Novus N1540-485 Instruction Manual Download Page 2

N1540 

NOVUS AUTOMATION

 

2

 

difk

 

Alarm of the maximum differential value. It triggers when 

the PV value is 

above

 the point defined by: 

ALrF+SPA1

 (using alarm 1 as an example). 

 

ALrF 

PV 

ALrF + SPA1 

 

 

ALrF 

PV 

ALrF + SPA1 

 

SPA1 positive 

SPA1 negative 

ierr

 

Sensor  breaks  alarm. It is activated when the Input 

presents problems such as interrupted sensor, bad 

connection, etc. 

Table 2 –

 

 

Alarm functions 

Note

: The figures are also valid for Alarm 2 (SPA2). 

Important note: 

Alarms configured with the 

ki

dif

  and 

dif.k

 

functions also trigger their associated output when a sensor fault is 

identified and signaled by the digital panel meter. A relay output, for 

example, configured to act as a High Alarm (

ki

), will operate when 

the SPAL value is exceeded and when the sensor connected to the 

input is broken. 

ALARM BLOCKING INITIAL 

The 

Initial

 

Blocking 

option inhibits the alarm from being recognized 

if an alarm condition is present in the process when the digital panel 

meter  is first energized. The alarm will be enabled only after the 

occurrence of no alarm condition. 
The initial blocking is useful, for example, when one of the alarms is 

set up as a minimum value alarm, which may cause the activation of 

the alarm soon upon the process start-up; an occurrence that may be 

undesirable in many cases. 
The initial blocking is not valid for the 

ierr

 (Sensor Break) function. 

OFFSET 

Allows fine adjustments to the PV indication, correcting measurement 

errors that appear, for example, after the replacement of the 

temperature sensor. 

MAXIMUM AND MINIMUM 

The  digital panel meter  memorizes the measured maximum and 

minimum values (peak and valley). The operator can observe these 

extreme values at any time. These two values are shown when 

pressing the 

F1

 (maximum) and 

F2 

(minimum) keys. Pressing both 

keys simultaneously will clear the memory for a new peak and valley 

detection. 
To clear the stored values and begin a new cycle of monitoring 

endpoints, just press the 

F1

  and 

F2

  keys 

simultaneously

. When 

turn off, the digital panel meter this information is not saved. 

CUSTOM LINEARIZATION 

This feature allows accurate measurement of the input signals with 

non-linear features. 
Linearization consists in dividing the calibration curve of the input 

signal into segments of variable gain. Each segment  consists of a 

start and an end. For each input value (

Inp

.xx), a respective output 

indication (

ovP

.xx) is defined. 

The input signal must present an always-crescent. 
Applied to 0-20 mA, 4-20 mA, 0-50 mV, 0-5 V and 0-10 V input types. 

24 VDC AUXILIARY VOLTAGE SOURCE  

The standard version of the 

N1540

  provides an auxiliary power 

supply for exciting field transmitters (terminals 11 and 13 on the rear 

panel). 

SERIAL COMMUNICATION 

Follows a description of the usual communication registers. For full 

documentation,  download the 

Registers Table N1540 for Serial 

Communication

 on our web site: 

www.novusautomation.com

 

Figure 1 –

 

Serial communication 

RS485 INTERFACE: ELECTRICAL CONNECTION 

The RS485 signals are: 

D1  D  D +  B 

Bidirectional data line.

 

Terminal 17 

D0 

𝑫𝑫

  D -  A 

Inverted bidirectional data line.

 

Terminal 18 

Optional connection that 

improves the communication 

performance.

 

Terminal 19 

GND 

Table 3 –

 

 RS485 

USB INTERFACE 

The USB interface is used to CONFIGURE, MONITOR or UPDATE 

the controller FIRMWARE. The user should use 

QuickTune

 

software, which offers features to create, view, save,  and open 

settings from the device or files on the computer. The tool for saving 

and opening configurations in files allows the user to transfer settings 

between devices and perform backup copies.  
For specific models, 

QuickTune

  allows to update the firmware 

(internal software) of the controller via the USB interface. 
For MONITORING purposes, the user can use any supervisory 

software (SCADA) or laboratory software that supports the Modbus 

RTU communication over a serial communication port. When 

connected to  a computer USB, the controller is recognized as a 

conventional serial port (COM x).  
The user must use 

QuickTune

  software or consult the DEVICE 

MANAGER on the Windows Control Panel to identify the COM port 

assigned to the controller. 
The user should consult the mapping of the Modbus memory in the 

controller communication manual and the documentation of the 

supervision software to start the MONITORING process. 
Follow the procedure below to use the USB communication of the 

device: 

1.

 

Download 

QuickTune 

software from our website and install it on 

the computer. The USB drivers necessary for operating the com-

munication will be installed with the software. 

2.

 

Connect the USB cable between the device and the computer. 

The controller does not have to be connected to a power supply. 

The USB will provide enough power to operate the communica-

tion (other device functions may not operate). 

3.

 

Run the 

QuickTune 

software, configure the communication and 

start the device recognition. 

Summary of Contents for N1540-485

Page 1: ...f equipment that require the same configuration FEATURES SIGNAL INPUT TYPE The input type to be used by the N1540 is defined in the equipment configuration Table 1 presents the input options available...

Page 2: ...t value Inp xx a respective output indication ovP xx is defined The input signal must present an always crescent Applied to 0 20 mA 4 20 mA 0 50 mV 0 5 V and 0 10 V input types 24 VDC AUXILIARY VOLTAG...

Page 3: ...mA OUT Note RS485 Communication and PV Retransmission features are mutually exclusive INSTALLATION CONNECTIONS The digital panel meter is meant for panel mounting The sequence of steps is Prepare a cu...

Page 4: ...PV PV Display Indication The value of the measured variable PV is shown on the upper display red Sp a1 Sp a2 Setpoint Alarm Alarm SP Value that defines the alarm activation point For the alarms set up...

Page 5: ...signal applied to the input ink Input High Calibration Indication of the full scale calibration signal applied to the input rstr Restore It restores the factory calibrations of the input disregarding...

Page 6: ...4 V 12 to 24 Vdc 24 Vac 10 20 Maximum consumption 6 VA ENVIRONMENTAL CONDITIONS Operation temperature 0 to 50 C Relative humidity 80 30 C For temperatures above 30 C reduce 3 per C Indoor use Installa...

Page 7: ...N1540 NOVUS AUTOMATION 7 Other error messages may indicate hardware problems requiring maintenance service WARRANTY Warranty conditions are available on our website www novusautomation com warranty...

Reviews: