IMPORTANT:
Nova recommends cleaning and disinfecting the
Meter with the following EPA Registered product - Clorox® Germicidal
Wipes, EPA Registration #67619-12, or Super Sani-Cloth® Germicidal
Disposable Wipes, EPA Registration #9480-4. Clorox Germicidal
Wipes are available from the following suppliers: Amazon.com:
http://www.amazon.com Clorox Healthcare: 1-800-234-7700 Office
Depot: http://www.officedepot.com.
NOTE:
Cleaning and disinfection may in rare cases damage the
device(s). Meter damage may include plastic housing cracks,
cloudiness, or frosting of the display, legibility or response issues with
the keypad, or battery compartment fluid leakage. Signs of Meter
performance deterioration may include failure to recover proper
control results or the inability to perform a blood test. If you observe
damage due to cleaning and disinfecting, please stop using the Meter
and contact Customer Service.
Basic Upkeep