3. PA-2S Amplifier Module 

 

3.1. PA-2S 

Components 

 
The following items are shipped with the PA-2S system: 
 

 

Amplifier module for the EPMS-07 system 

 

User manual 

 

3.2. System 

Description 

 
The PA-2S amplifier is a plug-in unit for the npi EPMS-07 modular system. It is designed to 
amplify small bio-electrical signals. Two channels are available and can be used 
independently from each other in AC or DC mode. DC offsets can be compensated either by 
the using the OFFSET controls (coarse and fine) or they can be eliminated using the AC 
position of the input switches. The inputs and outputs are BNC connectors. 
 
Input and output configurations are selectable. The PA-2S can be connected internally to 
other modules of the EPMS-07 system, e.g. to one channel of an EXT-08, to a DPA-2F or to 
an INT-20M module. Both INPUT and OUTPUT signals can be linked internally to other 
EPMS modules. The LC Display shows which amplifier / channel is used (see below) and 
indicates if the linear region of the amplifier is exceeded.  
 
The input impedance is 1 M

 (other impedances are available on request). The GAIN range 

is x10 to x1000 and can be set by a rotary switch (10-20-50-100-200-500-1000 switch). The 
OUTPUT is filtered by an 8 kHz lowpass filter. 

Summary of Contents for PA-2S

Page 1: ...PA 2S TWO CHANNEL AMPLIFIER MODULE FOR EPMS SYSTEMS with 8 kHz Lowpass Filter VERSION 1 5 npi 2006 npi electronic GmbH Hauptstrasse 96 D 71732 Tamm Germany Phone 49 0 7141 601534 Fax 49 0 7141 601266...

Page 2: ...Table of Contents About this Manual 3 1 Safety Regulations 4 2 EPMS 07 Modular Plug In System 5 2 1 Components of the EPMS 07 Housing 5 2 2 General System Description Operation 5 2 3 System Grounding...

Page 3: ...mall bio electrical signals please read the manual completely The experienced user should read at least chapter 3 Important Please read chapter 1 carefully It contains general information about the sa...

Page 4: ...nly at 115 230 Volt 60 50 Hz AC Please check for appropriate line voltage before connecting any system to mains Always use a three wire line cord and a mains power plug with a protection contact conne...

Page 5: ...plug in units are connected to power by a bus at the rear panel The plug in units must be kept in position by four screws M 2 5 x 10 The screws are important not only for mechanical stability but als...

Page 6: ...l ground is used on the BNC connectors or GROUND plugs of the modules that are inserted into the EPMS 07 housing The internal ground and mains ground protective earth can be connected by a wire using...

Page 7: ...be compensated either by the using the OFFSET controls coarse and fine or they can be eliminated using the AC position of the input switches The inputs and outputs are BNC connectors Input and output...

Page 8: ...element in lowercase letters Then a short description of the element is given The PA 2S has two channels with identical operation elements that are numbered and described only once 1 AMP No LED The P...

Page 9: ...1 Hz DC the input signal is DC coupled 0 the input of the PA 2S is grounded possible offsets are not affected 6 INPUT BNC BNC connector to connect the input signal Input impedance 1 M other impedance...

Page 10: ...rk Ogden D ed 1992 Microelectrode Techniques The Plymouth Workshop Handbook Second Edition The Company of Biologists Ltd Cambridge Windhorst U Johansson H eds 1999 Modern Techniques in Neuroscience Re...

Reviews: