background image

Setting Up the PicoDigitizer250 

The following procedure will guide you through the PicoDigitizer250 setup to run the Record/Playback example.  

Requirements 

 

PicoDigitizer box content 

 

A PC with Windows 7 64-bits and 10 GB of RAM. 

 

Visual Studio 2008 SP1 

 

An Ethernet cable 

 

A 1-MHz signal generator 

 

A 1-GHz oscilloscope 

Procedure 

1.

 

Install  the  ADP  software  tools  on  the  Windows  PC.  Follow  the  instructions  available  in  the 

ADP  uTCA  edition  -  Software 

installation guide for windows7.pdf 

document in the 

Install documentation

 folder of the installation DVD. 

2.

 

Connect the PC to the PicoDigitizer250 with the Ethernet cable. Use the Ethernet port 1 on the PicoDigitizer250 backplane. 

3.

 

Connect  an  MMCX  to  SMA  cable  between  the  ADAC250  ADC  A  connector  and  the  signal  generator.  Configure  the  signal 
generator to output a 0 dBm signal at 1MHz. 

4.

 

Connect an MMCX to SMA cable between the ADAC250 DAC A connector and the oscilloscope. 

5.

 

Power up the PicoDigitizer250 by plugging in the power supply and by clicking once on the backplane reset button.  

6.

 

The Perseus will boot and configure itself with the Record example bitstream previously written in Flash memory.  

7.

 

The PicoDigitizer Perseus is configured to use a static IP upon startup. The Perseus default IP address is 192.168.0.101 

8.

 

Start Microsoft Visual Studio. 

9.

 

On the 

File 

menu, point to 

Open

 and click 

Project/Solution

10.

 

Browse to the

 \perseus6010_adac250_record_playback\host\prj_win\ 

folder and select the 

perseus6010_adac250_playback_record.sln 

file. 

The host software project opens and you see the hierarchy of the project on the 

Solution Explorer 

tab. If you do not see the 

Solution Explorer 

tab, on the 

View 

menu, click 

Solution Explorer

11.

 

Select the build configuration Release x64. 

12.

 

On the

 Build 

menu, click 

Build Solution

13.

 

Double-click 

Launch_adac250_record_playback_demo.bat.

 

The test starts automatically. 

14.

 

When prompted, choose the test you wish to execute. The results for the Record (15) and the Playback(16) tests are shown 
below. 

15.

 

Record Test Result 

 

The following window shows a successful PicoDigitizer250 record test: 

 