DPA-2FX User Manual 

___________________________________________________________________________ 

___________________________________________________________________________ 
version 1.8 

 

 

page 5 

 

 

Figure 1: PWR-03D front panel view  

Figure 2: PWR-03D rear panel view 

 

Note

: This power supply is intended to be used with npi EPMS-E systems only. 

 

2.5. System 

Grounding 

EPMS-07 

The 19" cabinet is grounded by the power cable through the ground pin of the mains 
connector (= protective earth). In order to avoid ground loops the internal ground is isolated 
from the protective earth. The internal ground is used on the BNC connectors or GROUND 
plugs of the modules that are inserted into the EPMS-07 housing. The internal ground and 
mains ground (= protective earth) can be connected by a wire using the ground plugs on the 
rear panel of the instrument. It is not possible to predict whether measurements will be less or 
more noisy with the internal ground and mains ground connected. We recommend that you 
try both arrangements to determine the best configuration. 

EPMS-E-07 

The 19" cabinet is connected to the PROTECTIVE EARTH connector at the 
rear panel. The chassis is linked to protective earth only if the PWR-03D is 
connected. It can be connected also to the SYSTEM GROUND (SIGNAL 
GROUND) on the rear panel of the instrument (see Figure 3).  
 
 
 

Important:

: Always adhere to the appropriate safety measures.  

 
 
 
 

Figure 3: Rear panel connectors of the EPMS-E-07 

2.6. Technical 

Data 

19” rackmount cabinet, for up to 7 plug-in units 
Dimensions: 

3U high (1U=1 3/4” = 44.45 mm), 254 mm deep 

EPMS-07 

Power supply:  115/230 V AC, 60/50 Hz, fuse 2 A / 1 A slow, 45-60 W 

EPMS-E-07 

External power supply (for EPMS-E):  

115/230 V AC, 60/50 Hz, fuse 1.6/0.8 A, slow 

Dimensions of External power supply:  

(W x D x H) 225 mm x 210 mm x 85 mm 

Summary of Contents for DPA-2FX

Page 1: ...FOR THE DPA 2FX DIFFERENTIAL AMPLIFIER FILTER MODULE FOR EPMS SYSTEMS VERSION 1 8 npi 2014 npi electronic GmbH Bauhofring 16 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 support...

Page 2: ...PMS 07 Modular Plug In System 4 2 1 General System Description Operation 4 2 2 EPMS 07 Housing 4 2 3 EPMS E 07 Housing 4 2 4 PWR 03D 4 2 5 System Grounding 5 EPMS 07 5 EPMS E 07 5 2 6 Technical Data 5...

Page 3: ...only at 115 230 Volt 60 50 Hz AC Please check for appropriate line voltage before connecting any system to mains Always use a three wire line cord and a mains power plug with a protection contact conn...

Page 4: ...n of electromagnetic noise the power supply unit the power switch and the fuse are located at the rear of the housing 2 3 EPMS E 07 Housing The following items are shipped with the EPMS E 07 housing E...

Page 5: ...ound plugs on the rear panel of the instrument It is not possible to predict whether measurements will be less or more noisy with the internal ground and mains ground connected We recommend that you t...

Page 6: ...t are BNC connectors The input impedance is 10 M Thus the module can be used for direct recording of bioelectric signals e g from peripheral nerves Using the differential input configuration common mo...

Page 7: ...__________________________________________________________ ___________________________________________________________________________ version 1 8 page 7 3 4 Description of the Front Panel and Operati...

Page 8: ...ge switch 3 position switch for selecting the OFFSET range 0 1 V 1 V or 10 V 2 OFFSET potentiometer 10 turn potentiometer for compensating for a DC OFFSET Note The overall OFFSET compensation is the p...

Page 9: ...lter with bandpass characteristic can be implemented easily 6 OVER OVER LEDs LEDs indicating that the DPA 2FX exceeds it s linear range 10 V 7 OUT connector BNC connector providing the conditioned sig...

Page 10: ...ptors mediate expression of LTP in slices of rat visual cortex Eur J Neurosci 15 1641 1645 Huemmeke M Eysel U T Mittmann T 2004 Lesion induced enhancement of LTP in rat visual cortex is mediated by NM...

Page 11: ...0 1 1 V or 10 V set by toggle switch HIGHPASS Filter single pole attenuation 6 dB octave corner frequencies Hz 3 30 100 300 400 500 600 800 1k 3k 10k LOWPASS Filter single pole attenuation 6 dB octave...

Reviews: