Nova StatStrip

 Glucometer: 

QC Test 

Quick Reference Guide 

 

 

 

 

 

 

1.

 

From Home screen, 

press 

Login

 

 

2.

 

Enter Operator ID and 

press 

Accept

 

3.

 

From Patient Test 

screen, press 

QC

 

4.

 

Scan the glucose 

test strip barcode. 

 

 

 

 

5.

 

Scan the QC bottle. 

 

6.

 

Insert Test Strip into the 

Meter 

7.

 

Touch drop from QC 

bottle to strip 

8.

 

Result will appear 

within 6 seconds 

 

 

Remember: 

a.

 

QC bottles once opened are good for 3 months. 

b.

 

Test strips once opened are good for 6 months. 

c.

 

Failing one level of the QC requires that BOTH 

low and high levels be repeated. 

d.

 

If QC fails, troubleshoot: 

 

Does the tested QC level match to what 

was scanned? 

 

Are the strips/QC in date? 

 

Was the first QC drop discarded? 

 

Was the QC bottle mixed well? 

e.

 

If QC fails repeatedly, do NOT use meter. Fill out 

Point of Care Service Requisition and send this 

form together with the meter to the POCT Lab. 

9.

 

To accept result, 

press 

Accept

 

 

10.

 

Repeat Steps 3 to 9 for 

the other QC level. 

 