NSK Varios 750 Operation Manual Download Page 6

8

9

Fig. 20

Fig. 19

Water Selection Mode Key

Dry

Bottle

City Water

(4) Water Mode Selection

Select the mode with the Water Selection Key on the 
Front Panel. Bottle, City Water or Dry mode(No Irrigation 
Mode) is available. (Fig. 19)

(5) Operate Varios750/750LUX

Depress the Foot Control. Tip will vibrate. In case of the Varios 750 LUX, the handpiece donut -shape light will illuminate.
Adjust the power level to the proper range. Adjust the water spray as follows.

Turn the Water Adjustment Knob counterclockwise 

gradually to increase the supply volume. (Fig. 20)

Water Supply Volume Adjustment

Water Supply Volume Adjustment

· Always use the water supply unless you are doing procedure requiring dry mode. If water supply is insufficient, 

handpiece will overheat patient´s tooth surface can be injured.

· Prior to starting the procedure, verify that the water spray is clean and at adequate volume.

(                                             )

CAUTION

When using medicated solutions, clean the entire Irrigation System thoroughly.

CAUTION

(6) When Treatment is Finished.

Close the dental unit's water valve.

Thoroughly wash the Bottle Water Supply system. Refer to the  11. (4)  Auto Cleaning (Cleaning of Irrigation Tube

 (                                     ) ).

Release the Foot Control and power off the Control Unit by turning off the Power Switch.

Use of City Water

Use of City Water

Use of Bottle

Use of Bottle

Use of Bottle

Use of Bottle

When the Control Unit is shut down, the last mode settings in use are automatically retained in memory. When you 
power on the system for the next procedure, the system will reactivate in the same configuration as when shut 
down.

NOTICE

Use of City Water

Use of City Water

Increase

Increase

Water

Adjustment Knob

Water

Adjustment Knob

Bottle Water

Adjustment Knob

Bottle Water

Adjustment Knob

Increase

Increase

Memory Button

Fig. 21

Fig. 23

Fig. 22

   Safety Function

If you operate more than 10 minutes and operate mode set to “G”, also power setting to maximum, the Control Unit  
automatically reduce power. (The output display moves 3 lamps from “Max” to “4”. ) 

To increase power again after auto reduction, release the Foot Control completely and depress  it again.

   Memory Function

The Control Unit can be preset for water spray mode, 
power setting and operating mode. To set the desired 
preset conditions into memory follow the procedure 
below.

1) Set the desired levels.
2) Push the Memory Button for about 2 seconds or 

more, the Memory Light will illuminate and the 
Control Unit will produce a beep. The memory is set. 
(Fig. 21)

To enter new settings

To enter new settings

After confirming the settings of Water Spray Mode, Power Setting and Operating Mode, push the Memory Button for 
approx. 2 seconds. (The Memory light will illuminate.) The old memory is cleared and new settings are programmed.

Removing
tartar on the 
neck of tooth

Apply the top of the Tip on the 
tooth plane and move it 
sideways finely along the neck 
of tooth. (Fig. 22)

7. Provided Scaler Tips

Place the side of the Tip parallel 
with the tooth plane and finely 
move it up and down. (Fig. 23)

Insert the Tip between the teeth 
vertically and move it slightly. 
Always touch the tip on the 
tooth.

 G1

This Tip is designed for removing tartar on the neck of tooth (the tooth plane), 
between teeth, and below the gum margin. Set the level less than “     3”.

Removing
tartar on the 
tooth plane

Removing
interdental
calculus

···

···

···

G

Summary of Contents for Varios 750

Page 1: ...OPERATION MANUAL Please read this Operation Manual carefully before use and file for future reference Multi Function Ultrasonic Scaler Optic Non Optic OM E0144E 003 10 03 005 B ...

Page 2: ...tions are classified as follows in accordance with the seriousness of the risk Cautions for handling and operation WARNING C A U T I O N N O T I C E A hazard that could result in bodily injury or damage to the device if the safety instructions are not followed A hazard that could result in light or moderate bodily injury or damage to the device if the safety instructions are not followed General in...

