background image

NovaLynx Corporation 
 

 

 

 

225-RHP-2Sxx 

Page 7 

January 2021 

 

7

 

TROUBLESHOOTING 

 

If the sensor’s output signal appears to be in error or is absent, check the power connections. At the 
sensor cable, measure the battery or the input power source voltage with a voltmeter. Be sure that the 
instrument has been powered up correctly or wait for the next power ON cycle to occur. Check any 
batteries to be sure that they have sufficient charge and an adequate voltage level to power the 
instrument and that all connections are secure. Inspect the battery terminals to ensure that they are 
clean and solidly connected to the battery. 

 

7.1

 

Troubleshooting 4-20 mA Models 

Verify appropriate supply voltage. The transmitter requires a minimum of 10 and a maximum of 35 
VDC at its connection for proper operation. Choose a power supply with a voltage and current rating 
which meets this requirement under all operating conditions. If the power supply is unregulated, make 
sure voltage remains within these limits under all power line conditions. Ripple on the supply should 
not exceed 100 mV. 

Loop Resistance – The maximum allowable loop resistance depends on the power supply voltage. 
Maximum loop voltage drop must not reduce the transmitter voltage below the 10 VDC minimum. 
Maximum loop resistance can be calculated with the following equation. Vps is the power supply 
voltage. 

 

Some receivers, particularly loop powered indicators, may maintain a fixed loop voltage to power the 
device. This voltage drop must also be subtracted from the power supply voltage when calculating the 
voltage margin for the transmitter. The following equation takes this into account. Vrec is the receiver 
fixed voltage. 

 

 

7.2

 

Troubleshooting 0-5 V Output Models: 

Verify appropriate supply voltage. The 0-5 V output requires a DC supply of 15-35 V or an AC supply of 
15-29 V for proper operation maximum. Maximum output load is 5 mA. 

 

7.3

 

Determining the Source of a Failure 

The sensor is one part of the measurement system, and it is possible that the logger or PLC that the 
sensor is being used with is actually the cause of a problem. If the problem cannot be located using the 

Summary of Contents for 225-RHP-2S Series

Page 1: ...User Manual 225 RHP 2S11 4 20 mA output 225 RHP 2S33 0 5 V output Temperature and Relative Humidity Transmitter 225 RHP 2Sxx Transmitter with 380 RHRS Solar Radiation Shield Phone 530 823 7185 Email n...

Page 2: ...e items If packing facilities are not available take the equipment to the nearest Post Office UPS or other freight service and obtain assistance with the packaging Please write the RA on the outside o...

Page 3: ...NSTALLATION 5 5 1 Outdoors 5 5 2 Indoors 5 5 3 Duct Mount 5 6 WIRING 6 6 1 Dual 4 20 mA Output Wiring Model 225 RHP 2S11 6 6 2 Dual 0 5 V Output Wiring Model 225 RHP 2S33 6 7 TROUBLESHOOTING 7 7 1 Tro...

Page 4: ...tion shield NovaLynx 380 RHRS protects the sensor from rain and radiated heat when installed outdoors The weatherproof enclosure cord grip cable entry point and a filter protect the electronics and se...

Page 5: ...st 5 3 Duct Mount The tip of the probe is sometimes inserted into an HVAC duct In this application drill a 1 diameter hole in the duct at the desired location away from fans corners heating and coolin...

Page 6: ...The channels are common on the positive side of the current loop Sensor excitation power is derived from the RH channel so power must always be applied to that channel If the temperature channel is n...

Page 7: ...emains within these limits under all power line conditions Ripple on the supply should not exceed 100 mV Loop Resistance The maximum allowable loop resistance depends on the power supply voltage Maxim...

Page 8: ...the humidity sensor while watching for a change in the meter reading The change may take 10 to 40 seconds To increase the temperature reading hold the probe in your hand for several seconds If no chan...

Reviews: