nubeca RT-1242 Operation Instructions Download Page 3

SYMBOL INFORMATION

MESSAGE

WHAT IT MEANS

WHAT TO DO

Appear when ambient 

temperature is outside of 

the operating temperature 

range.

Only operate the thermometer 

in ambient temperature range of 

10°C to 40°C (50°F to 104°F).

Surface temperature 

measured outside of the 

measurement range (0.0 to 

100.0 °C). 

Review the instructions and 

re-start the measurement 

procedure.

Low or no power. 

Replace with new batteries.

Measured temperature 

below the measurement 

range (below 32.0 °C). 

Review instructions and repeat 

measurement. Make sure the 

distance from forehead is not 

greater than 7cm (2”), and that 

the forehead is clear.

Measured temperature 

above the measurement 

range (above 43.0 °C).

Review instructions and repeat 

measurement. Consult with a 

physician if the problem 

persists.

MAINTENANCE

• RT-1242 (TD-1242) has no user serviceable internal parts except battery 

  replacement.

• Always replace Sensor Cap (or place in cradle) when not in use.

• Store in a dry location free of dust and away from direct sunlight.

• Use a soft dry cloth to clean the plastic housing or a cloth dampened with a 

solution of water and mild detergent. Occasionally, 70% isopropanol solution 

may be used. Never submerge in liquid.

• The sensor window is recessed to assist in keeping it clean and free of 

debris. Inspect the lens and remove any debris. Smudges may be cleaned by 

gently wiping the window with a small foam-tipped swab (non-lining) 

moistened with 70% alcohol. Wait 10 minutes prior to taking temperatures.

TROUBLESHOOTING

SPECIFICATIONS

Model No.: RT-1242 (TD-1242)

Dimension & Weight: 155.46 (L) x 40.14 (W) x 39.45 (H) mm, 61.8g (without 

battery)

Power Source: 2 x 1.5V AAA alkaline batteries

Battery Life: With new batteries, approx. 3,000 measurements.

Displayed Temperature range

• Forehead: 32°C to 43°C (89.6°F to 109.4°F)

• Surface: 0°C to 100°C (32°F to 212°F)

Display Resolution: 0.1°C / 0.1°F

Accuracy: Meet the accuracy requirement specified in ASTM E1965-98 and 

ISO 80601-2-56

• Forehead:

±0.2°C (±0.4°F) for the range of 35.0°C to 42.0°C (95°F to 107.6°F) / ±0.3°C

(±0.5°F) for the range of <35.0°C (95°F) or >42.0°C(107.6°F)

• Surface: ±1°C (±2°F)

Temperature Unit: °C (Default) or °F

Operating Temperature range: 10°C to 40°C (50°F to 104°F)

Operating Humidity: below 85% relative humidity 

(non-condensing)

Operating Atmosphere Pressure: 700hPa to 1060hPa

Storage / Transportation Temperature range: -20°C to 60°C (-4°F to 140°F)

Storage / Transportation Humidity: below 85% relative humidity 

(non-condensing)

Memory Capacity: 30 measurements

Ingress Protection Rating: IP22

Expected Service Life: 3 years

The specifications may be changed without prior notice.

REFERENCE STANDARDS

Device Standard:

Device Corresponds to the requirements of the standard for infrared 

thermometers ASTM E1965-98, EN ISO 80601-2-56:2017, EN 

60601-1-2:2015, EN 60601-1:2006+A12:2014, EN 60601-1-6:2010

Electromagnetic Compatibility:

Device fulfills the stipulations of the standard EN 60601-1-2.

The stipulations of EU-Directive 93/42/EEC for Medical Devices Class Ila have 

been fulfilled.

Electromagnetic environment-guidance 

(for home and professional healthcare environment)

Emission test

RF emissions CISPR 11

RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker emissions IEC 61000-3-3

Group 1

Class B
Not applicable
Not applicable

Compliance

Manufacturer’s declaration-electromagnetic emissions

RT-1242 (TD-1242)  is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.

The customer or the user of the RT-1242 (TD-1242) should assure that it is used in such an environment.

RT-1242 (TD-1242)  uses RF energy only for its internal function.  

Therefore, its RF emissions are very low and are not likely to 

cause any interference in nearby electronic equipment.  
RT-1242 (TD-1242)  is suitable for use in all establishments, 

including domestic establishments and those directly connected 

to the public low-voltage power supply network that supplies 

buildings used for domestic purposes.   

Electromagnetic environment-guidance            

(for home and professional healthcare environment)

Electrostatic 

discharge(ESD) IEC 

61000-4-2

Compliance level

IEC 60601 test level

Immunity test

Manufacturer’s declaration-electromagnetic immunity

RT-1242 (TD-1242)  is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.

The customer or the user of the RT-1242 (TD-1242)  should assure that it is used in such an environment.

Floors should be wood, concrete or ceramic tile. If floors are 

covered with synthetic material, the relative humidity should 

be at least 30%
Mains power quality should be that of a typical home and 

professional healthcare environment.

Mains power quality should be that of a typical home and 

professional healthcare environment.

RT-1242 (TD-1242) power frequency magnetic fields 

should be at levels characteristic of a typical location in a 

typical home healthcare environment. 

Mains power quality should be that of a typical home and

professional healthcare environment. If the user of the 

RT-1242 (TD-1242) requires continued operation during 

power mains interruptions, it is recommended that the 

RT-1242 (TD-1242) be powered from an uninterruptible 

power supply or a battery.   