Page 3: ...ppen when using the Non NSK Tips Vibration failure caused by using nonconformable screws Patients accidental ingestion of damaged Tips Damage of thread ridge of handpiece Do not sterilize by ultraviolet light Handpiece could discolor If chemical solvent or antiseptic solution is deposited on this product immediately wipe it away Discoloration or deformation may occur if left as it is If water is s...

Page 4: ... Fig 4 NOTICE Pushing the White Ring the Quick Connector Release Ring on the Water Connector gently remove the Tube Fig 4 When the water tube is not connected mount the cover on the connector on the Control Unit NOTICE To remove the bottle pull up the bottle gently The Irrigation Bottle graduations are printed on both sides of the bottle and can be read accurately from the fill position or mounted ...

Page 5: ...t s water valve Set water pressure between 0 1 0 5MPa 1 5 kgf cm2 Check that the Bottle is filled to the proper level 6 Operating Procedures Use of City Water Use of City Water Use of Bottle Use of Bottle Tip Tip Wrench Tip Wrench Tighten Tighten Tighten Tighten Loosen Loosen Loosen Loosen Mark Mark Mark Mark Joint Joint Handpiece Handpiece Handpiece Plug Handpiece Plug Fig 16 Tip Case Label Set to...

Page 6: ... as when shut down NOTICE Use of City Water Use of City Water Increase Increase Water Adjustment Knob Water Adjustment Knob Bottle Water Adjustment Knob Bottle Water Adjustment Knob Increase Increase Memory Button Fig 21 Fig 23 Fig 22 Safety Function If you operate more than 10 minutes and operate mode set to G also power setting to maximum the Control Unit automatically reduce power The output di...

Page 7: ... the green yellow and red line to check wear of the Tip See below what each color means At NSK we recommend to replace a Tip when the Tip meets the yellow line wear of 1mm to guarantee safe and effective use Green No wear Tip is OK Tip replacement is not necessary Yellow Wear of 1mm Tip is showing some wear Tip replacement is recommended Red Wear of 2mm Tip is badly worn Tip replacement is necessa...

Page 8: ... released position and turn clockwise until it clicks Fig 38 6 Insert the Irrigation Tube into the Groove on the reverse side with slack between pump and groove 7 Connect the Irrigation Tube to the connector on the handpiece cord connector 8 Insert the Tube Connector to the Joint Ring until it clicks Optional Irrigation Pump Order No U424061 Assemble the Water Filter in the correct direction as sh...

Page 9: ... the mark on the cover with the same on the Handpiece Cord Push the cover to the cord unit it clicks Fig 44 Assemble the handpiece and the Handpiece Cord Optional Lamp VA Bulb Set A Set consist of 3 pieces Order No Y900107 5 Cleaning of Donut Shape Light Varios750LUX Wipe the debris off the end of the Optic Fibers at the handpiece with alcohol soaked cotton swab Fig 42 Following method is also ava...

Page 10: ... be subjected to autoclave sterilization Perform alcohol disinfection to the Control Unit Power Cord Foot Control Handpiece Cord including with reference after every patient WARNING ON STERILIZATION Fuse Power Connector Assembly Power Connector Assembly Fuse When trouble is found check the followings prior to consulting dealer 13 Troubleshooting No Poor vibration The Tip is bent or broken The Tip ...

Page 11: ...ly insert the Irrigation Tube into the Irrigation Connector inmost Replace with new O ring Refer to 11 1 Changing O ring Handpiece Varios 750 Contact your dealer Firmly insert the handpiece into the Handpiece Cord inmost Mount the Lamp correctly and securely Replace with new lamp Refer to 11 6 Changing Lamp Varios 750 LUX Contact dealer Power output will weaken autom atically while continuous oper...

Page 12: ...ot freeze up Humidity 30 75 Atmospheric pressure 700 1060 hPa Temperature 10 60 C Humidity 10 85 Atmospheric pressure 500 1060 hPa NE134 AC120V 50 60Hz AC230V 50 60Hz 28 32kHz 8W Max 42VA 0 1 0 5MPa 1 5kgf cm2 Varios750 No Varios750LUX Yes Guidance and manufacturer s declaration electromagnetic immunity The Varios750 Varios750LUX is intended for use in the electromagnetic environment specified bel...

Page 13: ...UX can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Varios750 Varios750LUX as recommended below according to the maximum output power of the communications equipment For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be est...

Reviews: