24 |
P a g e R e v . 1 . 1
I s s u e 3 0 - 1 1 - 2 0 1 8
Certification/Standards
CE, ATEX
2014/34/EU
/
EN 60079-0:2012 + A11:2013, EN 60079-
11: 2012, EN 60079-26: 2015
0158
II (1) G II (1) D ; [Ex ia
Ga] IIC; [Ex ia Da] IIIC
Functional safety: SIL2 according to IEC 61508/61511
EMC 2014/30/
EU
EN 61326-3-2:2008; EN 61000-6-3: 2007 + A1: 2011
EN 61000-4-2:2009; EN 61000-4-3: 2006 + A1: 2008 + A2: 2010
EN 61000-4-4:2012
2006 + A1: 2008 + A2: 2010
EN 61000-4-5:2014
EN 61000-4-6:2014
Oxygen Sensor
Specifications-Electrical
Models
OC-25, OC-26, OC-200 series
Range
0-25% Oxygen
Signal Output
300-375mV in Air OC25, OC-26. 135-160mV in Air OC-200
Technology/Lifespan
Electrochemical Solid State Long Life / 3-5+ Years application
dependant. Storage Life 1 Year.
Connections
Circular Pluggable IP67 Connector with cable to specified length.
Intrinsically Safe Sensor
Input Connection
Parameters.
voltage Uo
DC
12 V
current intensity Io
120 mA
power Po
0.55 W
max. outer inductivity Lo
0 mH
max. outer capacity Co
1.2
μF
Specifications-Mechanical
Dimensions OC-25
100mm High x 50mm Diameter
Dimensions OC-26
75mm High x 55 Diameter (KF40)
Dimensions OC-200
Variable length to order, typically 200mm long x 12mm Diameter probe.
Max. Diameter is 30mm.
Summary of Contents for SIL O2
Page 2: ...2 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8...
Page 5: ...5 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8...
Page 14: ...14 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8...
Page 28: ...Gas Measurement...
Page 29: ...Gas Measurement...
Page 30: ...Gas Measurement...
Page 31: ...Gas Measurement...
Page 32: ...Gas Measurement...
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 41: ......
Page 42: ......
Page 43: ......
Page 44: ......
Page 45: ......
Page 49: ......
Page 52: ......
Page 53: ......
Page 54: ......
Page 55: ......