ELPORATOR - User Manual
_________________________________________________________________________________________________________________
___________________________________________________________________________
version 1.0
page 8
3.3. Description of the Rear Panel
Figure 2: ELectroPORATOR rear panel view
MONITORING OUTPUT
connectors
(
1
) POTENTIAL 0.1 V/V connector
(
2
) CURRENT 100/10/1 mV/µA connector (scaling dependent on current scaling)
(
3
) RESISTANCE 1 V/MΩ connector
(
4
) POWER 10 V/W connector
(
5
) REMOTE connector
BNC connector with active low logic. A passive footswitch or a remote button can be connected
here.
(
6
) GROUND connector
Banana plug providing internal ground (see below).
(
7
) CHASSIS connector
Banana plug providing mains ground (see below).
(
8
) FUSE holder
Holder for the line fuse. For changing the fuse rotate the holder counter-clockwise using a screw
driver.
(
9
) LINE SELECT switch
Switch for selecting the line voltage. Switch to the right for 230 V AC, to the left for 115 V
AC. The selected voltage is indicated on the switch.
Caution
: Before turning on the instrument, make sure that the correct line voltage is selected.