CAUTION

• When operating the handpiece always consider the safety of the patient.

• Check for vibration, noise and overheating outside the patient's oral cavity before use. If any abnormalities are found, stop using the handpiece 

immediately and contact dealer.

• Should the handpiece function abnormally, cease operation immediately and return the handpiece to an authorized dealer for repair.

• Depressing the push-button while the handpiece is rotating will result in OVERHEATING of the handpiece head. Special caution must be exercised 

during use to keep cheek tissue AWAY from the push-button of the handpiece. Contact with cheek tissue may cause the push-button to depress and 

burn injury to the patient may occur.

• Do not applying excessive power in the direction which bur removes. When the bur drops into patient’s oral cavity, they may drink by mistake and get 

injured.

• This handpiece have built-in dynamo. Consumption Air generates electricity and LED become luminous. Therefore, do not perform autoclave 

sterilization which it heat and cools rapidly. 

• Do not allow any impact on to the Handpiece. Do not drop the Handpiece.

• Do not look into LED light directly.

CAUTION

• Ensure connect “Coupling and Hose” or “Handpiece and Coupling” firmly.

• Do not pull the Connector Ring while the handpiece is under air pressure. It may jump the handpiece out from the connector because of the 

air pressure. 

• Do not exceed the recommended air pressure. If it exceeds the pressure, it may bur jump out from the handpiece.

• Ensure clean, dry air supply. Moisture and debris in air supply may cause the motor to malfunction.

CAUTION

• Do not connect or disconnect the handpiece or bur until the motor has completely stopped.

• When used for crown removal and other procedures that would involve heavy vibration, the bur may become difficult to remove. Push the 

button firmly, and forcibly pull out the bur with a pair of pliers. Apply small amount of spray lubricant into the chuck. This can be prevented if 

the bur is repositioned a few times during the procedure.

• When abnormalities are in LED (dark, do not light on or flashing or so), contact dealer to repair.

• Do not attempt to disassemble the handpiece or tamper with the mechanism except NSK recommend on this Operation Manual.

• Always keep bur shanks clean. Dirt or debris in a chuck could cause poor bur concentricity or low chuck retention force.

• Do not exceed the drive motor speed recommended the bur manufacturers. If the rotary speed which the manufacturer of bur nominates is lower than 

spec of handpiece, do not use it.

• Do not connect or disconnect the handpiece or bur until the motor has completely stopped.

• NSK has not recommended the repair and disassembly of the cartridge without exception. Handpiece may perform abnormally (abnormal noise or 

abnormal vibration). Damage, failure or accidents are outside of our guarantee.

• Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.

• Use eye protectors, mask for safety when operating this handpiece.

• User is responsible for operation, maintenance and safety.

• If the handpiece has not been used for a long period, check for noise, vibration, and overheating before use.

• Treatment under the sufficient irrigation water(Spraying water).

1. Specification

Intended to use: 

This handpiece is designed only for clinical dental use by qualified personnel.

3. Check before treatment

Follow the check below, before use. If any abnormalities are found, stop using the handpiece immediately and contact the authorized NSK dealer.

1) Insert the Bur. (Do not exceed maximum rotation speed for any item.)

2) Check that there is no abnormal rotation, vibration, noise and overheating.

3) If there are no immediate apparent abnormalities, continue to rotate the handpiece for about 1 minute. Then check that there is no abnormal 

heating on the outer surface.

4) Operate it if there are not abnormalities above.

CAUTION

• Do not use non-standard burs. The ISO standard shank diameter is 

φ

1.59 - 

φ

1.60 mm.

• Do not exceed the bur speed recommended by the bur manufacturer. Moreover, do not use the bur which it has lower rotation speed. 

• Do not use long surgical burs. Do not use burs longer than 26mm.

• Do not use short shank burs. 

(

M600LG M4

M600LG B2

M600LG QD

)

• Always keep the bur shank clean. Entry of hard debris into the chuck via the bur shank could cause rotation slip and also prevent the bur from 

being securely located in the chuck.

