4.1

 

Power Supply Connection 

The 

200-WS-21P Power Pack

 (sold separately) is provided with ¼" 

spade connectors that fit the terminals at the lower left corner of 
the temperature alarm circuit board. If you are using a different 
power  supply,  crimp  the  provided  accessory  connectors  to  your 
12V AC or DC power source. Plug the AC Adapter into a wall outlet 
and  check  that  the  green  power  LED  on  the  temperature  alarm 
circuit board turns on. 

The display should look similar to the 
illustration  on  the  right.  This  is  the 
operating mode screen.  

NOTE:

  If  you  see  the  word  "Wind:" 

instead  of  "Temp:"  then  the  DIP 
switches  are  in  the  wrong  position. 
Refer  to  the  "DIP  Switch  Settings" 
section of  the  manual to  correct this 
situation.  

 

4.2

 

Menu Navigation 

The programming buttons are below the display. 

4.2.1

 

MENU Button 

Press the MENU button. One of three sub-menus will be displayed (depending on which was 
viewed last). Additional uses of the MENU button will be discussed in the following 
programming section. 

4.2.2

 

UP Button 

Use the UP button to cycle through the menus. 

 

1.

 

Alarm Delays – the delay (seconds) before and after an 
alarm threshold is crossed.  
 

2.

 

Temp Alarms

 

– the set points at which the alarm occurs.  

 

3.

 

Temps – displays the lowest and highest temperatures 
recorded since the CLEAR button was pressed.

  

NOTE: 

Do NOT use the DOWN button to cycle through the menus. A firmware error causes the 

unit to stop at the "Temp Alarms" menu. If this happens, reboot the processor by disconnecting 
power to the unit for a few seconds.

 

 

 MENU       UP 

 DOWN   CLEAR      GO 

ACin / DCin

 

ACin/DCgnd

200-WS-21P

AC Adapter

(sold separately)

(DC+)

(DC-)

A/DM

Menu

Up

Dn

Clr

Go

PROGRAMMING

BUTTONS

POWER LED

Temp: 71.0 (°F)
1:< 32    2:< 28

Temps (°F)     
Lo:-10.0 Hi:78.0

Temp Alarms(°F)
1:< 32    2:< 28

Alarm Delays (s)
ON: 5     OFF: 5

UP

UP

UP

Summary of Contents for 210-421W-A

Page 1: ...x A User Manual Dual Set Point Temperature Alarm 210 421 A Air Temperature 210 421W A Water Soil Temperature Air Temperature Sensor Water and Soil Temperature Sensor Phone 530 823 7185 Email nova nova...

Page 2: ...ns in the case of delicate or sensitive items If packing facilities are not available take the equipment to the nearest Post Office UPS or other freight service and obtain assistance with the packagin...

Page 3: ...P Button 6 4 2 3 DOWN Button 7 4 2 4 CLEAR Button 7 4 2 5 GO Button 7 4 3 Program Alarm Thresholds 7 4 4 Program Alarm Delays 8 4 5 DIP Switch Settings 8 5 TEMPERATURE SENSOR INSTALLATION 9 5 1 Air Te...

Page 4: ...the unit is programmed by push button controls without the need to connect a computer All set points and the minimum and maximum temperatures are stored in non volatile memory DIP switch settings sele...

Page 5: ...connectors supplied Relay specifications Form C SPDT N O and N C Contact rating 3A 24Vdc 115 VAC Alarm ON OFF delay range 0 to 99 seconds Timing accuracy 2 Measurement range 40 to 160 F 40 to 60 C In...

Page 6: ...tion 4 2 Menu Navigation The programming buttons are below the display 4 2 1 MENU Button Press the MENU button One of three sub menus will be displayed depending on which was viewed last Additional us...

Page 7: ...te to the Temp Alarms menu 1 Press GO until the display shows the operating mode screen see GO Button above 2 Press MENU to bring up one of the three sub menu displays 3 Press UP to reach the Temp Ala...

Page 8: ...creen 4 5 DIP Switch Settings The DIP switches are pre set at the factory and should not need to be adjusted However sometimes during installation they get changed accidentally 1 The Dual Set Point co...

Page 9: ...t is on the underside of the bracket The clamp will provide strain relief for the cable and ensure the sensor will not fall out For Pipe Mounting 1 Remove the U bolt from the mounting bracket 2 Secure...

Page 10: ...AC receptacle for power If the controller is inside a building the sensor cable may need to be routed outside so that the sensor can be exposed ambient air conditions 6 2 Controller Mounting There are...

Page 11: ...of this wire to a good earth ground an electrical box with an earth connection or a metal cold water pipe that actually enters the ground Failure to provide a ground may cause erratic operation and ma...

Page 12: ...u have programmed the temperature alarm to the settings you desire and the temperature alarm system is operating properly before you connect power to your external devices 8 ALARM FUNCTIONS The temper...

Page 13: ...Go PROGRAMMING BUTTONS MODE WIND UNITS TEMP UNITS BUZZER WIND MPH DEG F BUZZ TEMP KPH DEG C SILENT ON OFF Wt Wu Tu Si DIP SETTINGS DUAL SET POINT TEMPERATURE ALARM 210 421 A and 210 421W A Alarms OFF...

Reviews: