background image

TxBlock-USB 

NOVUS AUTOMATION 

4/5 

CONFIGURING TXBLOCK-USB WITH TXCONFIG  APPLICATION 

FOR ANDROID: 
Once the connection between the smartphone and the TxBlock-

USB is made, the device will send the following message: 

 

Fig. 13 – USB device connected 

For the correct operation of the application, the option "Open 

TxConfig when this USB device is connected?" must be checked and 

then click the OK button. 
After that, it will automatically run the TxConfig application, which 

has already been previously installed on the smartphone, and will 

display the following message: 

 

Fig. 14 – Reading device 

The application will automatically switch to the configuration screen 

(see Fig. 11), where it will be possible to configure the TxBlock-USB

OPERATION 

The sensor offset can be changed through the 

TxConfig II

 software 

or the 

TxConfig 

app. 

The USB cable may be connected to the transmitter without causing 

any measurement errors.  See item 

Zero Correction

  in the chapter 

CONFIGURATION of this manual. 
The uer must choose the most suitable sensor and range to the 

process. The chosen range must not exceed the maximum range of 

measurement defined for the sensor and should not be smaller than 

the minimum range for the same sensor. 
It is important to note that the transmitter accuracy is based on the 

maximum range of the sensor used, even when a narrower range is 

programmed.  
Example

 

The Pt100 sensor in the range 0 to 100 °C and accuracy of 

0.12 %, the maximum error will be 1.02 °C (0.12 % de 850 °C). 

 

The Pt100 sensor in the range 500 to 600 °C and accuracy of 0.19 %, 

the maximum error will be 1.61 °C (0.19 % of the 850 °C). 

Note: When measurements are made at the transmitter, see if the 

Pt100 excitation current required by the calibrator is compatible with 

the Pt100 excitation current used in the transmitter: 0.8 mA. 

SPECIFICATIONS 

Sensor input: User defined. The supported sensors are listed in 

Table 1, along with their maximum ranges. 

Thermocouples: Types J, K, R, S, T, N, E and B accoding to 

IEC 60584 (ITS-90). Impedance >> 1 M

 

Pt100

Type  3-wire, Excitation 0.8 mA, 

α

= 0.00385, 

according IEC 60751 (ITS-90). 

 

For 2-wire sensors, tie terminals 3 and 4 

together. 

Pt1000

Type 3-wire, Excitation 0.65 mA, 

α

= 0.00385, 

according IEC 60751 (ITS-90). 

 

For 2-wire sensors, tie terminals 3 and 4 

together. 

NTC R

25°C

:  

10 k

 ±1 %, B

25/85

 = 3435 

Voltage:  

0 to 50 mVdc. Impedance >> 1 M

 

Sensor Type 

Maximum Measurement 

Range 

Minimum Measurement 

Range 

Voltage 

0 to 50 mV 

5 mV 

Thermocouple K 

-150 to 1370 °C 

100 °C 

Thermocouple J 

-100 to 760 °C 

100 °C 

Thermocouple R 

-50 to 1760 °C 

400 °C 

Thermocouple S 

-50 to 1760 °C 

400 °C 

Thermocouple T 

-160 to 400 °C 

100 °C 

Thermocouple N 

-270 to 1300 °C 

100 °C 

Thermocouple E 

-90 to 720 °C 

100 °C 

Thermocouple B 

500 to 1820 °C 

400 °C 

Pt100 

-200 to 650 °C 

40 °C 

Pt1000 

-200 to 650 °C 

40 °C 

NTC 

-30 to 120°C 

40 °C 

Table 1 – Sensors accepted by the transmitter 

Switch-on delay: < 2.5 s. The accuracy is only guaranteed after 15 

min. 
Terms of reference
: Ambient: 25 ° C; voltage: 24 Vdc, load: 250 

settling time: 10 minutes. 
Temperature Effect
: < ±0.16 % / 25 °C 
Response time
: Typical 1.6 s. 
Maximum voltage allowed at input terminals no sensor: 3 V. 
RTD current: 800 

µ

A. 

RTD cable resistance effect: 0.005 °C / 

 

Maximum allowable cable resistance for RTD: 25 

Sensor Type 

Typical 

Accuracy 

Minimun 

Accuracy 

Pt100 / Pt1000  

(-150 to 400 °C) 

0.10 % 

0.12 % 

Pt100 / Pt1000  

(-200 to 650 °C) 

0.13 % 

0.19 %  

mV, K, J, T, E, N, R, S, B 

0.1 % (*) 

0.15 % (*) 

NTC 

0.3 °C 

0.7 °C 

Table 2 – Calibration error, percentage of the full measurement range

 

(*) Add cold junction compensation: < ± 1 °C. 
Power supply influence: 0.006 % / V typical (percentage of the full 

measure range). 
Output: 4-20 mA or 20-4 mA current, 2-wired; linear in relation to the 

temperature measurement by the selected sensor. 
Output Resolution: 2 

µ

A. 

Power supply: 10 to 35 Vdc, across the transmitter; 
Maximum load (RL):  RL (max.) = (Vdc – 10) / 0.02 [

]

 

 

 Where: Vdc= Power supply voltage (10-35 Vdc) 

Operating Temperature: -40 to 85 °C 
Humidity: 0 to 90 % RH 
Electromagnetic Compatibility: EN 61326-1:2006 

Summary of Contents for TxBlock-USB

Page 1: ...l 0 8 Nm RECOMMENDATIONS FOR INSTALLATION Sensor signals conductors must go through the plant system separate from power leads loop if possible in grounded conduits The instruments must be powered from the instrumentation power supply circuit In control and monitoring applications is essential to consider what can happen when any part of the system fails It is recommended the use of suppressors in...

Page 2: ...install execute the TxConfigIISetup exe file and follow the instructions The setup application is only available for Android smartphones and also can be downloaded for free in the Google Play Store To configure the device through the application an On The Go OTG adapter is required The procedure for using this adapter will be explained throughout this manual Fig 8 USB cable connection During the s...

Page 3: ...le typically used for refrigeration Maximum Output current goes to 22 0 mA up scale typically used for heating 5 Zero Correction It corrects small deviations presented in the transmitter output for example when the sensor is replaced 6 Send Configuration It applies the new setup Once sent the setup will be immediately adopted by the transmitter 7 Read Configuration Reads the current setup in the t...

Page 4: ...ed sensors are listed in Table 1 along with their maximum ranges Thermocouples Types J K R S T N E and B accoding to IEC 60584 ITS 90 Impedance 1 MΩ Pt100 Type 3 wire Excitation 0 8 mA α 0 00385 according IEC 60751 ITS 90 For 2 wire sensors tie terminals 3 and 4 together Pt1000 Type 3 wire Excitation 0 65 mA α 0 00385 according IEC 60751 ITS 90 For 2 wire sensors tie terminals 3 and 4 together NTC...

Page 5: ...device and its alarms are not intended to protect against device failures Independent safety devices should be always provided if personnel or property are at risk Device performance and specifications may be affected by its environment and installation It s user s responsibility to assure proper grounding shielding cable routing and electrical noise filtering in accordance with local regulations ...

Reviews: