Norsonic nor1256 Instruction Manual Download Page 8

4

Nor1256

Sound Calibrator

Recalibration

At regular intervals – at least once per year – proper 
operation of the sound calibrator should be verified, 
preferably by an accredited acoustic calibration labo-
ratory or the Norsonic factory. The calibration should 
be performed as specified for periodic verification 
in the International standard for sound calibrators 
IEC 60942 (2003), Annex B.
The level should preferably be measured with a labo-
ratory standard microphone type LS2, according to 
IEC 61094-1, where the pressure sensitivities at the 
different frequencies are known with a sufficiently high 
accuracy.

Various types and makes of microphones may have 

different effective load volume. Due to its working  
principle, the Nor1256 has a large effective coupler 
volume. The variations in sound pressure level due 
to variations in effective load volume among micro-
phones are therefore for most applications insignifi-
cant. An equivalent microphone volume of 250 mm³ is 
used as reference load volume for ½” microphones. 
This corresponds to most ½” measuring microphones 
with protecting grid mounted like the Norsonic types 
1220, 1225, 1226, 1227 and 1230. If no adaptor ring is 
used, the nominal reference front volume is 1333 mm³. 
This corresponds to most 1” measuring microphones 
with the protection grid mounted.

The sound pressure levels generated by the calibrator 
may be adjusted by the following procedure (NOTE: it 
will not be possible to switch the calibrator OFF during 
calibration mode):

The nominal attenuation of the A-weighting is 0 dB 

at 1 kHz (reference frequency) and 8,6 dB at 250 Hz. 
The nominal A-weighted sound pressure levels for 
the calibrator will therefore at 250 Hz be 85,4 dB 
and 105,4 dB, respectively. The attenuations of the 
C-weighting are for both frequencies 0,0 dB and C-
weighted levels correspond therefore with the un-
weighted levels.

Most sound level meters are designed to show the 

correct free-field sound pressure level. This means 
that the indication shall correspond to the level in a 
free, progressive field before the sound level meter 
was placed in the field. The size of the microphone 
– and eventually the sound level meter body – will  
modify the pressure just in front of the microphone – 
especially at higher frequencies. For a ½”microphone 
(type WS2F) the pressure increase will typically be 
0,1 dB at 1 kHz. It is therefore a common recom-
mendation to adjust the sound level meters applying 
such microphones to display 93,9 dB and 113,9 dB, 
respectively at 1 kHz.
If microphones other than WS2P are used, the read-
ing from the sound level meter should be corrected 
accordingly.  See the list below for some microphones 
in common use

Summary of Contents for nor1256

Page 1: ...INSTRUCTION MANUAL nor1256 Class 1 sound calibrator Complies with IEC 60942 class 1 ...

Page 2: ... instrumentation described herein Further more Norsonic AS assumes no responsibility for any infringement of the intellectual property rights of third parties wherever applica ble which would result from such use Norsonic AS reserves the right to amend any of the information given in this manual in order to take account of new developments If you wish to communicate with us please feel welcome Our...

Page 3: ...nor1256 Class 1 Sound Calibrator ...

Page 4: ......

Page 5: ...indicated by the LED on the front turning from red to green The feedback ensures that the sound level is nearly independent of the acoustic load volume of the microphone By pressing the dB button the sound pressure level may be switched between 94 dB 114 dB By pressing the Hz button the frequency of the sound may be switched between 1 kHz and 250 Hz Due to the principle of operation the calibratio...

Page 6: ...for the LED above the dB button to change from red to green The level and frequency is indicated in the display of the calibrator After powering the sound calibrator the first combination of frequncy level is 94 dB re 20µPa and 1 kHz Select the appropriate level and frequency for your calibration by pressing the dB or Hz button respectively Normally the background level should be more than 20 dB l...

Page 7: ...ressed until the red LED turns on Acclimatisation Is always a good practice to let the calibrator and the sound measuring equipment reach the prevailing en vironmental conditions mainly temperature before using a sound calibrator If the sound calibrator and the microphone to be calibrated have different temper atures an unstable conditions exits and the calibration will not be reliable Unit for te...

Page 8: ...responds to most 1 measuring microphones with the protection grid mounted The sound pressure levels generated by the calibrator may be adjusted by the following procedure NOTE it will not be possible to switch the calibrator OFF during calibration mode The nominal attenuation of the A weighting is 0 dB at 1 kHz reference frequency and 8 6 dB at 250 Hz The nominal A weighted sound pressure levels f...

Page 9: ...14 0 CEL 186 3F 192 2 192 2F 94 0 bzw 114 0 93 9 bzw 113 9 GRAS 40 AE 40 AF 94 0 bzw 114 0 93 9 bzw 113 9 GRAS Low Noise 40 HL 94 0 93 9 GRAS 40 AU 40 AIR 40 AQ 94 0 bzw 114 0 94 0 bzw 114 0 GRAS 40 AG 40 AK 40 GK 94 0 bzw 114 0 94 0 bzw 114 0 Larson Davis 2540 2541 94 0 bzw 114 0 93 9 bzw 113 9 Mikrotech Gefell MK 102 1 94 0 bzw 114 0 93 7 bzw 113 7 Mikrotech Gefell MK 221 MK222 MK223 94 0 bzw 11...

Page 10: ...hile the display is blinking press Hz to increase the reading or dB to reduce the reading When the correct value is displayed press C F and the value for humidity start to blink While the display is blinking press Hz to increase the reading or dB to reduce the reading When the correct value for humidity is displayed press C F and the calibration mode is terminated and the instrument returns to nor...

Page 11: ...114 0 0 2 dB 94 0 0 2 dB Distortion Max 2 0 Stabilizing time 20 seconds Sensitivity to environmental conditions As specified for IEC 60942 class 1 Microphone size 1 1 and below Adapter for 1 2 micro phone Nor4583 included Power supply Battery operated two LR03 alkaline batteries AAA size also classified as 24A by ANSI NEDA Voltage for specified operation 2 V 3 4 V Automatic switch off if the volta...

Page 12: ......

Page 13: ...of the LVD and EMC Directive The direction of maximum sensitivity emission is with the display facing the antenna This product has been manufactured in compliance with the provisions of the relevant internal Norsonic production standards All our products are tested individually before they leave the factory Calibrated equipment traceable to national and international standards has been used to car...

Page 14: ......

Page 15: ......

Page 16: ... for acoustics from sound calibrators micro phones preamplifers via small handheld sound level meters to advanced yet portable real time analysers but also building acoustics analysers and complete community industry and airport noise monitoring systems Contact your local representative or the factory for information on our complete range of instrumentation ...

Reviews: