Novis ProCair Prime User Manual Download Page 1

U S E R   G U I D E

ProCair Prime

Summary of Contents for ProCair Prime

Page 1: ...U S E R G U I D E ProCairPrime...

Page 2: ...product improvement and reserves the right to amend specifications presented in this guide Information correct at time of production May2018 2018 Novis Healthcare All rights reserved Important Notice...

Page 3: ...tions 7 8 System Preparation 9 12 Patient Set Up 13 Operation Control Handset 14 Operation Modes of Operation 15 Operation Quick Twist CPR 15 Operation Transport Mode 16 Operation Deflation Storage 17...

Page 4: ...lood and lymphatic fluids through these tissues to provide essential oxygen and remove metabolic waste A full size movable foam sheet can be positioned at the bottom of the mattress working in conjunc...

Page 5: ...Carry bag It is recommended that all packing materials and the User Guide be kept in the carry bag provided for ease of storage and or transport HEAD END FOOT END FOOT END HEAD END RIGHT RIGHT LEFT LE...

Page 6: ...fessionals of sound mind with the ability to read and understand English and Westernised Arabic Numerals Working Environment Temperature 15 C to 35 C 59 F to 95 F Humidity 30 to 75 non condensing Ship...

Page 7: ...cell movement As part of a sensible pressure injury prevention strategy avoid wearing clothing that may cause areas of localised damage due to creases seams objects in pockets etc Never use sharp obje...

Page 8: ...o not use in the presence of smoking materials or open flame air flowing through the mattress will support combustion Do not open the control unit risk of electrical shock Refer servicing to qualified...

Page 9: ...nd On a profiling bed secure the straps around the moveable sections of the base Ensure the buckles are securely fastened and straps are pulled tight Do not secure mattress straps to bed side rails st...

Page 10: ...connectors are secure by pushing the male and female ends together there should be no movement If a connection is open a click will be heard once connector is firmly closed System Preparation 6Feed p...

Page 11: ...ts to pressure setting 02 upon full initial inflation Press AUTO SET to activate automatic pressure setting This feature monitors patient weight every alternation cycle and automatically selects the a...

Page 12: ...onvert to hybrid mode 1 Unzip zippers from the foot end of the mattress and remove the top cover 2Undo 4 press studs on one side of the mattress base cover and remove the foam sheet from its compartme...

Page 13: ...deflated You may need to unzip the cover to feel the cells for inflation 2With the patient lying supine unzip one side of the top cover just past sacral region lower spine 3Slide your hand underneath...

Page 14: ...tton the system will automatically detect patient weight every alternating cycle and adjust pressure accordingly if necessary E Manual Pressure Select After initial inflation the system defaults to se...

Page 15: ...tage the ProCair Prime Universal Therapy System will stop functioning and PF arlam code will be displayed on the Alarm Indicator accompanied by an audible alarm The system will return to its normal op...

Page 16: ...t to the CLOSED position Switch off control unit by long pressing the Power button for 3 secs Open top cover disconnect the air connectors from the control unit and connect the male connector to the f...

Page 17: ...from the system detach the mattress from the bed by unfastening all mattress straps place handset and power cord on the centre of the mattress then fold and roll the mattress from head end to foot end...

Page 18: ...d Foam Sheet Cleaning Swab the mattress base air cells and the foam sheet with a solution of sodium hypochlorite or similar up to 10 000 ppm available chlorine Dry thoroughly before reassembly Do not...

Page 19: ...her zipper run the zipper along the side of the mattress towards the centre of the head end 3 Repeat with other zipper The top cover can now be detached from the mattress base To reattach the top cove...

Page 20: ...ply Check for loose connection on plug and main power is switched on Check condition of power cord and plug Check if mains socket is faulty Control unit does not operate no display or lights Check con...

Page 21: ...2 0 1A MIN MAX PRESSURE 20 60 mmHg 6 mmHg PROTECTION TYPE Class II Type BF INGRESS PROTECTION RATING IP22 HANDSET MODEL HS NBE HANDSET PART NO FO NVS0001 HANDSET CASING MATERIAL ABS MATTRESS DIMENSION...

Page 22: ...onsible and contact your local authority on available options to recycle this product at its end of life Service Life The expected service life of a control unit and a mattress is highly dependent on...

Page 23: ...ation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the...

Page 24: ...60 Hz 50 Hz NOTE UT is the AC mains voltage prior to application of the test level Conducted RF IEC 61000 4 6 3 Vrms 0 15 MHz 80 MHz 3 Vrms 0 15 MHz 80 MHz Portable and mobile RF communications equipm...

Page 25: ...Pulse modulation B 217 Hz 0 2 0 3 9 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation B 18 Hz 2 0 3 28 28 870 930 1 720 1 700 1 990 GSM 1800 CDMA 1900 GSM 1900...

Page 26: ...aired may be replaced by refurbished good of the same type rather than being repaired Refurbished parts may be used to repair goods Novis Healthcare will not be held responsible for any repair other t...

Page 27: ...27 ProCair Prime User Guide Notes...

Page 28: ...Pressure care and patient handling specialists novis com au 1300 738 885...

Reviews: