CHAPTER 8 ERROR MESSAGE INTERPRETING 

 

Error 
Code 

Description 

ERR 01 

The cuff is not well-wrapped 

ERR 02 

Overpressure protection 

ERR 03 

No valid pulse is detected 

ERR 04 

Excessive motion artifact 

ERR 05 

The measurement fails 

ERR 06 

Air leakage 

ERR 07 

Self-test fail 

ERR 08 

Pneumatic system error 

ERR 09 

Signal saturation 

ERR 10

 

Air leakage

 

ERR 11 

Hardware fault 

ERR 12 

Measurement timeout 

ERR 15 

Low battery 

 

Note: the above error codes appear when the NIBP measurement fails. 

 

Summary of Contents for Elegant-1010

Page 1: ...ent Monitor North vision Tech Inc 1 Wearable Multi parameter Patient Monitor Elegant 1010 User Manual Version 2 0 Issued by Sara Lo Date 7 20 2020 Reviewed by Albert Huang Date 7 20 2020 Document No UM PM10 202007 01 North Vision Tech Inc ...

Page 2: ... usage operation repair maintenance and storage etc as well as the safety procedures to protect both the user and equipment Refer to the respective chapters for details The Manual is published in English and we have the ultimate right to explain the Manual Marks in the Manual Caution must be followed to avoid endangering the operator and the patient Attention must be followed to avoid causing dama...

Page 3: ...ng Check the device accessories including cable and cuff before use Stop use if any damage and aging is found Please check the monitor completely to verify that the accessories can function safely and normally Do NOT use the device under the flammable gas condition or in any environment that may lead to explosion The device is forbidden to be used together with other equipment not specified in thi...

Page 4: ...urement 11 4 2 SpO2 Measurement 12 CHAPTER 5 TECHNICAL SPECIFICATIONS 13 5 1 Blood Pressure Measurement 13 5 2 SpO2 Measurement Optional 13 5 3 Pulse Rate Measurement Optional 13 5 4 Dimension and Weight 13 CHAPTER 6 MAINTENANCE AND SERVICE 14 6 1 Technical Maintenance 14 6 1 1 Daily Examination 14 6 1 2 Battery Maintenance 14 6 2 Cleaning and Disinfection of the Main Unit 14 6 3 Cleaning and Disi...

Page 5: ...asive blood pressure NIBP functional oxygen saturation SpO2 and pulse rate PR 1 3 Conformation The Elegant 1010 Monitor consists of the main unit functional accessories cuff SpO2 sensor and the software package running on the host system 1 4 Product Name and Model Product Name North vision Wearable Multi parameter Patient Monitor Model Name Elegant 1010 Monitor 1 5 Working Environment and Power Su...

Page 6: ... is successfully connected with the device 4 external DC power input socket Micro USB interface 5 sensor connector It is used to connect SpO2 probe 6 Cuff it is used for wrapping patient s upper arm for NIBP measurement It is self bound to the device and cannot be disassembled 7 Symbols and description Symbol Description Symbol Description North vision Brand name SN Serial number Type BF applied p...

Page 7: ...a the APP software installed in a host device such as smart phone tablet and etc so before making measurement please install APP software first and make sure the APP software connects to the Elegant 1010 Monitor via wireless of Blue tooth successfully The APP software is installed in the following procedure 1 Copy the APK file to the folder of Download in smart phone s internal space as shown belo...

Page 8: ...User Manual for Multi parameter Patient Monitor North vision Tech Inc 8 2 Run the APK execution file 1 Enter the menu screen 2 Open the File Manager and enter the folder of Download 3 Click the APK file ...

Page 9: ...ent Monitor North vision Tech Inc 9 4 If your smartphone is not allowed to install any apps obtained from unknown sources please confirm to accept it for normal working in your device 5 Return to the directory and click the APK file again ...

