PDES-01DM User Manual 

version 1.3 

 

page 9 

(3)

 

EXHAUST connector 

Connector for connecting a “retain” pressure source or for EXHAUST. If no “retain” pressure 
is connected, the system is exhausted as long as no pressure is applied to the pipette. 
 

(4)

 

IN connector 

Connector for connecting the filtered, dry and oil-free gas source. The gas source must be 
filtered by a 5 µm filter element. 

Important

: Do not connect sources of OXYGEN or COMBUSTIBLE gases!! 

 

Note

: npi recommends an input pressure of 1 bar / 14 psi more than the maximum output 

pressure. 
 

(5)

 

Pressure regulator 

Rotary knob for regulating the pressure of the ejection channel. The pressure is indicated by the 
manometer (#

1

). 

 

(6)

 

OUT connector 

Connector for connecting the injection micropipette, or for connecting the tubing to external 
valve, e.g. the micropipette holder 

micro

JECT. The EJECT pressure is supplied to this 

connector. 
 

(7)

 

EXTERNAL VALVE connector (optional) 

Connector for the electrical connection of the external valve. 
 

(8)

 

TIMING MONITOR 

BNC connector providing a TTL signal that indicates the status of the regulated injection valve 
(HI = valve open, LO = valve closed). 
 

(9)

 

PRESSURE MONITOR 

BNC connector monitoring the pressure output; sensitivity: 1 V / bar or 0.1 V / psi, respectively. 
 

(10)

 

REMOTE CONTROL connector 

BNC connector for connecting a remote control, e.g. a pushbutton. 
 

(11)

 

TTL INPUT 

BNC connector for connecting a TTL signal for EXTERN operation or for using the internal 
timer in TIMER operating mode (see also #

12

). 

 

Summary of Contents for PDES-01DM

Page 1: ...PDES 01DM PNEUMATIC DRUG EJECTION MODULE WITH ANALOG TIMER FOR EPMS 07 SYSTEMS VERSION 1 3 npi 2016 npi electronic GmbH Bauhofring 16 D 71732 Tamm Germany Phone 49 0 7141 0730230 Fax 49 0 7141 9730240...

Page 2: ...Grounding 6 EPMS 07 EPMS 03 6 EPMS E 07 6 2 8 Technical Data 6 EPMS 07 EPMS E 07 and EPMS H 07 6 EPMS 07 and EPMS H 07 6 EPMS E 07 6 EPMS 03 6 3 PDES System 7 3 1 System Description 7 3 2 PDES 01AM C...

Page 3: ...LECTRICITY Electronic equipment is sensitive to static discharges Some devices such as sensor inputs are equipped with very sensitive FET amplifiers which can be damaged by electrostatic charge and mu...

Page 4: ...ront covers Figure 1 Left front view of empty EPMS 07 housing In order to avoid induction of electromagnetic noise the power supply unit the power switch and the fuse are located at the rear of the ho...

Page 5: ...ernal power supply PWR 03D is capable of driving up to 3 EPMS E housings Each housing is connected by a 6 pole cable from one of three connectors on the front panel of the PWR 03D to the rear panel of...

Page 6: ...d to the SYSTEM GROUND SIGNAL GROUND on the rear panel of the instrument see Figure 4 The chassis can be linked to PROTECTIVE EARTH by connecting it to the PWR 03D with the supplied 6 pole cable and b...

Page 7: ...front panel If the internal timer is used the eject time is preset by a control and the time base can be selected by a toggle switch The operational status is indicated by a red green LED The pressure...

Page 8: ...utton External valve slow switch time 10 ms 3 3 Description of the Front Panel Figure 5 PDES 01DM front panel view the numbers are related to those in the text below 1 Manometer Digital display for th...

Page 9: ...re is indicated by the manometer 1 6 OUT connector Connector for connecting the injection micropipette or for connecting the tubing to external valve e g the micropipette holder microJECT The EJECT pr...

Page 10: ...x1s x0 01s x0 01s switch Switch to set the time base of the timer 14 range x10 s x1 s x0 1 s or x0 01 s The total ejection time is the reading of this switch multiplied by the reading of the timer po...

Page 11: ...e The valve which is controlled by the PDES is located on the remote valve plate see Figure 6 Tubing and cable of the external valves are connected at the front panel of the PDES system refer to chapt...

Page 12: ...y not connected Figure 7 microJECT 1 pipette connector 2 connector for EJECT pressure 3 connector for EXHAUST pressure optional 4 electrical connection Important In experiments with high pressures 1 b...

Page 13: ...connect a pushbutton to the REMOTE CONTROL connector 10 Figure 5 at the front panel The LED is green when no ejection occurs and red when the pressure is applied to the pipette Select the operation m...

Page 14: ...PDES 01DM User Manual version 1 3 page 14 Figure 9 Operation modes of the PDES 01DM...

Page 15: ...iol 587 5859 5877 Lin B J Chen T W Schild D 2007 Cell type specific relationships between spiking and Ca2 i in neurons of the Xenopus tadpole olfactory bulb Journal of Physiology 582 163 175 Marandi N...

Page 16: ...kidney 293 cells and Xenopus Laevis oocytes J Neurosci 19 7426 7433 Fast drug application with microJECT EHM Courjaret R Troger M Deitmer J W 2009 Suppression of GABA input by A1 adenosine receptor a...

Page 17: ...red light valve is closed in EXTERN TIMER CONT operation mode green light valve is open Off no light valve is closed Mode selection three position toggle switch TIMER CONT EXTERN Minimum valve switch...

Reviews: