Operator's Guide
437D140 Rev. I
DICOM Interface
13-11
Query / Cancel
When the user presses the
button, the query will begin executing. The Query button's caption will
change to
while a query is in progress. This will allow the user to cancel the current query that is in
progress.
The query will return one of three possible responses:
1. No Results
If no results match the criteria selected, then the system will display a dialog indicating such as shown in Figure
13-7, left.
Figure 13-7:
No Query Results (left) vs. Error in Query (right)
2. Error in Query
If an error occurs in the query or the query can not be performed for one reason or another, the system will return
an error code to the user as shown in Figure 13-7 above, right. If the Worklist Query cannot be performed, the
user may not be able to retrieve the patient information from the Worklist server, and the patient may need to be
entered manually. Refer to "Preparing Patient Records" on page 4-21 for detailed instructions on adding or
updating a patient manually.
If the user encounters this message, there may be a configuration problem for the DICOM Worklist settings.
Verify your Worklist settings in the DICOM Preferences window. Refer to "DICOM Preferences Dialog Box:
the Site Tab" on page 13-4. If the problem persists, contact Technical Support.
3. Query Results Area
Should the system successfully execute a query, the results will be returned to the Illuminatus DICOM Patient
Worklist dialog box and populated in the Query Results area as shown in Figure 13-8:
Summary of Contents for Elite 437A150
Page 1: ...DXA Bone Densitometer Operator s Guide Elite Model 437A150 Operator s Guide 437D140 Rev I...
Page 22: ...This page intentionally left blank Operator s Guide 437D140 Rev I 1 8...
Page 56: ...This page intentionally left blank Operator s Guide 437D140 Rev I 2 34...
Page 66: ...11 Click Operator s Guide 437D140 Rev I 3 10 Installation and Setup...
Page 108: ...This page intentionally left blank Operator s Guide 437D140 Rev I 3 52...
Page 124: ...Sample Printed QA Results Report Operator s Guide 437D140 Rev I 4 16 Basic Operation...
Page 137: ...Operator s Guide 437D140 Rev I Basic Operation 4 29...
Page 166: ...This page intentionally left blank Operator s Guide 437D140 Rev I 5 24...
Page 198: ...Operator s Guide 437D140 Rev I 7 20 Scanning AP Spine...
Page 206: ...This page intentionally left blank Operator s Guide 437D140 Rev I 7 28...
Page 238: ...Figure 8 6 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 8 32 Scanning Hip...
Page 290: ...This page intentionally left blank Operator s Guide 437D140 Rev I 10 30...
Page 326: ...Operator s Guide 437D140 Rev I 11 36 Additional Techniques...
Page 412: ...This page intentionally left blank Operator s Guide 437D140 Rev I 14 16...
Page 428: ...This page intentionally left blank Operator s Guide 437D140 Rev I 16 10...
Page 446: ...Label Locations continued Operator s Guide 437D140 Rev I 17 18 Technical Reference...
Page 448: ...Scanner Labels continued Operator s Guide 437D140 Rev I 17 20 Technical Reference...
Page 449: ...Operator s Guide 437D140 Rev I Technical Reference 17 21 Scanner Labels continued...
Page 459: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 9...
Page 466: ...Operator s Guide 437D140 Rev I This page intentionally left blank...
Page 484: ...Operator s Guide 437D140 Rev I This page intentionally left blank...