Page 10: ...egant 1010 software icon to place it on the desktop 4 Create User Account You can just login directly if you have an account otherwise please create a user account first 1 Click the icon of Elegant 1010 2 Click Login button 3 Click Signup button 4 Key in account s information 5 Input the account and password 6 Login successfully ...

Page 11: ...h left Before making measurement do not eat drink smoke take exercise or have a bath Operating procedure 1 When putting on the cuff spread the cuff and wrap it around the upper arm evenly to appropriate tightness The correct cuff position refers to figure 4 2 Note locate the cuff in such a way that direction mark directs to the lower arm Figure 4 2 Cuff placement 2 Adjust the testee s position Rig...

Page 12: ...n 4 photoplethysmogram 5 pulse bar graph Operation procedure 1 Connect the SpO2 probe to the connector on the sensor connector 2 When the device is powered on insert one finger index finger is preferred the nail should be not too long into the clip of the probe 3 Wait for 2 seconds the device will auto begin to make measurement 4 The measured result will be displayed on the screen as shown in figu...

Page 13: ...ing range 35 100 2 SpO2 measuring accuracy Arms is not greater than 3 for SpO2 range from 70 to 100 Note Arms is defined as root mean square value of deviation according to ISO 9919 3 Resolution 1 4 Low limit alarm value for SpO2 90 Note when the measured SpO2 90 then the over limit SpO2 value blinks 5 3 Pulse Rate Measurement Optional 1 Pulse Rate measuring range 30bpm 240bpm 2 Pulse rate measuri...

Page 14: ...damaged Whether the monitor is on or off the built in battery will be charged as long as the monitor is connected to the provided AC adapter and AC power is on When the battery is full it will stop charging for protecting the battery from damage In order to maintain battery power and prolong battery lifetime please charge the battery routinely Generally charge the battery every 3 months if the dev...

Page 15: ... mis connected parts and try again No blood pressure result The cuff is wrapped around the arm incorrectly Wrap the cuff around the arm correctly No SpO2 result SpO2 probe is not inserted to the sensor connector or the finger is out of probe Plug SpO2 probe to the sensor connector and check if the finger is inserted into the probe properly Can not connect to the host device 1 Bluetooth version is ...

Page 16: ...R 02 Overpressure protection ERR 03 No valid pulse is detected ERR 04 Excessive motion artifact ERR 05 The measurement fails ERR 06 Air leakage ERR 07 Self test fail ERR 08 Pneumatic system error ERR 09 Signal saturation ERR 10 Air leakage ERR 11 Hardware fault ERR 12 Measurement timeout ERR 15 Low battery Note the above error codes appear when the NIBP measurement fails ...

Page 17: ...ion Tech Inc 17 CHAPTER 9 PACKING LIST Item Description Quantity Check 1 Elegant 1010 Monitor One piece OK 2 User Manual One piece OK 3 AC power adapter charger One piece OK 4 SpO2 probe One piece Optional 5 Charging cable One piece OK 6 Handbag One piece OK ...

Page 18: ...m is applicable to adult with age over 18 years Blood Pressure Level SYS mmHg DIA mmHg Optimal blood pressure 120 80 Normal blood pressure 130 85 High normal blood pressure 130 139 85 89 Hypertension Grade 1 140 159 90 99 Hypertension Grade 2 160 179 100 109 Hypertension Grade 3 180 110 正常血壓 正常高值 1 級高血壓 2 級高血壓 3 級高血壓 收縮壓 mmHg 舒張壓 mmHg ...

Page 19: ...l No 15 Gongye E 2nd Rd East Dist HsinChu City 30075 Taiwan R O C Tel 886 3 5771038 Fax 886 3 5771039 E mail ahuang north vision com Website http www north vision com Version 2 0 Issued date July 20 2020 Information is subject to change and or updating without notice Copyright 2020 North vision All rights reserved European Representative CMC Medical Devices Drugs S L C Horacio Lengo N 18 CP 29006 ...

Reviews: