Malfunctions and troubleshooting
Malfunction Cause Solution
Refer to the operating
instructions
Device is not functional
(Indicator light is not on)
Control unit not switched on Set the power switch “I/O”
to “I”
7.1 Switching the device on and off
Power connection not
established
Connect the control unit to
the mains power supply
6.1. Connection to the power
supply
Incorrect operating voltage
Check the mains voltage
6.1. Connection to the power
supply
Faulty fuse
Replace the fuse
9.1 Replacing the control unit fuse
Pump doesn’t work
(Indicator light is on)
Control switch is on too low
position or “Off”
Raise pump performance by
turning control switch up
7.2 Regulation of infiltration
process
Tubing set incorrectly
inserted
Insert tubing set correctly
(note the direction)
6.2 Device preparation
Incorrect operation
Check operating instructions 6.2 Device preparation
Pedal was not pressed
Press pedal
7.2 Regulation of infiltration
process
Roller clamp is closed
Open Roller clamp all the way 6.2 Device preparation
Pedal doesn’t work
(Indicator light is on)
Pedal is not connected
Connect pedal with the
socket on rear of device
5.0 Device overview
6.2 Device preparation
Incorrect operation
Check operating instructions 7.2 Regulation of infiltration
process
6.2 Device preparation
If the problem cannot be solved please contact your supplier or an authorized service center. The addresses are provided in
the appendix of this operation manual under “Service center”.
11
Spare parts list with order numbers
Accessories
REF
ON/OFF pedal, IPX8 --------------------------------------------------------------------------------------------------------------- 1513
VARIO pedal, IPX8 ---------------------------------------------------------------------------------------------------------------- 1501
Stand for irrigation fluid bottle ------------------------------------------------------------------------------------------------ 1770
Tubing set with spike, roller clamp and Luer Lock, sterile, 4 m, single-use --------------------------------------- 6022
Operation manual for Dispenser DP 30 on CD-ROM ----------------------------------------------------------------- 31678*
*A printed version of the operation manual is available at Nouvag AG, free of charge. At order please designate
that you wish to receive a printed version.
To order additional parts, please contact our customer service department.