Limitations
The Nova Max Test Strips give accurate results when the follow-
ing limitations are observed:
• The test strips should not be used to diagnose diabetes or to
test newborns.
• Each test strip is for single use only. Do not reuse. Use a new
sterile Nova Max Test Strip each time you test.
• Your test strips are for personal use only. DO NOT share with
others.
• Use only fresh capillary whole blood.
Do not use serum or
plasma.
• There is no effect on blood glucose values for altitudes up to
10,000 feet (3000 meters) above sea level.
• Refer to the Owner’s Guide for operating temperature range
for the monitor.
• Extremes in humidity (higher than 90% and lower than 10%)
may affect results.
• The Nova Max Test Strips are calibrated against plasma.
• The anticoagulant sodium heparin may be used. EDTA is not
recommended for use with Nova Max Test Strips.
• Interferences for elevated levels of acetaminophen, tolaza-
mide, uric acid, bilirubin, ephedrine, and methyldopa may
affect results.
Appendix