Electrical fast 

transient/burst IEC 

61000-4-4
Surge IEC 

61000-4-5

Power frequency(50, 

60 Hz) magnetic field 

IEC 61000-4-8

Voltage Dips, short 

interruptions and 

voltage variations on 

power supply input 

lines IEC 61000-4-11

Contact:±8 kV 

Air±2 kV,±4 kV,

±8 kV,±15 kV
Not applicable

Not applicable 

Not applicable

Not applicable

30 A/m

50 Hz and 60 Hz

Voltage dips:

Not applicable 

Not applicable 

Not applicable 

Voltage interruptions:

Not applicable

Contact:±8 kV 

Air±2 kV,±4 kV,

±8 kV,±15 kV
+ 2kV for power supply lines

+ 1kV for input/output lines

+ 0.5kV, +1kV line(s) to line(s)

+ 0.5kV, +1kV,+ 2kV line(s) to 

earth

30 A/m

50 Hz or 60 Hz

Voltage dips:

0 % UT; 0,5 cycle

0 % UT; 1 cycle

70 % UT; 25/30 cycles

Voltage interruptions:

0 % UT; 250/300 cycle

NOTE

 UT is the a.c. mains voltage prior to application of the test level.

Separation distance according to frequency of transmitter

m

0,01

0,1

1

10

100

N/A
N/A
N/A
N/A
N/A

0,12
0,38

1,2
3,8

12

0,23
0,73

2,3
7,3
23 

Rated maximum output power 

of transmitter

W

150 kHz to 80 MHz

d =1,2√P

80 MHz to 800 MHz

d =1,2√P

800 MHz to 2,7 GHz

d =2,3√P

Recommended separation distance between

portable and mobile RF communications equipment and RT-1242 (TD-1242)

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be 

estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter 

in watts (W) according to the transmitter manufacturer.

NOTE1

 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE2

 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 

structures, objects and people.

385

380 –390

TETRA 400

Pulse

modulation b)

18 Hz

1,8

0,3

27

27

710

745

780

704 – 787

LTE Band 13,

17

Pulse

modulation b)

217 Hz

0,2

0,3

9

9

5 240

5 500

5 785

5 100 –

5 800

WLAN 802.11

a/n

Pulse

modulation b)

217 Hz

0,2

0,3

9

9

810

870

930

800 – 960

GSM 800/900,

TETRA 800,

iDEN 820,

CDMA 850,

LTE Band 5

Pulse

modulation b)

18 Hz

2

0,3

28

28

2 450

2 400 –

2 570

Bluetooth, 

WLAN,

802.11 b/g/n, 

RFID 2450, 

LTE Band 7

Pulse

modulation b)

217 Hz

2

0,3

28

28

1 720

1 845

1 970

1 700 –

1 990

GSM 1800;

CDMA 1900;

GSM 1900; 

DECT;

LTE Band 1, 3,

4, 25; UMTS

Pulse

modulation b)

217 Hz

2

0,3

28

28

450

430 – 470

GMRS 460,

FRS 460

FM c)

±5 kHz deviation

1 kHz sine

2

0,3

28

28

Test 

frequency

(MHz)

Maximum 

power

(W)

Distance

(m)

IMMUNITY 

TEST LEVEL

(V/m)

Compliance LEVEL

(V/m) (for home and 

professional healthcare)

Band 

a)

(MHz)

Service 

a)

Modulation 

b)

Manufacturer’s declaration-electromagnetic immunity

Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment

RT-1242 (TD-1242) is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.

The customer or the user of RT-1242 (TD-1242)should assure that it is used in such an environment.

NOTE

   If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or 

ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.

b) The carrier shall be modulated using a 50 % duty cycle square wave signal.

c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual 

modulation, it would be worst case.

Electromagnetic environment-guidance

(for home and professional healthcare environment)

Conducted RF

IEC 61000-4-6

Radiated RF 

IEC 61000-4-3

Compliance level

IEC 60601 test level

Immunity test

Manufacturer’s declaration-electromagnetic immunity

RT-1242 (TD-1242)  is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.

The customer or the user of the RT-1242 (TD-1242)  should assure that it is used in such and environment.

Recommended separation distance:

d = 1,2 √P

d = 1,2 √

P  80MHz to 800 MHz

d = 2,3 √

P  800MHz to 2,7 GHz 

Where P is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter 

manufacturer and d is the recommended separation 

distance in metres (m).

Interference may occur in the vicinity of equipment marked 

with the following symbol:

Not applicable

Not applicable

10 V/m

80 MHz – 2,7 GHz 

80 % AM at 1 kHz 

3 Vrms:

0,15 MHz – 80 MHz

6 Vrms: 

in ISM and amateur

radio bands between

0,15 MHz and 80 MHz 

80 % AM at 1 kHz 

10 V/m

80 MHz – 2,7 GHz 

80 % AM at 1 kHz 

NOTE1

 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE2

 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 

structures, objects and people.

Portable and mobile RF communications equipment should 

be used no closer to any part of RT-1242(TD-1242) 

including cables, than the recommended separation 

distance calculated from the equation applicable to the 

frequency of the transmitter.   

RT-1242 (TD-1242) is intended for use in an electromagnetic environment (for home and professional healthcare) in which radiated RF 

disturbances are controlled. The customer or the user of RT-1242 (TD-1242) can help prevent electromagnetic interference by 

maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and RT-1242 (TD-1242) 

as recommended below, according to the maximum output power of the communications equipment.   

Symbol

Referent

Consult instructions for use

This device does not belong to household waste and must be 

returned to a collection point for recycling electric and electronic 

devices according to local laws. If it contains batteries, the 

batteries should be removed and disposed in accordance with local 

regulations for separate collection of spent batteries.

Manufacturer

Serial number

Caution

Authorized representative in the European Community

CE mark

FDA mark

Symbol

Referent

Type BF applied part

Temperature limit

Humidity limitation

Resistant to liquid ingress

RoHS Compliance

IP22

0123

Reviews: