the adaptor.
NO SET
PUSH STRT
END OF DOSE
PROG
DOOR
AIR
BATT
BATT E + F
OCC IN
>
The programming feature is blocked in the set-up mode of the pump
>
Refer to instruction booklet
to de-block.
LOCK
ER01 - ER99
NO PLUG SYMBOL
OCC OUT
Problem
Instructions
(while the pump is
connected to the mains)
clr
fi ll set
start/stop
start/stop
start/stop
>
Stop the pump by pressing the
>
Let it run briefl y by repressing the
key
>
Stop the pump again with key, ensuring there has been no OCC OUT alarm
>
Remove the set from the pump and reinsert it into the pump
>
Restart the pump by pressing
.
OCC OUT
(repeated)
start/stop
start/stop
start/stop
start/stop