• Fully depress the Push Button and insert the bur into the chuck until it is secure.

• Test that the bur is secure by gently pulling and pushing the bur without depressing the Push Button.

4. Insertion /Removal of the Bur

2. Connecting/Disconnecting Handpiece

M500LG M4/B2, M600LG M4/B2

2-1 Connecting

1) Insert the handpiece into the hose connector.

2) Tighten the Hose Nut. (Fig.1)

2-2 Disconnect

Loosen the Hose Nut and pull out the Hose.

M500LG QD, M600LG QD
2-3 Connecting

1) Insert the Coupling into the Hose Connector.

2) Tighten the Hose Nut.

3) Pull the Connector Ring.

4) Push into the handpiece into Coupling.

2-4 Disconnect

Slide back the Connector Ring as shown in Fig.2.

SUPPLY AIR PRESSURE Setting

Set the Air pressure at the “Air Check point” on Fig. 1 and Fig. 2 The setting values are on the specification 

table. Do not exceed the recommended air pressure by using Multi Gauge.

Multi Gauge: Midwest 4 Hole  Order code Z109400 / Borden2/3 Hole  Order code Z109200

4-1 To inserts the Bur

1) Insert the bur until it stops.(   )

2) Depress the Push Button.(   )

3) Insert the bur into the chuck until it is stops.(   )

4) Release the button.

4-2 To remove the Bur

Depress the Push Button firmly and remove the bur.(   )

CAUTION

• Be sure to hold the handpiece firmly to prevent the handpiece from slipping out of hand by the spray pressure when lubricating.

• Supply lubricant until it comes out of the handpiece head.(at least 2 seconds)

• Hold the spray can upright.

9. Non-Retraction Valve (QD Coupling)

To prevent oral fluids drawn into the water line, QD coupling has Non-Retraction valve. If water is 

beginning to leaking from handpiece, replace non-retraction valve.

1) Remove the back-end gasket.

2) Replace Non-Retraction Valve.

10. Warranty

NSK warrants our product against poor manufacturing, poor workmanship and defects in materials. NSK reserves the right to analyze and determine 

the cause of any problem. Warranty is voided should the handpiece be not used correctly. Replacement parts are available for seven years beyond 

discontinuation of the model.

5-4 Lubrication

Lubricate by PANA SPRAY Plus

1) Mount the Arrow Head Spray Nozzle into the Spray Port.(Fig.7)

2) Remove the Bur from the handpiece.

3) Insert the Arrow Head Spray Nozzle in the drive Air Tube of the 

handpiece and spray for approximately 2-3 seconds.

1) Mount the Spray Nozzle into the Spray Port.(Fig.8)

2) Remove the bur from the handpiece.

3) Insert the Spray Nozzle in rear of the handpiece and spray for approximately 2-3 seconds.

5. Maintenance

After the treatment of each patient, perform maintenance below.

5-1 Cleaning (Clean-Head System)

1) Scrub dirt and debris from the handpiece.(Fig.4)

2) Half fill a cup with clean water. 

3) Rotate the handpiece and immerse half of the handpice head. (Fig. 5)

4) Rotate and stop intermittently the handpiece about 2 to 3 seconds. (About 3 times)

5) By the function of the clean head system, the clean head hole can be washed.

6) Stop the handpiece and wipe it dry. If the dirt could not remove from the hole, remove the cartridge and clean by brush.

7) Perform lubrication (4-3 Lubrication) and sterilization. (5-4 Lubrication, 5-5 Sterilization)

5-5 Sterilization

Autoclave sterilization is recommended. Sterilization is required first time you use and after each patient as noted below.

This handpiece and motor can be autoclavable up to Max.135°C.

Lubricate by Care3 plus

NSK Care3 Plus automatic handpiece cleaning and lubrication system.

After connecting the handpiece to the correct adaptor, activate the Care3 System per the Care3 System Operation Manual.

Autoclave Procedure:

1) Clean and lubricate the handpiece.

2) Insert into an autoclave pouch. Seal the pouch.

3) Autoclavable up to max. 135˚C.

ex.) Autoclave for 3 min. at 134°C.

4) Keep the handpiece in the autoclave pouch to keep it clean until you use it.

* EN13060/EN ISO17665-1 recommends autoclaving for 3 minutes (minimum holding time) at 134°C.

NSK recommends Class B sterilization.

M500LG M4

Mini

Mini

ISO9168 Type 2 4 hole

380,000 - 450,000 min

-1

0.20MPa(2.0kgf/cm

2

) - 0.25MPa(2.5kgf/cm

2

)

Quattro Spray

Push Button Chuck

ISO 1797-1 Type3 

φ

1.59-1.60mm  Short Shank Bur

9.0mm

21mm

φ

2mm

45±5NL/min

M500LG B2

ISO9168 Type 1 2 hole

40±5NL/min

M500LG QD

M600LG M4

Standard

Standard

Mini

Standard

330,000 - 430,000

 min

-1

380,000 - 450,000

 min

-1

330,000 - 430,000

 min

-1

380,000 - 450,000

 min

-1

330,000 - 430,000

 min

-1

ISO1797-1 Type3 

φ

1.59-1.60mm Standard Bur

ISO 1797-1 Type3 

φ

1.59-1.60mm  Short Shank Bur

ISO1797-1 Type3 

φ

1.59-1.60mm Standard Bur

ISO 1797-1 Type3 

φ

1.59-1.60mm  Short Shank Bur

ISO1797-1 Type3 

φ

1.59-1.60mm Standard Bur

10.5mm

9.0mm

10.5mm

9.0mm

10.5mm

25mm

21mm

25mm

21mm

25mm

M600LG B2

M600LG QD

QD Coupling

Model

Head Type

Connection Type

Rotation Speed

Drive Air Pressure

Water Spray Type

Chuck Type

Bur Type

Min. Chucking Length

Max. Bur Length

Max. Working Part 

Diameter

Max. Air Consumption

(0.25MPa)

Head Diameter x Height)

φ

10.6mm x 12.4mm

φ

12.1mm x 13.3mm

φ

10.6mm x 12.4mm

φ

12.1mm x 13.3mm

45±5NL/min

50±5NL/min

φ

10.6mm x 12.4mm

φ

12.1mm x 13.3mm

Water Pressure

Chip Air Pressure

0.05MPa(0.5kgf/cm

2

) - 0.20MPa(2.0kgf/cm

2

)

Optic

0.15MPa(1.5kgf/cm

2

) - 0.25MPa(2.5kgf/cm

2

)

-

0.15MPa(1.5kgf/cm

2

) - 0.25MPa(2.5kgf/cm

2

)

Glass Rod

Gasket

Non-Retraction Valve

Drive Air Tube

MidWest

4 hole

Bordent

2 hole

Handpiece

Arrow Head Spray Nozzle

PANA SPRAY plus/PANA SPRAY

Fig. 7

M500LG QD, M600LG QD

M500LG M4/B2, M600LG M4/B2

Fig. 13

Fig. 3

CAUTION

Use only clean water to maintain the Clean Head System.

Fig. 4

CAUTION

• To prevent infectiveness, Autoclave the handpiece every patient.

• Do not autoclave the handpiece, even when it is bagged, together with other instruments, from which chemicals are not sufficiently cleaned. 

This is to avoid discoloration and/or damage to the plating of the handpiece. 

• In case the sterilizer chamber temperature could exceed 135°C during the dry cycle, skip the dry cycle.

• Do not heat or cool the handpiece quickly. Rapid change in temperature could break the glass rod or dynamo or LED parts.

6. Changing Turbine Cartridge

1) Mount a dummy bur in the chuck.

2) Turn the Wrench counter clockwise and remove the Head Cap by Head Cap Wrench.

3) Push up the dummy bur to remove Cartridge.

4) Clean the head interior. (Wipe extra oil after lubrication)

5) Align the Alignment Pin and insert Cartridge. (Fig. 8)

6) Turn the Wrench clockwise and mount the Head Cap. (Fig. 9)

* Replacement Cartridge: SX-MU03 Mini Cartridge Order Code P1001250

SX-SU03 Standard Cartridge Order Code P1002250

2) Take out the cleaning wire from the back end of the brush holder. Insert the wire straight into the spray ports with caution. Do not forcibly insert the 

wire into the port. Damaged or disoriented port could cause the spray diverge or directed away from the bur, and the cooling efficiency deteriorates. 

3) After use, clean the cleaning wire. Push into the brush holder with the pointed end of the wire inward.

7

Spray Port Cleaning

When spray port are clogged, or spray does not exit evenly from three ports, clean the ports as follows:

1) Remove the bur from the chuck. (Refer to “4. Insertion/Removal the Bur” and “6. 

Replacing the Cartridge”)

8. Replacing the O-rings (QD Coupling)

Worn or damaged O-ring may cause air or water to leak. 

When water is found in the exhaust air or water leak from the coupled point, replace four O-rings as shown in Fig.10.

Replacement O-ring Set (4pcs) : Order Code Y900592

Fig. 12

O-Ring

Clean Head Hole

Clean Head Hole

Clean Head Hole

Fig. 5

CAUTION

Do not use a pointed tool or sharp edged to clean the Cellular Glass Optic Rod. They may damage the mirror finish and reduce the light 

transmission efficiency. If Cellular Glass Optic Rod end has become damaged, return it to dealer for repair.

5-3 Cleaning

Scrub dirt and debris from the handpiece, and wipe clean with alcohol-immersed cotton swab or cloth. Do not 

use a wire brush.

* Refer the manual of Thermo-Disinfector for performing it.

This handpiece can be washed via Thermo Disinfector.

5-2 Cleaning (Glass Rod)

Wipe clean the Cellular Glass Optic Rod entry point and exit point with an alcohol-immersed cotton swab. Remove all debris and oil.(Fig.6)

Fig. 6

Cotton Swab

Glass Rod Optic

Glass Rod Optic

Fig. 10

O-Ring

O-Ring

CAUTION

• Tighten the head cap after changing the cartridge. If attachment is loose, there are the causes of an injury in 

oral cavity or possibility of accidental ingestion. 

• Use only NSK purity cartridge. Competitive cartridge is not covered under warranty.

• The screw of head cap is very small. Do not bind tight using a Wrench from the start. After tightening by the 

hand, use the Wrench.

• Check that the O-Rings are placed correct position and then install the cartridge.(Fig.10)

Fig. 9

Cartridge

Arraignment Slit

Head Cap

Head Cap 

Wrench

Arraignment Pin

Fig. 1

1

Cartridge

Spray Port

Cleaning Wire

Cleaning Wire

Fig. 1

Handpiece

Hose Nut

Air Check Point

Fig. 2

Air Check Point

Hose Nut

Connector Ring

Handpiece

Fig. 8

Spray Nozzle

Handpiece

PANA SPRAY plus/PANA SPRAY

• Do not wash, soak, or wipe off the handpiece with/in the oxidation potential solution (strong acid, super acid solution) sterilized solution, strong 

alkaline solution or functional water.

• The handpiece is kept in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. Moreover, air does not include a dust, 

salt and sulfur.

• Do not touch the handpiece just after the autoclave. It may very hot.

• NSK do not recommend you to sterilize on plasma sterilization or EOG sterilization.

High Speed Air Turbine Handpiece

N-600 167X67mm

PANTONE 2945C PANTONE 7546C

DynaLED

002

OM-T0394E

Please read this Operation Manual carefully before use, 

and file for future reference.

2017.03.17   02 S

Summary of Contents for DynaLED

Page 1: ...t with cheek tissue may cause the push button to depress and burn injury to the patient may occur Do not applying excessive power in the direction which bur removes When the bur drops into patient s o...

Page 2: ...dpiece is designed only for clinical dental use by quali ed personnel M500LG M4 Mini Mini ISO9168 Type 2 4 hole 380 000 450 000 min 1 0 20MPa 2 0kgf cm2 0 25MPa 2 5kgf cm2 Quattro Spray Push Button Ch...

Page 3: ...tectors mask for safety when operating this handpiece User is responsible for operation maintenance and safety If the handpiece has not been used for a long period check for noise vibration and overhe...

Page 4: ...fication Intended to use This handpiece is designed only for clinical dental use by quali ed personnel 3 Check before treatment Follow the check below before use If any abnormalities are found stop us...

Page 5: ...mm Do not exceed the bur speed recommended by the bur manufacturer Moreover do not use the bur which it has lower rotation speed Do not use long surgical burs Do not use burs longer than 26mm Do not u...

Page 6: ...e Then check that there is no abnormal heating on the outer surface 4 Operate it if there are not abnormalities above 4 1 To inserts the Bur 1 Insert the bur until it stops 2 Depress the Push Button 3...

Page 7: ...the Connector Ring 4 Push into the handpiece into Coupling 2 4 Disconnect Slide back the Connector Ring as shown in Fig 2 SUPPLY AIR PRESSURE Setting Set the Air pressure at the Air Check point on Fi...

Page 8: ...ler 1 Insert the Bur Do not exceed maximum rotation speed for any item 2 Check that there is no abnormal rotation vibration noise and overheating 3 If there are no immediate apparent abnormalities con...

Page 9: ...approximately 2 3 seconds 1 Mount the Spray Nozzle into the Spray Port Fig 8 2 Remove the bur from the handpiece 3 Insert the Spray Nozzle in rear of the handpiece and spray for approximately 2 3 seco...

Page 10: ...ld the spray can upright 5 5 Sterilization Autoclave sterilization is recommended Sterilization is required rst time you use and after each patient as noted below This handpiece and motor can be autoc...

Page 11: ...ubrication 5 5 Sterilization Autoclave Procedure 1 Clean and lubricate the handpiece 2 Insert into an autoclave pouch Seal the pouch 3 Autoclavable up to max 135 C ex Autoclave for 3 min at 134 C 4 Ke...

Page 12: ...der Code P1001250 SX SU03 Standard Cartridge Order Code P1002250 CAUTION Do not use a pointed tool or sharp edged to clean the Cellular Glass Optic Rod They may damage the mirror nish and reduce the l...

Page 13: ...t autoclave the handpiece even when it is bagged together with other instruments from which chemicals are not suf ciently cleaned This is to avoid discoloration and or damage to the plating of the han...

Page 14: ...ion 5 Align the Alignment Pin and insert Cartridge Fig 8 6 Turn the Wrench clockwise and mount the Head Cap Fig 9 Replacement Cartridge SX MU03 Mini Cartridge Order Code P1001250 SX SU03 Standard Cart...

Page 15: ...ot suf ciently cleaned This is to avoid discoloration and or damage to the plating of the handpiece In case the sterilizer chamber temperature could exceed 135 C during the dry cycle skip the dry cycl...

Page 16: ...ht into the spray ports with caution Do not forcibly insert the wire into the port Damaged or disoriented port could cause the spray diverge or directed away from the bur and the cooling ef ciency det...

Page 17: ...llows 1 Remove the bur from the chuck Refer to 4 Insertion Removal the Bur and 6 Replacing the Cartridge Fig 10 O Ring O Ring CAUTION Tighten the head cap after changing the cartridge If attachment is...

Page 18: ...a fraise venait se d tacher dans la cavit buccale du patient ce dernier pourrait l avaler accidentellement et se blesser Cette pi ce main pr sente une dynamo int gr e L air consomm g n re de l lectric...

