Page 20
4.1.13 Menu 13
– Sensor damping
Oxygen sensors output change when subjected to pressure changes.
The damping option is used to allow the user to reduce the effects of sudden
changes by applying digital filtering. The larger the number the more the signal
damping that is applied.
Note:
The minimum setting is 5.
•
Press MENU to open the menu system.
•
Using the NEXT and PREVIOUS buttons, select menu option:
E:13
•
Press ENTER. The display shows the damping factor that is
applied to the sensor.
•
Use the INCREASE and DECREASE buttons to set the required
damping level.
•
Press ENTER to store the new value.
Note: Pressing the MENU button rather than the ENTER button
exits without any change.
•
Press MENU to close the menu system.
Note: The signal damping setting will be displayed on exit
while the MENU key is pressed.
4.1.14 Menu 14
– New sensor data
This option allows the initial sensor calibration data to be set. It is used to
predict the remaining sensor life.
•
Press MENU to open the menu system.
•
Using the NEXT and PREVIOUS buttons, select menu option:
E:14
•
Press ENTER. The display displays E:14
•
Press ENTER to store the new data in the memory.
Note: Pressing the MENU button rather than the ENTER button
exits without any change.
•
Press MENU to close the menu system.