Exceptional Situations

Problem

Causes and Solutions

No display appears on the 

display screen.

No power.

Measurement values 

appear too high or too low.

Check the battery installation for proper 

placement of the batteries polarities.

Replace all the worn batteries with 4 new 

ones.

Blood pressure varies constantly. Many 

factors including stress, time of day, how 

you wrap the cuff, may affect your blood 

pressure. Review the sections "Proper Use 

of the Arm Cuff" and "Take a Measurement".

Eliminate Fault

Symbol

Cause

Correction

When the batteries 

power are too low.

The course of 

inflating appears 

error.

When measurement 

fails.

Replace all of the worn batteries 

with new ones.

Wrap the cuff correctly and tightly.
Ensure the connection is close

between the air plug and the unit.
Inflate over again after ensuring.

Measure over again according to 

correct way.

Do not move your arm and body,

and keep quiet.

Error Indicators

22

The following symbol will appear on the display when measuring abnormally.

23

Care and Maintenance

Do not clean the body and cuff with 

naphtha, thinner or gasoline etc.

Do not wet the cuff or attempt to 

clean the cuff with water.

Store the unit in a clean and dry 

location. Do not subject the unit to 

extreme hot or cold temperature, 

humidity and direct sunlight.

Remove the batteries if the unit will 

not be used for 3 months or longer.

Always replace all of the batteries 

with new ones at the same time. 

Use the unit consistent with the instructions provided in this manual. 

Use only authorized parts and accessories.

 naphtha

thinner

gasoline

Maintenance

Keep the unit in the storage case 

when not in use.
Clean the unit with a soft dry cloth. 

Do not use any abrasive or volatile 

cleaners. 
Never immerse the unit or any 

components in water.

Care

If needed, please use ultraviolet radiation for disinfection.

Avoid subjecting the monitor to strong shocks or vibrations, 

such as dropping the unit on the floor.

+

-

+

+

-

-

+

-

Summary of Contents for BA2015

Page 1: ...Model No BA2219 A00...

Page 2: ...3 Konw Your Unit 4 5 IHB Introduction 6 Before Taking A Measurement 7 Quick Reference Guide 8 Battery lnstallation Replacement 9 Adjusting Settings 1 To set year 10 2 To set date and time 10 11 3 To...

Page 3: ...or moderate injury to the user or patient or damage to the equipment or other property Read all information in the instruction manual and any other literature included in the box before using the uni...

Page 4: ...Hose Air Plug Insert to Air Socket Metal Arm Cuff Warning 1 Do not use rechargeable batteries 2 When using optional AC adapter the AC adapter must comply with the requirements of standard IEC60601 1 3...

Page 5: ...d bathing for 30 minutes before taking a measurement 2 Stress raises blood pressure Avoid taking measurements during stressful times 3 The cuff can be applied to your left or right wrist Always measur...

Page 6: ...ative polarities match the polarities of the battery compartment as indicated Remove the battery cover from the battery compartment Replace the battery cover Press the START STOP button All symbols ap...

Page 7: ...the month and press the SET button and continue with next step Then set the day the time system conversion the hour and the minute by using the same way Continue with step 3 2 To set date and time 10...

Page 8: ...the START STOP button to confirm user and turn off You can stop the setting anytime when you press the START STOP button to save the current setting and turn off If select to display the date and time...

Page 9: ...plug is inserted properly in the main unit 1 15 Proper Use of the Arm Cuff 8 Relax your arm keep the palm upward and the fingers curving naturally then turn on the unit and start to measure Wrap the...

Page 10: ...uff starts to inflate automatically Press the START STOP button 1 In order to accelerate the measurement time for the systolic pressure more than 180 mmHg press and hold the memory button until the un...

Page 11: ...memory To read the average value Press the button The average value symbol will display the average value of the latest 3 times measurement taken Turn to next The 1st latest measurement value The 2nd...

Page 12: ...nts the blood pressure produced by relaxation of the heart muscle Blood Circulation 21 About Blood Pressure Note When a person s systolic and diastolic pressures fall into different categories the hig...

Page 13: ...ve your arm and body and keep quiet Error Indicators 22 The following symbol will appear on the display when measuring abnormally 23 Care and Maintenance Do not clean the body and cuff with naphtha th...

Page 14: ...y using the oscillometric method The unit is intended for usingin only adult population not applied to the other populations such as neonatal baby It can t be used while the wirst arm has bleeding or...

Page 15: ...60601 1 2 Medical electrical equipment Part 1 2 General requirements for basic safety and essential performance Collateral standard Electromagnetic compatibility Requirements and tests EN1060 1 Non in...

Page 16: ...1000 4 8 6 kV contact 8 kV air 6 kV contact 8 kV air 3A m 3A m Floors should be wood concrete or ceramic tile If floor are covered with synthetic material the relative humidity should be at least 30 P...

Page 17: ...be necessary such as reorienting or relocating the Digital Blood Pressure Monitor Radiated RF IEC 61000 4 3 IEC 60601 test Compliance level level Electromagnetic environment guidance Recommended sepa...

Reviews: