LipoSurg
Operating instructions LipoSurg, REF 31648, V48/18
21
EN
10
Malfunction and troubleshooting
Malfunction
Cause
Solution
Refer to operating instruction
Device doesn’t work
Control unit is
not switched on
Set the
power switch “I/O”
to “I”
7.1 Switching the device on and
off
Power connection not estab-
lished
Connect the control unit to
the mains power supply
6.2 Connection to the power
supply
Incorrect operating voltage
Check the mains voltage
6.2 Connection to the power
supply
Faulty fuse
Replace the fuse
9.1 Replacing the control unit
fuse
Processor error
Switch main switch “I/O” to
position “O”, wait 10 sec. and
switch back to position “I”.
7.1 Switching the device on
and off
No stroke movement at the
Conform Cannula Handpiece
“LIPO” function is not select-
ed
Select function “LIPO”
7.2 Operation elements
Motor to drive cannula not
activated
Activate motor to drive can-
nula by pressing pedal
7.6 Operation with VARIO pedal
Motor to drive cannula not
connected
Insert motor plug into the
socket at the devices front
5.0 Device overview
6.3.4 Preparing the Conform
Cannula Handpiece
Transmission capsule not
connected properly to the
motor
Press transmission capsule
firmly to the motor until it
engages.
6.3.4 Preparing the Conform
Cannula Handpiece
No delivery of infiltration
solution at the cannula
“INFIL” is not selected
Select function “INFIL”
7.2
Operation elements
Peristaltic pump is not
switched on
Activate peristaltic pump by
pressing the pedal
7.6 Operation with VARIO pedal
Tube set not properly inserted Insert tube set correctly
(watch flow direction)
6.3.5 Preparing the tubing set
Tube set is blocked
Replace tube set
6.3.5 Preparing the tubing set
Infiltration container is not
ventilated
Open air bleed valve at
tube set
6.3.5 Preparing the tubing set
Tube set is dripping
Replace tube set
6.3.5 Preparing the tubing set
Roller clamp is closed
Open roller clamp all the way
it goes
6.3.5 Preparing the tubing set
Pedal is not functioning
Pedal not connected
Insert the pedal plug into the
socket at the devices back
6.3.3 Preparing the pedal
5.0 Device overview
Control unit is not switched
on
Use mains switch, in the back
of the device, to switch on
7.1 Switching device on and off
Faulty operation
Read operation manual care-
fully
7.6 Operation with VARIO pedal
If a fault cannot be rectified, please contact your supplier or an authorized service center. The addresses are provided on
the last page of the operating instructions.