Page 19: ...tra ner un dysfonctionnement de la pi ce main bruits ou vibrations anormaux Les dommages les dysfonctionnements ou les accidents ne sont pas couverts par notre garantie dans ces cas l N essuyez pas ne...

Page 20: ...ect e de la pi ce main Veillez assurer une arriv e d air s che et propre L humidit et les d bris pr sents dans l arriv e d air peuvent entra ner un dysfonctionnement du moteur 1 Sp cifications Utilisa...

Page 21: ...a vibration et la surchauffe avant de l utiliser Le traitement doit tre r alis avec suf samment d eau d irrigation eau pulv ris e ATTENTION N utilisez pas de fraises non standards Le diam tre ISO stan...

Page 22: ...apparentes imm diates continuez de faire tourner la pi ce main pendant environ 1 minute V ri ez ensuite s il n y a pas de chauffe anormale sur la surface externe 4 Vous pouvez commencer utiliser la pi...

Page 23: ...ain M500LG M4 B2 M600LG M4 B2 M500LG QD M600LG QD 2 3 Montage 1 Ins rez le raccord dans le connecteur du tuyau 2 Serrez l embout du tuyau 3 Tirez l anneau du connecteur 4 Poussez la pi ce main dans le...

Page 24: ...main et contactez votre revendeur agr 1 Ins rez la fraise ne d passez pas la vitesse de rotation maximale pour chaque pi ce 2 V ri ez qu il n y ait pas de rotation de vibration de bruit ou de surchau...

Page 25: ...aut e du spray dans le raccord pr vu cet effet Fig 7 2 Otez la fraise de la pi ce main 3 Ins rez l embout t te biseaut e du spray dans l ori ce d arriv e d air de la pi ce main et pulv risez pendant e...

Page 26: ...2 secondes Tenez le spray droit 5 5 St rilisation La st rilisation par autoclave est recommand e La st rilisation est recommand e avant la premi re utilisation et apr s chaque traitement comme indiqu...

Page 27: ...ion 5 5 St rilisation Proc dure d autoclave 1 Nettoyez et lubri ez la pi ce main 2 Ins rez dans un sachet pour autoclave Scellez le sachet 3 St rilisez en autoclave 135 C max par ex st rilisez par aut...

Page 28: ...he R f rence P1001250 SX SU03 Cartouche standard R f rence P1002250 ATTENTION N utilisez pas d outil pointu pour nettoyer le barreau de quartz Cela pourrait endommager la nition du barreau de quartz e...

Page 29: ...dans un sac avec d autres instruments dont les r sidus de produits chimiques n auraient pas t suf samment nettoy s Ceci a pour but d viter la d coloration et ou l endommagement du rev tement de la pi...

Page 30: ...a lubri cation 5 Alignez l ergot d alignement et ins rez la cartouche Fig 9 6 Tournez la cl dans le sens horaire et montez le bouchon de t te Fig 9 Cartouche de rechange SX MU03 Mini cartouche R f ren...

Page 31: ...loration et ou l endommagement du rev tement de la pi ce main Si la temp rature de la chambre de st rilisation peut exc der 135 C pendant le cycle sec passez le cycle de s chage Ne chauffez et ne refr...

Page 32: ...pas le l dans le port Un port endommag ou d plac pourrait modi er la trajectoire du spray ou l loigner de la fraise et r duire ainsi l ef cacit du refroidissement 3 Apr s utilisation nettoyez le l de...

Page 33: ...us au point 4 Insertion Retrait de la fraise et au point 6 Remplacement de la cartouche de turbine Fig 10 Joint Joint ATTENTION Serrez le bouchon de t te apr s avoir chang la cartouche Si le bouchon d...

Page 34: ...esa cayese en la cavidad oral del paciente ste podr a tragarla por error y lesionarse Esta pieza de mano tiene una dinamo integrada El consumo de aire genera electricidad que ilumina el LED Por lo tan...

Page 35: ...arato en agua muy cida o en soluciones esterilizantes Utilice gafas protectoras y m scara para su seguridad mientras est trabajando con esta pieza de mano El usuario es responsable del funcionamiento...

Page 36: ...hos en el suministro de aire pueden ocasionar que el motor funcione incorrectamente 1 Caracter sticas t cnicas Uso previsto Esta pieza de mano est dise ada para un uso cl nico dental llevado a cabo ni...

Page 37: ...guna de las fresas de abajo ya que esto podr a causar la ruptura o desgaste de la fresa Fresa doblada deformada desgastada oxidada o de ciente Fresa que muestra una raja en el borde o en el eje Fresa...

Page 38: ...nuto A continuaci n compruebe que no hay calentamiento anormal en la super cie externa 4 Puede comenzar a utilizar la pieza de mano si no se ha producido ninguna de las anomal as anteriormente descrit...

Page 39: ...iento en el conector de tubo 2 Ajuste la tuerca de tubo 3 Tire del anillo de conexi n 4 Presione la pieza de mano en el acoplamiento 2 4 Disconnect Deslice hacia atr s el anillo de conexi n tal y como...

Page 40: ...nserte la fresa No exceda la velocidad de rotaci n m xima para ning n art culo 2 Compruebe que no hay una rotaci n anormal vibraciones ruido y sobrecalentamiento 3 Si no hay anomal as aparentes de inm...

Page 41: ...erte la boquilla de pulverizador del cabezal en echa en el tubo de conducci n de aire de la pieza de mano y pulverice durante unos dos a tres segundos 1 Monte la boquilla de pulverizador en el conduct...

Page 42: ...r hacia arriba 5 5 Esterilizaci n Se recomienda esterilizar con autoclave Es necesaria una esterilizaci n antes del primer uso y tras cada tratamiento tal y como se indica a continuaci n La pieza de m...

Page 43: ...n 5 5 Esterilizaci n Procedimiento con autoclave 1 Limpie y lubrique la pieza de mano 2 Introducir en un estuche de autoclave Selle el estuche 3 Esterilice con autoclave hasta 135 C m ximo ej Esterili...

Page 44: ...l Fig 9 Sustituci n de cartucho Mini cartucho SX MU03 C digo de pedido P1001250 Cartucho est ndar SX SU03 C digo de pedido P1002250 PRECAUCI N No utilice una herramienta puntiaguda o cortante para lim...

Page 45: ...n otros instrumentos cuyos productos qu micos no se hayan limpiado de forma su ciente Esto evita la decoloraci n y o el da o del revestimiento de la pieza de mano En caso de que la temperatura de la c...

Page 46: ...tirar el cartucho 4 Limpie el interior del cabezal Retire el aceite extra despu s de la lubricaci n 5 Alinee la patilla de alineaci n e inserte el cartucho Fig 9 6 Gire la llave en el sentido de las a...

Page 47: ...En caso de que la temperatura de la c mara esterilizadora superase los 135 C durante el ciclo de secado s ltese el ciclo de secado No caliente ni enfr e la pieza de mano r pidamente Los cambios r pid...

Page 48: ...auci n No introduzca el cable por la fuerza en el conducto Un conducto da ado o descolocado puede hacer que el pulverizador se separe o se aleje de la fresa reduciendo as la e cacia del enfriado 3 Des...

Page 49: ...de la fresa y 6 Cambio del cartucho de turbina 2 Saque el cable de limpieza de la parte trasera del soporte del cepillo Introduzca el cable directamente en Fig 10 Junta Junta PRECAUCI N Apriete el ta...

Page 50: ...da ada puede causar un escape de aire o de agua Si se encuentra agua en el aire de salida o sale agua en un punto de acoplamiento sustituya a las cuatro juntas hist ricas tal y como se muestra la gura...

Page 51: ...2017 03 17 02 S...

Reviews: