NSK Care3 Plus

When using NSK Care3 Plus automatic handpiece cleaning and 
lubrication system, refer to the Care3 System instructions.

6-3 Sterilization

Sterilize the product by autoclave sterilization. Remove the bur after 
each patient and sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.

Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or 
3 min. at 134°C.

3) The handpiece should remain in the autoclave pouch until required 

for use.

Do not autoclave the product with other instruments even when it 
is in a pouch. This is to prevent possible discoloration and damage 
to the product from chemical residue on other instruments.
Keep the product in suitable atmospheric pressure, temperature, 
humidity, ventilation, and sunlight. The air should be free from 
dust, salt and sulphur.
Immediately after use, the product should be cleaned, lubricated 
and sterilized. If blood remains on the external or internal surfaces 
it can become clotted and cause rust.
Do not heat or cool the product too quickly. Rapid change in 
temperature could cause damage to the product.
If the sterilizer chamber temperature may exceed 135°C during 
the dry cycle then delete the dry cycle.
Autoclave sterilization is recommended for the product. The 
validity of other sterilization methods is not confirmed.
Do not touch the product immediately after autoclaving as it will 
be very hot and must remain in a sterile condition.

CAUTION

Summary of Contents for FX65

Page 1: ......

Page 2: ...Open Close 1 2 Chuck Open Chuck Close...

Page 3: ...3...

Page 4: ...ct nor tamper with the mechanism except as recommend by NSK in this Operation Manual Do not allow any impact on to the product Do not drop the product Operators and all others in the area must wear ey...

Page 5: ...be autoclaved prior to use Perform regular function and maintenance checks If the product is not used for a long period check it is functioning correctly before using on a patient To avoid clinical d...

Page 6: ...he bur shank clean Entry of debris into the chuck via the bur shank could cause bur rotation slip and also prevent the bur from being securely located in the chuck Do not exceed the bur speed recommen...

Page 7: ...e effect of lubrication and could cause corrosion inside of the product To clean the product never use any solvent such as benzine or thinner CAUTION 6 2 Lubrication NSK PANA SPRAY Plus Apply NSK PANA...

Page 8: ...le discoloration and damage to the product from chemical residue on other instruments Keep the product in suitable atmospheric pressure temperature humidity ventilation and sunlight The air should be...

Page 9: ...Maintenance Checks Perform periodical maintenance checks every three months referring to the check sheet below If any abnormalities are found contact your Authorized NSK Dealer 9 Warranty NSK products...

Page 10: ...ors handling the disposal of medical equipment as well as the risks of environmental contamination caused thereof a surgeon or a dentist is required to confirm the equipment is sterile Ask specialist...

Page 11: ...5 4846 48 8 43 5 3 5 6 8 4 96 5 9 0 3 3 8 96 3 8 460 3 68 2 8 6 7 3 6432 38 6 375468 8 43 3 846 3 6432 38 624 7 3 846 2 5 22 96 645 3 5 2 3 2 3 6 8 6 22 22 22 25 6 896 4 43 37 8 43 92 8 82475 6 6 779...

Page 12: ......

Reviews: