background image

Tetrode Amplifier User Manual 

_______________________________________________________________________________________________________________

 

___________________________________________________________________________ 
version 1.1 

 

 

page 7 

3. Headstage 

 

 

 

Figure 4: one headstage of the EXT-T2 

 
Headstage Elements 
 

P

EL

 

Connector for the tetrode (see also Figure 5) 

REF 

Not connected 

GND 

Ground connector 

 
 

 

Ground connector (GND) 

Channel 1 connector (CH0) 

Channel 3 connector (CH2) 

Channel 4 connector (CH3) 

Channel 2 connector (CH1) 

 

Figure 5: tetrode connector 

 

4. Technical Data 

 
Input resistance: >10

12

 

, range ±1 V 

Output: 

gain (x200, x500, x1k), selectable by DIL switch 

 

Output range: ±12 V into 1 k

 / ±1 V into 50 

 load 

Low pass filter: 

500, 2k, 8k Hz, selectable by DIL switch 

High pass filter: 

DC, 0.1, 0.3, 1, selectable by DIL switch 

Power requirements: 

12…18 V AC, 1 A min.; wall power supply 

Dimensions: 

19” rackmount cabinet, 19” (483 mm), 10” (250 mm), 1.75” 
(44 mm) 

Headstage Size: 

70 x 26 x 26 mm 

Holding Bar: 

length: 150 mm; diameter: 8 mm. 

Summary of Contents for EXT-T2

Page 1: ...ON FOR THE 8 CHANNEL EXTRACELLULAR AMPLIFIER SYSTEM FOR 2 TETRODES VERSION 1 1 npi 2010 npi electronic GmbH Hauptstrasse 96 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 support n...

Page 2: ..._______________________ ___________________________________________________________________________ version 1 1 page 2 Table of Contents 1 Safety Regulations 3 2 EXT T2 Amplifier Filter 4 2 1 System D...

Page 3: ...vice is to be operated only at 115 230 Volt 60 50 Hz AC Please check for appropriate line voltage before connecting any system to mains Always use a three wire line cord and a mains power plug with a...

Page 4: ...s single ended configuration against GND 2 2 Description of the Front Panel Figure 1 front panel view of EXT T2 1 HEADSTAGE connector Connector for the HEADSTAGE of TETRODE A 2 OUTPUT 0 OUTPUT 3 conne...

Page 5: ...Hz gain 500 LP 2 kHz OFF ON ON ON OFF ON OFF HP 0 1 Hz gain 1000 LP 2 kHz OFF ON ON OFF ON ON ON HP 0 1 Hz gain 200 LP 500 Hz OFF ON ON ON ON ON ON HP 0 1 Hz gain 500 LP 500 Hz OFF ON ON ON OFF OFF O...

Page 6: ...DIL switch and jumper 2 3 Description of the Rear Panel Figure 3 rear panel view of EXT T2 1 GROUND This connector is linked to the internal system ground which has no connection to the 19 cabinet CH...

Page 7: ...Ground connector GND 2 Channel 1 connector CH0 3 Channel 3 connector CH2 4 Channel 4 connector CH3 5 Channel 2 connector CH1 Figure 5 tetrode connector 4 Technical Data Input resistance 1012 range 1 V...

Reviews: