NOVOTEST.UT-1.000 O.M.
Operation manual
Ultrasonic Thickness Gauge NOVOTEST UT-1M
Page. 11
2.3.2 Operation in the "MEASUREMENT" mode
To enter the "MEASUREMENTS" mode press the key «
» or «
» when you are in the
main menu, select the desired mode and confirm your selection by pressing «
», Then the device
will go into the "MEASUREMENT" mode (fig. 2.6).
Figure 2.6 - "MEASUREMENT" mode
When the device is in the "MEASUREMENT" mode, the display shows the battery charge level,
the transmitter position indication, the selected monitoring material, the established propagation
velocity of the ultrasound in the control material, and the measured value.
The speed of the ultrasound in the control material is changed by means of the keys «
» and
«
». Discreteness of speed setting 1 m / s.
By pressing the key «
»The action selection menu is opened in which you can delete the
current measured value "DELETE" Delete all measured values "DELETE". ALL ", go to the main
menu" EXIT ", and also cancel the press and close the "CANCEL" selection menu.
2.3.2.1 Measuring modes
The device operates in several measurement modes, to switch between modes, use keys «
»,
After it is pressed, the menu for selecting the mode of operation appears: "NORM", "STAT", "B
SCAN" and "CALIBER".
The "NORM" mode is intended for carrying out the usual measurement without outputting
additional statistics and information.
The "STAT" mode is intended for measuring with the output of additional statistics (figure 2.7)
with the indication of the maximum and minimum measured values, as well as the number of
measurements.
Figure 2.7 - Measurement mode with additional statistics output
The "B SCAN" mode is used to display the B-Scanned profile of the thickness of the controlled
article (Fig. 2.8). Also, the "B SCAN" mode is used to search for corrosion damage, identify local
abrasions and delaminations when checking control objects with a graphic B-scan.
Figure 2.8 - Measurement mode with the B-Scan output
Battery charge
Material
US velocity
Transducer position